Search criteria

1 vulnerability by Baxter/ Hillrom

CVE-2022-26389 (GCVE-0-2022-26389)

Vulnerability from cvelistv5 – Published: 2025-02-07 17:07 – Updated: 2025-02-07 18:49
VLAI?
Summary
An improper access control vulnerability may allow privilege escalation.This issue affects:  * ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior;  * ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior;  * ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior;  * ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior.
CWE
  • CWE-284 - Improper Access Control
Assigner
Credits
An anonymous user reported these vulnerabilities to Hillrom.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2022-26389",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2025-02-07T18:49:13.343999Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2025-02-07T18:49:22.506Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "ELI 380 Resting  Electrocardiograph",
          "vendor": "Baxter/ Hillrom",
          "versions": [
            {
              "lessThanOrEqual": "2.6.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph",
          "vendor": "Welch Allyn",
          "versions": [
            {
              "lessThanOrEqual": "2.3.1",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "ELI 250c/BUR 250c Resting Electrocardiograph",
          "vendor": "Welch Allyn",
          "versions": [
            {
              "lessThanOrEqual": "2.1.2",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph",
          "vendor": "Welch Allyn",
          "versions": [
            {
              "lessThanOrEqual": "2.2.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "An anonymous user reported these vulnerabilities to Hillrom."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "An improper access control vulnerability may allow privilege escalation.\u003cp\u003eThis issue affects:\u0026nbsp;\u003c/p\u003e\u003cul\u003e\u003cli\u003eELI 380 Resting Electrocardiograph:\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eVersions 2.6.0 and prior;\u003c/span\u003e\u0026nbsp;\u003c/li\u003e\u003cli\u003eELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eVersions 2.3.1 and prior\u003c/span\u003e;\u0026nbsp;\u003c/li\u003e\u003cli\u003eELI 250c/BUR 250c Resting Electrocardiograph:\u0026nbsp;\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eVersions 2.1.2 and prior;\u0026nbsp;\u003c/span\u003e\u003c/li\u003e\u003cli\u003eELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: \n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eVersions 2.2.0 and prior\u003c/span\u003e.\u003c/li\u003e\u003c/ul\u003e"
            }
          ],
          "value": "An improper access control vulnerability may allow privilege escalation.This issue affects:\u00a0\n\n  *  ELI 380 Resting Electrocardiograph:\n\nVersions 2.6.0 and prior;\u00a0\n  *  ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:\n\nVersions 2.3.1 and prior;\u00a0\n  *  ELI 250c/BUR 250c Resting Electrocardiograph:\u00a0Versions 2.1.2 and prior;\u00a0\n  *  ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: \n\nVersions 2.2.0 and prior."
        }
      ],
      "impacts": [
        {
          "capecId": "CAPEC-233",
          "descriptions": [
            {
              "lang": "en",
              "value": "CAPEC-233 Privilege Escalation"
            }
          ]
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 7.7,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "LOW",
            "integrityImpact": "LOW",
            "privilegesRequired": "LOW",
            "scope": "CHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:N/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-284",
              "description": "CWE-284 Improper Access Control",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-02-07T17:07:27.929Z",
        "orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
        "shortName": "Baxter"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-22-167-01"
        },
        {
          "url": "https://hillrom.com/en/responsible-disclosures/"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eHillrom has released software updates for all impacted devices to \naddress these vulnerabilities. New product versions that mitigate these \nvulnerabilities are available as follows:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003eWelch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023\u003c/li\u003e\n\u003cli\u003eWelch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022\u003c/li\u003e\n\u003cli\u003eWelch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eHillrom recommends users upgrade to the latest product versions. \nInformation on how to update these products can be found on the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://hillrom.com/en/responsible-disclosures/\"\u003eHillrom disclosure page\u003c/a\u003e\u003c/p\u003e.\n\n\u003cbr\u003e"
            }
          ],
          "value": "Hillrom has released software updates for all impacted devices to \naddress these vulnerabilities. New product versions that mitigate these \nvulnerabilities are available as follows:\n\n\n\n  *  Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023\n\n  *  Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022\n\n  *  Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023\n\n\n\n\nHillrom recommends users upgrade to the latest product versions. \nInformation on how to update these products can be found on the  Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ \n\n."
        }
      ],
      "source": {
        "advisory": "ICSMA-22-167-01",
        "discovery": "EXTERNAL"
      },
      "title": "Improper Access Control Vulnerability in ELI Electrocardiograph Devices",
      "workarounds": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eHillrom recommends the following workarounds to help reduce risk:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003eApply proper network and physical security controls.\u003c/li\u003e\n\u003cli\u003eEnsure a unique encryption key is configured for ELI Link and Cardiograph.\u003c/li\u003e\n\u003cli\u003eWhere possible, use a firewall to prevent communication on Port 21 \nFTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet \nservice.\u003c/li\u003e\n\u003c/ul\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "Hillrom recommends the following workarounds to help reduce risk:\n\n\n\n  *  Apply proper network and physical security controls.\n\n  *  Ensure a unique encryption key is configured for ELI Link and Cardiograph.\n\n  *  Where possible, use a firewall to prevent communication on Port 21 \nFTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet \nservice."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
    "assignerShortName": "Baxter",
    "cveId": "CVE-2022-26389",
    "datePublished": "2025-02-07T17:07:27.929Z",
    "dateReserved": "2022-03-03T22:04:24.534Z",
    "dateUpdated": "2025-02-07T18:49:22.506Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}