Search criteria
8 vulnerabilities found for CMS8000 Patient Monitor by Contec Health
CVE-2025-1204 (GCVE-0-2025-1204)
Vulnerability from cvelistv5 – Published: 2025-02-25 17:00 – Updated: 2025-02-25 17:23
VLAI?
Summary
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Severity ?
CWE
- CWE-912 - Hidden Functionality
Assigner
References
| URL | Tags | |||||||
|---|---|---|---|---|---|---|---|---|
|
||||||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor |
Affected:
0
|
Credits
Claroty - Team82
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-1204",
"options": [
{
"Exploitation": "poc"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-02-25T17:22:58.082071Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-25T17:23:02.222Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "CMS8000 Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "0"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "Claroty - Team82"
}
],
"datePublic": "2025-02-25T17:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "The \"update\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device."
}
],
"value": "The \"update\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"attackVector": "NETWORK",
"baseScore": 7.7,
"baseSeverity": "HIGH",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "PASSIVE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-912",
"description": "CWE-912 Hidden Functionality",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-02-25T17:00:16.133Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"third-party-advisory"
],
"url": "https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org"
},
{
"tags": [
"government-resource"
],
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
}
],
"source": {
"discovery": "UNKNOWN"
},
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003eIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\u003c/p\u003e"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120."
}
],
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2025-1204",
"datePublished": "2025-02-25T17:00:16.133Z",
"dateReserved": "2025-02-10T17:00:02.210Z",
"dateUpdated": "2025-02-25T17:23:02.222Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2025-0683 (GCVE-0-2025-0683)
Vulnerability from cvelistv5 – Published: 2025-01-30 18:17 – Updated: 2025-02-12 20:41
VLAI?
Title
Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
Summary
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text
patient data to a hard-coded public IP address when a patient is hooked
up to the monitor. This could lead to a leakage of confidential patient
data to any device with that IP address or an attacker in a
machine-in-the-middle scenario.
Severity ?
CWE
- CWE-359 - Exposure of Private Personal Information to an Unauthorized Actor
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor |
Affected:
All versions
|
Credits
An anonymous researcher reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2025-01-31T15:35:10.389Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"url": "https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
},
{
"url": "https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/"
}
],
"title": "CVE Program Container",
"x_generator": {
"engine": "ADPogram 0.0.1"
}
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-0683",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-01-30T19:06:59.525559Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-12T20:41:36.789Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "CMS8000 Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All versions"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "An anonymous researcher reported these vulnerabilities to CISA."
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
}
],
"value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"attackVector": "NETWORK",
"baseScore": 8.2,
"baseSeverity": "HIGH",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "NONE",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "HIGH",
"attackVector": "NETWORK",
"availabilityImpact": "NONE",
"baseScore": 5.9,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-359",
"description": "CWE-359 Exposure of Private Personal Information to an Unauthorized Actor",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-01-31T17:07:47.786Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003ePlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"\u003esafety communication\u003c/a\u003e.\u003c/p\u003e\n\n\u003cbr\u003e"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
}
],
"source": {
"advisory": "ICSMA-25-030-01",
"discovery": "EXTERNAL"
},
"title": "Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2025-0683",
"datePublished": "2025-01-30T18:17:35.867Z",
"dateReserved": "2025-01-23T18:11:20.770Z",
"dateUpdated": "2025-02-12T20:41:36.789Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2025-0626 (GCVE-0-2025-0626)
Vulnerability from cvelistv5 – Published: 2025-01-30 18:17 – Updated: 2025-03-01 17:12
VLAI?
Title
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Summary
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Severity ?
CWE
- CWE-912 - Hidden Functionality
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor |
Affected:
All versions
|
Credits
An anonymous researcher reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2025-01-31T15:37:36.861Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"url": "https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
},
{
"url": "https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/"
}
],
"title": "CVE Program Container",
"x_generator": {
"engine": "ADPogram 0.0.1"
}
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-0626",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-01-30T19:08:00.729447Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-12T20:41:36.928Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "CMS8000 Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All versions"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "An anonymous researcher reported these vulnerabilities to CISA."
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eThe \"monitor\" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.\u003c/span\u003e"
}
],
"value": "The \"monitor\" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"attackVector": "NETWORK",
"baseScore": 7.7,
"baseSeverity": "HIGH",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "PASSIVE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "HIGH",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-912",
"description": "CWE-912 Hidden Functionality",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-03-01T17:12:24.132Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
}
],
"source": {
"advisory": "ICSMA-25-030-01",
"discovery": "EXTERNAL"
},
"title": "Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003eIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\u003c/p\u003e\u003cp\u003ePlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"\u003esafety communication\u003c/a\u003e.\u003c/p\u003e"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
}
],
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2025-0626",
"datePublished": "2025-01-30T18:17:29.717Z",
"dateReserved": "2025-01-21T17:29:48.728Z",
"dateUpdated": "2025-03-01T17:12:24.132Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2024-12248 (GCVE-0-2024-12248)
Vulnerability from cvelistv5 – Published: 2025-01-30 18:17 – Updated: 2025-02-12 16:53
VLAI?
Title
Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor
Summary
Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.
Severity ?
9.8 (Critical)
CWE
- CWE-787 - Out-of-bounds Write
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor |
Affected:
Firmware version smart3250-2.6.27-wlan2.1.7.cramfs
Affected: Firmware version CMS7.820.075.08/0.74(0.75) Affected: Firmware version CMS7.820.120.01/0.93(0.95) |
Credits
An anonymous researcher reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2024-12248",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-01-31T19:13:06.864223Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-12T16:53:16.892Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "CMS8000 Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "Firmware version smart3250-2.6.27-wlan2.1.7.cramfs"
},
{
"status": "affected",
"version": "Firmware version CMS7.820.075.08/0.74(0.75)"
},
{
"status": "affected",
"version": "Firmware version CMS7.820.120.01/0.93(0.95)"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "An anonymous researcher reported these vulnerabilities to CISA."
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eContec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.\u003c/p\u003e\u003cbr\u003e"
}
],
"value": "Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "NETWORK",
"baseScore": 9.3,
"baseSeverity": "CRITICAL",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 9.8,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-787",
"description": "CWE-787 Out-of-bounds Write",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-01-31T16:57:45.995Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003ePlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"\u003esafety communication\u003c/a\u003e.\u003c/p\u003e\n\n\u003cbr\u003e"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
}
],
"source": {
"advisory": "ICSMA-25-030-01",
"discovery": "EXTERNAL"
},
"title": "Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2024-12248",
"datePublished": "2025-01-30T18:17:22.814Z",
"dateReserved": "2024-12-05T15:20:44.585Z",
"dateUpdated": "2025-02-12T16:53:16.892Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2025-1204 (GCVE-0-2025-1204)
Vulnerability from nvd – Published: 2025-02-25 17:00 – Updated: 2025-02-25 17:23
VLAI?
Summary
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Severity ?
CWE
- CWE-912 - Hidden Functionality
Assigner
References
| URL | Tags | |||||||
|---|---|---|---|---|---|---|---|---|
|
||||||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor |
Affected:
0
|
Credits
Claroty - Team82
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-1204",
"options": [
{
"Exploitation": "poc"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-02-25T17:22:58.082071Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-25T17:23:02.222Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "CMS8000 Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "0"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "Claroty - Team82"
}
],
"datePublic": "2025-02-25T17:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "The \"update\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device."
}
],
"value": "The \"update\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"attackVector": "NETWORK",
"baseScore": 7.7,
"baseSeverity": "HIGH",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "PASSIVE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-912",
"description": "CWE-912 Hidden Functionality",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-02-25T17:00:16.133Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"third-party-advisory"
],
"url": "https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org"
},
{
"tags": [
"government-resource"
],
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
}
],
"source": {
"discovery": "UNKNOWN"
},
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003eIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\u003c/p\u003e"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120."
}
],
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2025-1204",
"datePublished": "2025-02-25T17:00:16.133Z",
"dateReserved": "2025-02-10T17:00:02.210Z",
"dateUpdated": "2025-02-25T17:23:02.222Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2025-0683 (GCVE-0-2025-0683)
Vulnerability from nvd – Published: 2025-01-30 18:17 – Updated: 2025-02-12 20:41
VLAI?
Title
Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
Summary
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text
patient data to a hard-coded public IP address when a patient is hooked
up to the monitor. This could lead to a leakage of confidential patient
data to any device with that IP address or an attacker in a
machine-in-the-middle scenario.
Severity ?
CWE
- CWE-359 - Exposure of Private Personal Information to an Unauthorized Actor
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor |
Affected:
All versions
|
Credits
An anonymous researcher reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2025-01-31T15:35:10.389Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"url": "https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
},
{
"url": "https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/"
}
],
"title": "CVE Program Container",
"x_generator": {
"engine": "ADPogram 0.0.1"
}
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-0683",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-01-30T19:06:59.525559Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-12T20:41:36.789Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "CMS8000 Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All versions"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "An anonymous researcher reported these vulnerabilities to CISA."
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
}
],
"value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"attackVector": "NETWORK",
"baseScore": 8.2,
"baseSeverity": "HIGH",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "NONE",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "HIGH",
"attackVector": "NETWORK",
"availabilityImpact": "NONE",
"baseScore": 5.9,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-359",
"description": "CWE-359 Exposure of Private Personal Information to an Unauthorized Actor",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-01-31T17:07:47.786Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003ePlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"\u003esafety communication\u003c/a\u003e.\u003c/p\u003e\n\n\u003cbr\u003e"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
}
],
"source": {
"advisory": "ICSMA-25-030-01",
"discovery": "EXTERNAL"
},
"title": "Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2025-0683",
"datePublished": "2025-01-30T18:17:35.867Z",
"dateReserved": "2025-01-23T18:11:20.770Z",
"dateUpdated": "2025-02-12T20:41:36.789Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2025-0626 (GCVE-0-2025-0626)
Vulnerability from nvd – Published: 2025-01-30 18:17 – Updated: 2025-03-01 17:12
VLAI?
Title
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Summary
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Severity ?
CWE
- CWE-912 - Hidden Functionality
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor |
Affected:
All versions
|
Credits
An anonymous researcher reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2025-01-31T15:37:36.861Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"url": "https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
},
{
"url": "https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/"
}
],
"title": "CVE Program Container",
"x_generator": {
"engine": "ADPogram 0.0.1"
}
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-0626",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-01-30T19:08:00.729447Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-12T20:41:36.928Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "CMS8000 Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All versions"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "An anonymous researcher reported these vulnerabilities to CISA."
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eThe \"monitor\" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.\u003c/span\u003e"
}
],
"value": "The \"monitor\" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"attackVector": "NETWORK",
"baseScore": 7.7,
"baseSeverity": "HIGH",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "PASSIVE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "HIGH",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-912",
"description": "CWE-912 Hidden Functionality",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-03-01T17:12:24.132Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
}
],
"source": {
"advisory": "ICSMA-25-030-01",
"discovery": "EXTERNAL"
},
"title": "Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003eIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\u003c/p\u003e\u003cp\u003ePlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"\u003esafety communication\u003c/a\u003e.\u003c/p\u003e"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
}
],
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2025-0626",
"datePublished": "2025-01-30T18:17:29.717Z",
"dateReserved": "2025-01-21T17:29:48.728Z",
"dateUpdated": "2025-03-01T17:12:24.132Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2024-12248 (GCVE-0-2024-12248)
Vulnerability from nvd – Published: 2025-01-30 18:17 – Updated: 2025-02-12 16:53
VLAI?
Title
Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor
Summary
Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.
Severity ?
9.8 (Critical)
CWE
- CWE-787 - Out-of-bounds Write
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor |
Affected:
Firmware version smart3250-2.6.27-wlan2.1.7.cramfs
Affected: Firmware version CMS7.820.075.08/0.74(0.75) Affected: Firmware version CMS7.820.120.01/0.93(0.95) |
Credits
An anonymous researcher reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2024-12248",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-01-31T19:13:06.864223Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-12T16:53:16.892Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "CMS8000 Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "Firmware version smart3250-2.6.27-wlan2.1.7.cramfs"
},
{
"status": "affected",
"version": "Firmware version CMS7.820.075.08/0.74(0.75)"
},
{
"status": "affected",
"version": "Firmware version CMS7.820.120.01/0.93(0.95)"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "An anonymous researcher reported these vulnerabilities to CISA."
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eContec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.\u003c/p\u003e\u003cbr\u003e"
}
],
"value": "Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "NETWORK",
"baseScore": 9.3,
"baseSeverity": "CRITICAL",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 9.8,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-787",
"description": "CWE-787 Out-of-bounds Write",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-01-31T16:57:45.995Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003ePlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"\u003esafety communication\u003c/a\u003e.\u003c/p\u003e\n\n\u003cbr\u003e"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
}
],
"source": {
"advisory": "ICSMA-25-030-01",
"discovery": "EXTERNAL"
},
"title": "Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2024-12248",
"datePublished": "2025-01-30T18:17:22.814Z",
"dateReserved": "2024-12-05T15:20:44.585Z",
"dateUpdated": "2025-02-12T16:53:16.892Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}