3 vulnerabilities found for lifecare_pca_infusion_system_firmware by pfizer
FKIE_CVE-2015-1012
Vulnerability from fkie_nvd - Published: 2019-03-25 19:29 - Updated: 2024-11-21 02:24
Severity ?
Summary
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
References
| URL | Tags | ||
|---|---|---|---|
| ics-cert@hq.dhs.gov | https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 | Mitigation, Third Party Advisory, US Government Resource | |
| af854a3a-2127-422b-91ae-364da2661108 | https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 | Mitigation, Third Party Advisory, US Government Resource |
Impacted products
| Vendor | Product | Version | |
|---|---|---|---|
| pfizer | lifecare_pca_infusion_system_firmware | * | |
| pfizer | lifecare_pca_infusion_system | - |
{
"configurations": [
{
"nodes": [
{
"cpeMatch": [
{
"criteria": "cpe:2.3:o:pfizer:lifecare_pca_infusion_system_firmware:*:*:*:*:*:*:*:*",
"matchCriteriaId": "FAA18842-CF1B-4B64-95AB-57404B664C84",
"versionEndIncluding": "5.0",
"vulnerable": true
}
],
"negate": false,
"operator": "OR"
},
{
"cpeMatch": [
{
"criteria": "cpe:2.3:h:pfizer:lifecare_pca_infusion_system:-:*:*:*:*:*:*:*",
"matchCriteriaId": "BC81FC0C-02D8-4BD7-83EB-7E510D241A50",
"vulnerable": false
}
],
"negate": false,
"operator": "OR"
}
],
"operator": "AND"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access."
},
{
"lang": "es",
"value": "Las claves inal\u00e1mbricas se almacenan como texto en claro en la versi\u00f3n 5 de Hospira LifeCare PCA Infusion System. Seg\u00fan Hospira, la versi\u00f3n 3 del LifeCare PCA Infusion System no est\u00e1 indicada para un uso inal\u00e1mbrico, no est\u00e1 equipada con capacidades inal\u00e1mbricas y no se deber\u00eda modificar para que se use en modo inal\u00e1mbrico en un entorno cl\u00ednico. Hospira ha desarrollado una nueva versi\u00f3n del PCS Infusion System, la 7.0, que aborda las vulnerabilidades identificadas. En la versi\u00f3n 7.0, el Puerto 20/FTP y el Puerto 23/TELNET est\u00e1n cerrados por defecto para evitar el acceso no autorizado."
}
],
"id": "CVE-2015-1012",
"lastModified": "2024-11-21T02:24:29.490",
"metrics": {
"cvssMetricV2": [
{
"acInsufInfo": false,
"baseSeverity": "MEDIUM",
"cvssData": {
"accessComplexity": "LOW",
"accessVector": "NETWORK",
"authentication": "NONE",
"availabilityImpact": "NONE",
"baseScore": 5.0,
"confidentialityImpact": "PARTIAL",
"integrityImpact": "NONE",
"vectorString": "AV:N/AC:L/Au:N/C:P/I:N/A:N",
"version": "2.0"
},
"exploitabilityScore": 10.0,
"impactScore": 2.9,
"obtainAllPrivilege": false,
"obtainOtherPrivilege": false,
"obtainUserPrivilege": false,
"source": "nvd@nist.gov",
"type": "Primary",
"userInteractionRequired": false
}
],
"cvssMetricV30": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "NONE",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
"version": "3.0"
},
"exploitabilityScore": 3.9,
"impactScore": 3.6,
"source": "nvd@nist.gov",
"type": "Primary"
}
]
},
"published": "2019-03-25T19:29:00.290",
"references": [
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"vulnStatus": "Modified",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-312"
}
],
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"description": [
{
"lang": "en",
"value": "CWE-200"
}
],
"source": "nvd@nist.gov",
"type": "Primary"
}
]
}
CVE-2015-1012 (GCVE-0-2015-1012)
Vulnerability from cvelistv5 – Published: 2019-03-25 18:20 – Updated: 2024-08-06 04:26
VLAI?
Summary
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Severity ?
No CVSS data available.
CWE
- CWE-312 - Cleartext storage of sensitive information CWE-312
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Hospira | LifeCare PCA Infusion System |
Affected:
<= 5.0
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-06T04:26:11.620Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "LifeCare PCA Infusion System",
"vendor": "Hospira",
"versions": [
{
"status": "affected",
"version": "\u003c= 5.0"
}
]
}
],
"datePublic": "2015-05-13T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-312",
"description": "Cleartext storage of sensitive information CWE-312",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2019-03-25T18:20:12",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"ID": "CVE-2015-1012",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "LifeCare PCA Infusion System",
"version": {
"version_data": [
{
"version_value": "\u003c= 5.0"
}
]
}
}
]
},
"vendor_name": "Hospira"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Cleartext storage of sensitive information CWE-312"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2015-1012",
"datePublished": "2019-03-25T18:20:12",
"dateReserved": "2015-01-10T00:00:00",
"dateUpdated": "2024-08-06T04:26:11.620Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2015-1012 (GCVE-0-2015-1012)
Vulnerability from nvd – Published: 2019-03-25 18:20 – Updated: 2024-08-06 04:26
VLAI?
Summary
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Severity ?
No CVSS data available.
CWE
- CWE-312 - Cleartext storage of sensitive information CWE-312
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Hospira | LifeCare PCA Infusion System |
Affected:
<= 5.0
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-06T04:26:11.620Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "LifeCare PCA Infusion System",
"vendor": "Hospira",
"versions": [
{
"status": "affected",
"version": "\u003c= 5.0"
}
]
}
],
"datePublic": "2015-05-13T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-312",
"description": "Cleartext storage of sensitive information CWE-312",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2019-03-25T18:20:12",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"ID": "CVE-2015-1012",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "LifeCare PCA Infusion System",
"version": {
"version_data": [
{
"version_value": "\u003c= 5.0"
}
]
}
}
]
},
"vendor_name": "Hospira"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Cleartext storage of sensitive information CWE-312"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2015-1012",
"datePublished": "2019-03-25T18:20:12",
"dateReserved": "2015-01-10T00:00:00",
"dateUpdated": "2024-08-06T04:26:11.620Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}