Search criteria

2 vulnerabilities found for IntelliBridge Enterprise (IBE) by Philips

CVE-2020-12023 (GCVE-0-2020-12023)

Vulnerability from cvelistv5 – Published: 2020-06-11 19:11 – Updated: 2025-06-04 22:02
VLAI?
Title
Philips IntelliBridge Enterprise IBE Insertion of Sensitive Information into Log File
Summary
Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files.
CWE
Assigner
Impacted products
Vendor Product Version
Philips IntelliBridge Enterprise (IBE) Affected: 0 , < Versions B.12 (custom)
Create a notification for this product.
Credits
Customer Indiana University Health reported this vulnerability to Philips.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T11:48:57.842Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://www.us-cert.gov/ics/advisories/icsma-20-163-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "IntelliBridge Enterprise (IBE)",
          "vendor": "Philips",
          "versions": [
            {
              "lessThan": "Versions B.12",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Customer Indiana University Health reported this vulnerability to Philips."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePhilips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files.\u003c/p\u003e"
            }
          ],
          "value": "Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "ADJACENT_NETWORK",
            "availabilityImpact": "NONE",
            "baseScore": 2,
            "baseSeverity": "LOW",
            "confidentialityImpact": "LOW",
            "integrityImpact": "NONE",
            "privilegesRequired": "HIGH",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:A/AC:H/PR:H/UI:N/S:U/C:L/I:N/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-532",
              "description": "CWE-532",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-06-04T22:02:51.755Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-163-01"
        },
        {
          "url": "https://www.philips.com/a-w/security/security-advisories/product-security-2020.html#2020_archive"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePhilips plans a new release (IBE B.13) by the end of Q4 2020 that \nremediates the security risk by not logging the plain text user \ncredentials in the log file.\u003c/p\u003e\u003cp\u003eUsers with questions regarding their specific Philips IntelliBridge \nEnterprise installation should contact their local Philips service \nsupport team, or regional service support.\u003c/p\u003e\u003cp\u003eUsers can see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"http://www.philips.com/productsecurity\"\u003ePhilips advisory\u003c/a\u003e\u0026nbsp;for more details, and contact the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.usa.philips.com/healthcare/solutions/customer-service-solutions\"\u003ePhilips Customer Service Solutions\u003c/a\u003e team for additional guidance. Please see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/productsecurity\"\u003ePhilips product security website\u003c/a\u003e for the latest security information for Philips products. \n\n\u003cbr\u003e\u003c/p\u003e"
            }
          ],
          "value": "Philips plans a new release (IBE B.13) by the end of Q4 2020 that \nremediates the security risk by not logging the plain text user \ncredentials in the log file.\n\nUsers with questions regarding their specific Philips IntelliBridge \nEnterprise installation should contact their local Philips service \nsupport team, or regional service support.\n\nUsers can see the  Philips advisory http://www.philips.com/productsecurity \u00a0for more details, and contact the  Philips Customer Service Solutions https://www.usa.philips.com/healthcare/solutions/customer-service-solutions  team for additional guidance. Please see the  Philips product security website https://www.philips.com/productsecurity  for the latest security information for Philips products."
        }
      ],
      "source": {
        "advisory": "ICSMA-20-163-01",
        "discovery": "EXTERNAL"
      },
      "title": "Philips IntelliBridge Enterprise IBE Insertion of Sensitive Information into Log File",
      "workarounds": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eAs an interim mitigation to this vulnerability, Philips recommends the following:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003eThe IBE transaction logs are only accessible with administrative \nprivileges. An additional account can be created on the IBE system with \nlimited privileges, for service engineers.\u003c/li\u003e\n\u003cli\u003eReduce log retention to an acceptable timeframe that allows for recovery activities.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eUsers with questions regarding their specific Philips IntelliBridge \nEnterprise installation should contact their local Philips service \nsupport team, or regional service support.\u003c/p\u003e\u003cp\u003eUsers can see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"http://www.philips.com/productsecurity\"\u003ePhilips advisory\u003c/a\u003e\u0026nbsp;for more details, and contact the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.usa.philips.com/healthcare/solutions/customer-service-solutions\"\u003ePhilips Customer Service Solutions\u003c/a\u003e team for additional guidance. Please see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/productsecurity\"\u003ePhilips product security website\u003c/a\u003e for the latest security information for Philips products. \n\n\u003c/p\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "As an interim mitigation to this vulnerability, Philips recommends the following:\n\n\n\n  *  The IBE transaction logs are only accessible with administrative \nprivileges. An additional account can be created on the IBE system with \nlimited privileges, for service engineers.\n\n  *  Reduce log retention to an acceptable timeframe that allows for recovery activities.\n\n\n\n\nUsers with questions regarding their specific Philips IntelliBridge \nEnterprise installation should contact their local Philips service \nsupport team, or regional service support.\n\nUsers can see the  Philips advisory http://www.philips.com/productsecurity \u00a0for more details, and contact the  Philips Customer Service Solutions https://www.usa.philips.com/healthcare/solutions/customer-service-solutions  team for additional guidance. Please see the  Philips product security website https://www.philips.com/productsecurity  for the latest security information for Philips products."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "ID": "CVE-2020-12023",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Philips IntelliBridge Enterprise (IBE)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS)"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "n/a"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "INSERTION OF SENSITIVE INFORMATION INTO LOG FILE CWE-532"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.us-cert.gov/ics/advisories/icsma-20-163-01",
              "refsource": "MISC",
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-163-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2020-12023",
    "datePublished": "2020-06-11T19:11:03",
    "dateReserved": "2020-04-21T00:00:00",
    "dateUpdated": "2025-06-04T22:02:51.755Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2020-12023 (GCVE-0-2020-12023)

Vulnerability from nvd – Published: 2020-06-11 19:11 – Updated: 2025-06-04 22:02
VLAI?
Title
Philips IntelliBridge Enterprise IBE Insertion of Sensitive Information into Log File
Summary
Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files.
CWE
Assigner
Impacted products
Vendor Product Version
Philips IntelliBridge Enterprise (IBE) Affected: 0 , < Versions B.12 (custom)
Create a notification for this product.
Credits
Customer Indiana University Health reported this vulnerability to Philips.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T11:48:57.842Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://www.us-cert.gov/ics/advisories/icsma-20-163-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "IntelliBridge Enterprise (IBE)",
          "vendor": "Philips",
          "versions": [
            {
              "lessThan": "Versions B.12",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Customer Indiana University Health reported this vulnerability to Philips."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePhilips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files.\u003c/p\u003e"
            }
          ],
          "value": "Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "ADJACENT_NETWORK",
            "availabilityImpact": "NONE",
            "baseScore": 2,
            "baseSeverity": "LOW",
            "confidentialityImpact": "LOW",
            "integrityImpact": "NONE",
            "privilegesRequired": "HIGH",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:A/AC:H/PR:H/UI:N/S:U/C:L/I:N/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-532",
              "description": "CWE-532",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-06-04T22:02:51.755Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-163-01"
        },
        {
          "url": "https://www.philips.com/a-w/security/security-advisories/product-security-2020.html#2020_archive"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePhilips plans a new release (IBE B.13) by the end of Q4 2020 that \nremediates the security risk by not logging the plain text user \ncredentials in the log file.\u003c/p\u003e\u003cp\u003eUsers with questions regarding their specific Philips IntelliBridge \nEnterprise installation should contact their local Philips service \nsupport team, or regional service support.\u003c/p\u003e\u003cp\u003eUsers can see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"http://www.philips.com/productsecurity\"\u003ePhilips advisory\u003c/a\u003e\u0026nbsp;for more details, and contact the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.usa.philips.com/healthcare/solutions/customer-service-solutions\"\u003ePhilips Customer Service Solutions\u003c/a\u003e team for additional guidance. Please see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/productsecurity\"\u003ePhilips product security website\u003c/a\u003e for the latest security information for Philips products. \n\n\u003cbr\u003e\u003c/p\u003e"
            }
          ],
          "value": "Philips plans a new release (IBE B.13) by the end of Q4 2020 that \nremediates the security risk by not logging the plain text user \ncredentials in the log file.\n\nUsers with questions regarding their specific Philips IntelliBridge \nEnterprise installation should contact their local Philips service \nsupport team, or regional service support.\n\nUsers can see the  Philips advisory http://www.philips.com/productsecurity \u00a0for more details, and contact the  Philips Customer Service Solutions https://www.usa.philips.com/healthcare/solutions/customer-service-solutions  team for additional guidance. Please see the  Philips product security website https://www.philips.com/productsecurity  for the latest security information for Philips products."
        }
      ],
      "source": {
        "advisory": "ICSMA-20-163-01",
        "discovery": "EXTERNAL"
      },
      "title": "Philips IntelliBridge Enterprise IBE Insertion of Sensitive Information into Log File",
      "workarounds": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eAs an interim mitigation to this vulnerability, Philips recommends the following:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003eThe IBE transaction logs are only accessible with administrative \nprivileges. An additional account can be created on the IBE system with \nlimited privileges, for service engineers.\u003c/li\u003e\n\u003cli\u003eReduce log retention to an acceptable timeframe that allows for recovery activities.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eUsers with questions regarding their specific Philips IntelliBridge \nEnterprise installation should contact their local Philips service \nsupport team, or regional service support.\u003c/p\u003e\u003cp\u003eUsers can see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"http://www.philips.com/productsecurity\"\u003ePhilips advisory\u003c/a\u003e\u0026nbsp;for more details, and contact the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.usa.philips.com/healthcare/solutions/customer-service-solutions\"\u003ePhilips Customer Service Solutions\u003c/a\u003e team for additional guidance. Please see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/productsecurity\"\u003ePhilips product security website\u003c/a\u003e for the latest security information for Philips products. \n\n\u003c/p\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "As an interim mitigation to this vulnerability, Philips recommends the following:\n\n\n\n  *  The IBE transaction logs are only accessible with administrative \nprivileges. An additional account can be created on the IBE system with \nlimited privileges, for service engineers.\n\n  *  Reduce log retention to an acceptable timeframe that allows for recovery activities.\n\n\n\n\nUsers with questions regarding their specific Philips IntelliBridge \nEnterprise installation should contact their local Philips service \nsupport team, or regional service support.\n\nUsers can see the  Philips advisory http://www.philips.com/productsecurity \u00a0for more details, and contact the  Philips Customer Service Solutions https://www.usa.philips.com/healthcare/solutions/customer-service-solutions  team for additional guidance. Please see the  Philips product security website https://www.philips.com/productsecurity  for the latest security information for Philips products."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "ID": "CVE-2020-12023",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Philips IntelliBridge Enterprise (IBE)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS)"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "n/a"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "INSERTION OF SENSITIVE INFORMATION INTO LOG FILE CWE-532"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.us-cert.gov/ics/advisories/icsma-20-163-01",
              "refsource": "MISC",
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-163-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2020-12023",
    "datePublished": "2020-06-11T19:11:03",
    "dateReserved": "2020-04-21T00:00:00",
    "dateUpdated": "2025-06-04T22:02:51.755Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}