Search criteria

4 vulnerabilities found for Intellispace Cardiovascular (ISCV) by Philips

CVE-2025-2229 (GCVE-0-2025-2229)

Vulnerability from cvelistv5 – Published: 2025-03-13 18:17 – Updated: 2025-03-17 20:04
VLAI?
Summary
A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.
CWE
Assigner
Impacted products
Vendor Product Version
Philips Intellispace Cardiovascular (ISCV) Affected: 0 , ≤ 4.1 (custom)
Create a notification for this product.
Credits
Joe Dillon reported these vulnerabilities to Philips.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2025-2229",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2025-03-13T19:30:31.475698Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2025-03-13T19:30:42.059Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Intellispace Cardiovascular (ISCV)",
          "vendor": "Philips",
          "versions": [
            {
              "lessThanOrEqual": "4.1",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Joe Dillon reported these vulnerabilities to Philips."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "A token is created using the username, current date/time, and a fixed \nAES-128 encryption key, which is the same across all installations."
            }
          ],
          "value": "A token is created using the username, current date/time, and a fixed \nAES-128 encryption key, which is the same across all installations."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "LOCAL",
            "availabilityImpact": "NONE",
            "baseScore": 7.7,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "LOCAL",
            "baseScore": 8.5,
            "baseSeverity": "HIGH",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "NONE",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "HIGH",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-1391",
              "description": "CWE-1391",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-03-17T20:04:56.100Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01"
        },
        {
          "url": "https://www.philips.com/a-w/security/security-advisories.html"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePhilips recommends the following mitigations:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResolved in ISCV 4.2 build 20589, which was released in May 2019.\u003c/span\u003e\n\n\u003c/li\u003e\n\n\u003cli\u003ePhilips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\u003c/li\u003e\n\u003cli\u003ePlease contact a local Philips sales (service) representative to learn how to engage this upgrade process.\u003c/li\u003e\n\u003cli\u003eFor managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eRefer to the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/a-w/security/security-advisories.html\"\u003ePhilips advisory\u003c/a\u003efor more details.\n\n\u003cbr\u003e\u003c/p\u003e"
            }
          ],
          "value": "Philips recommends the following mitigations:\n\n\n\n  *  \n\nResolved in ISCV 4.2 build 20589, which was released in May 2019.\n\n\n\n\n  *  Philips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\n\n  *  Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.\n\n  *  For managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\n\n\n\n\nRefer to the  Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details."
        }
      ],
      "source": {
        "advisory": "ICSMA-25-072-01",
        "discovery": "EXTERNAL"
      },
      "title": "Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2025-2229",
    "datePublished": "2025-03-13T18:17:31.256Z",
    "dateReserved": "2025-03-11T20:11:49.638Z",
    "dateUpdated": "2025-03-17T20:04:56.100Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2025-2230 (GCVE-0-2025-2230)

Vulnerability from cvelistv5 – Published: 2025-03-13 18:14 – Updated: 2025-03-17 20:05
VLAI?
Summary
A flaw exists in the Windows login flow where an AuthContext token can be exploited for replay attacks and authentication bypass.
CWE
  • CWE-287 - Improper Authentication
Assigner
Impacted products
Vendor Product Version
Philips Intellispace Cardiovascular (ISCV) Affected: 0 , ≤ 5.1 (custom)
Create a notification for this product.
Credits
Joe Dillon reported these vulnerabilities to Philips.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2025-2230",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2025-03-13T19:33:45.514056Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2025-03-13T19:34:02.967Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Intellispace Cardiovascular (ISCV)",
          "vendor": "Philips",
          "versions": [
            {
              "lessThanOrEqual": "5.1",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Joe Dillon reported these vulnerabilities to Philips."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "A flaw exists in the Windows login flow where an AuthContext token can \nbe exploited for replay attacks and authentication bypass."
            }
          ],
          "value": "A flaw exists in the Windows login flow where an AuthContext token can \nbe exploited for replay attacks and authentication bypass."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "LOCAL",
            "availabilityImpact": "NONE",
            "baseScore": 7.7,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "LOCAL",
            "baseScore": 8.5,
            "baseSeverity": "HIGH",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "NONE",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "HIGH",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-287",
              "description": "CWE-287 Improper Authentication",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-03-17T20:05:56.939Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01"
        },
        {
          "url": "https://www.philips.com/a-w/security/security-advisories.html"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePhilips recommends the following mitigations:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResolved in ISCV 5.2, which was released in September 2020.\u003c/span\u003e\n\n\u003c/li\u003e\n\n\u003cli\u003ePhilips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\u003c/li\u003e\n\u003cli\u003ePlease contact a local Philips sales (service) representative to learn how to engage this upgrade process.\u003c/li\u003e\n\u003cli\u003eFor managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eRefer to the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/a-w/security/security-advisories.html\"\u003ePhilips advisory\u003c/a\u003efor more details.\n\n\u003cbr\u003e\u003c/p\u003e"
            }
          ],
          "value": "Philips recommends the following mitigations:\n\n\n\n  *  \n\nResolved in ISCV 5.2, which was released in September 2020.\n\n\n\n\n  *  Philips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\n\n  *  Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.\n\n  *  For managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\n\n\n\n\nRefer to the  Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details."
        }
      ],
      "source": {
        "advisory": "ICSMA-25-072-01",
        "discovery": "EXTERNAL"
      },
      "title": "Philips Intellispace Cardiovascular (ISCV) Improper Authentication",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2025-2230",
    "datePublished": "2025-03-13T18:14:43.573Z",
    "dateReserved": "2025-03-11T20:14:50.017Z",
    "dateUpdated": "2025-03-17T20:05:56.939Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2025-2229 (GCVE-0-2025-2229)

Vulnerability from nvd – Published: 2025-03-13 18:17 – Updated: 2025-03-17 20:04
VLAI?
Summary
A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.
CWE
Assigner
Impacted products
Vendor Product Version
Philips Intellispace Cardiovascular (ISCV) Affected: 0 , ≤ 4.1 (custom)
Create a notification for this product.
Credits
Joe Dillon reported these vulnerabilities to Philips.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2025-2229",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2025-03-13T19:30:31.475698Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2025-03-13T19:30:42.059Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Intellispace Cardiovascular (ISCV)",
          "vendor": "Philips",
          "versions": [
            {
              "lessThanOrEqual": "4.1",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Joe Dillon reported these vulnerabilities to Philips."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "A token is created using the username, current date/time, and a fixed \nAES-128 encryption key, which is the same across all installations."
            }
          ],
          "value": "A token is created using the username, current date/time, and a fixed \nAES-128 encryption key, which is the same across all installations."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "LOCAL",
            "availabilityImpact": "NONE",
            "baseScore": 7.7,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "LOCAL",
            "baseScore": 8.5,
            "baseSeverity": "HIGH",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "NONE",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "HIGH",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-1391",
              "description": "CWE-1391",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-03-17T20:04:56.100Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01"
        },
        {
          "url": "https://www.philips.com/a-w/security/security-advisories.html"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePhilips recommends the following mitigations:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResolved in ISCV 4.2 build 20589, which was released in May 2019.\u003c/span\u003e\n\n\u003c/li\u003e\n\n\u003cli\u003ePhilips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\u003c/li\u003e\n\u003cli\u003ePlease contact a local Philips sales (service) representative to learn how to engage this upgrade process.\u003c/li\u003e\n\u003cli\u003eFor managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eRefer to the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/a-w/security/security-advisories.html\"\u003ePhilips advisory\u003c/a\u003efor more details.\n\n\u003cbr\u003e\u003c/p\u003e"
            }
          ],
          "value": "Philips recommends the following mitigations:\n\n\n\n  *  \n\nResolved in ISCV 4.2 build 20589, which was released in May 2019.\n\n\n\n\n  *  Philips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\n\n  *  Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.\n\n  *  For managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\n\n\n\n\nRefer to the  Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details."
        }
      ],
      "source": {
        "advisory": "ICSMA-25-072-01",
        "discovery": "EXTERNAL"
      },
      "title": "Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2025-2229",
    "datePublished": "2025-03-13T18:17:31.256Z",
    "dateReserved": "2025-03-11T20:11:49.638Z",
    "dateUpdated": "2025-03-17T20:04:56.100Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2025-2230 (GCVE-0-2025-2230)

Vulnerability from nvd – Published: 2025-03-13 18:14 – Updated: 2025-03-17 20:05
VLAI?
Summary
A flaw exists in the Windows login flow where an AuthContext token can be exploited for replay attacks and authentication bypass.
CWE
  • CWE-287 - Improper Authentication
Assigner
Impacted products
Vendor Product Version
Philips Intellispace Cardiovascular (ISCV) Affected: 0 , ≤ 5.1 (custom)
Create a notification for this product.
Credits
Joe Dillon reported these vulnerabilities to Philips.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2025-2230",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2025-03-13T19:33:45.514056Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2025-03-13T19:34:02.967Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Intellispace Cardiovascular (ISCV)",
          "vendor": "Philips",
          "versions": [
            {
              "lessThanOrEqual": "5.1",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Joe Dillon reported these vulnerabilities to Philips."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "A flaw exists in the Windows login flow where an AuthContext token can \nbe exploited for replay attacks and authentication bypass."
            }
          ],
          "value": "A flaw exists in the Windows login flow where an AuthContext token can \nbe exploited for replay attacks and authentication bypass."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "LOCAL",
            "availabilityImpact": "NONE",
            "baseScore": 7.7,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "LOCAL",
            "baseScore": 8.5,
            "baseSeverity": "HIGH",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "NONE",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "HIGH",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-287",
              "description": "CWE-287 Improper Authentication",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-03-17T20:05:56.939Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01"
        },
        {
          "url": "https://www.philips.com/a-w/security/security-advisories.html"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePhilips recommends the following mitigations:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResolved in ISCV 5.2, which was released in September 2020.\u003c/span\u003e\n\n\u003c/li\u003e\n\n\u003cli\u003ePhilips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\u003c/li\u003e\n\u003cli\u003ePlease contact a local Philips sales (service) representative to learn how to engage this upgrade process.\u003c/li\u003e\n\u003cli\u003eFor managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eRefer to the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/a-w/security/security-advisories.html\"\u003ePhilips advisory\u003c/a\u003efor more details.\n\n\u003cbr\u003e\u003c/p\u003e"
            }
          ],
          "value": "Philips recommends the following mitigations:\n\n\n\n  *  \n\nResolved in ISCV 5.2, which was released in September 2020.\n\n\n\n\n  *  Philips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\n\n  *  Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.\n\n  *  For managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\n\n\n\n\nRefer to the  Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details."
        }
      ],
      "source": {
        "advisory": "ICSMA-25-072-01",
        "discovery": "EXTERNAL"
      },
      "title": "Philips Intellispace Cardiovascular (ISCV) Improper Authentication",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2025-2230",
    "datePublished": "2025-03-13T18:14:43.573Z",
    "dateReserved": "2025-03-11T20:14:50.017Z",
    "dateUpdated": "2025-03-17T20:05:56.939Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}