Search criteria
15 vulnerabilities found for cms8000_firmware by contechealth
FKIE_CVE-2022-3027
Vulnerability from fkie_nvd - Published: 2022-09-13 15:15 - Updated: 2024-11-21 07:18
Severity ?
5.7 (Medium) - CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H
5.7 (Medium) - CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:H/A:N
5.7 (Medium) - CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:H/A:N
Summary
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
References
| URL | Tags | ||
|---|---|---|---|
| ics-cert@hq.dhs.gov | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource | |
| af854a3a-2127-422b-91ae-364da2661108 | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource |
Impacted products
| Vendor | Product | Version | |
|---|---|---|---|
| contechealth | cms8000_firmware | - | |
| contechealth | cms8000 | - |
{
"configurations": [
{
"nodes": [
{
"cpeMatch": [
{
"criteria": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
"matchCriteriaId": "2C197D62-6F35-4B87-A721-BDB696EA240F",
"vulnerable": true
}
],
"negate": false,
"operator": "OR"
},
{
"cpeMatch": [
{
"criteria": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
"matchCriteriaId": "3A0CD9FA-68D7-4EEE-93A5-97275D84E2D3",
"vulnerable": false
}
],
"negate": false,
"operator": "OR"
}
],
"operator": "AND"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information."
},
{
"lang": "es",
"value": "El dispositivo CMS8000 no controla o sanea apropiadamente el nombre SSID de un nuevo punto de acceso Wi-Fi. Un actor de la amenaza podr\u00eda crear un SSID con un nombre malicioso, incluyendo caracteres no est\u00e1ndar que, cuando el dispositivo intenta conectarse al SSID malicioso, el dispositivo puede ser explotado para escribir archivos arbitrarios o mostrar informaci\u00f3n incorrecta"
}
],
"id": "CVE-2022-3027",
"lastModified": "2024-11-21T07:18:39.823",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 5.7,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"exploitabilityScore": 2.1,
"impactScore": 3.6,
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "NONE",
"baseScore": 5.7,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:H/A:N",
"version": "3.1"
},
"exploitabilityScore": 2.1,
"impactScore": 3.6,
"source": "nvd@nist.gov",
"type": "Primary"
}
]
},
"published": "2022-09-13T15:15:09.257",
"references": [
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"vulnStatus": "Modified",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-284"
}
],
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"description": [
{
"lang": "en",
"value": "NVD-CWE-noinfo"
}
],
"source": "nvd@nist.gov",
"type": "Primary"
}
]
}
FKIE_CVE-2022-36385
Vulnerability from fkie_nvd - Published: 2022-09-13 15:15 - Updated: 2024-11-21 07:12
Severity ?
6.8 (Medium) - CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
6.8 (Medium) - CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
6.8 (Medium) - CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Summary
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
References
| URL | Tags | ||
|---|---|---|---|
| ics-cert@hq.dhs.gov | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource | |
| af854a3a-2127-422b-91ae-364da2661108 | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource |
Impacted products
| Vendor | Product | Version | |
|---|---|---|---|
| contechealth | cms8000_firmware | - | |
| contechealth | cms8000 | - |
{
"configurations": [
{
"nodes": [
{
"cpeMatch": [
{
"criteria": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
"matchCriteriaId": "2C197D62-6F35-4B87-A721-BDB696EA240F",
"vulnerable": true
}
],
"negate": false,
"operator": "OR"
},
{
"cpeMatch": [
{
"criteria": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
"matchCriteriaId": "3A0CD9FA-68D7-4EEE-93A5-97275D84E2D3",
"vulnerable": false
}
],
"negate": false,
"operator": "OR"
}
],
"operator": "AND"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device."
},
{
"lang": "es",
"value": "Un actor de la amenaza con acceso moment\u00e1neo al dispositivo puede conectar una unidad USB y llevar a cabo una actualizaci\u00f3n de firmware malicioso, lo que resulta en cambios permanentes en la funcionalidad del dispositivo. No se presenta autenticaci\u00f3n ni controles para evitar que un actor de amenaza modifique maliciosamente el firmware y lleve a cabo un ataque drive-by para cargar el firmware en cualquier dispositivo CMS8000"
}
],
"id": "CVE-2022-36385",
"lastModified": "2024-11-21T07:12:55.063",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"exploitabilityScore": 0.9,
"impactScore": 5.9,
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"exploitabilityScore": 0.9,
"impactScore": 5.9,
"source": "nvd@nist.gov",
"type": "Primary"
}
]
},
"published": "2022-09-13T15:15:08.480",
"references": [
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"vulnStatus": "Modified",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-284"
}
],
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"description": [
{
"lang": "en",
"value": "NVD-CWE-noinfo"
}
],
"source": "nvd@nist.gov",
"type": "Primary"
}
]
}
FKIE_CVE-2022-38100
Vulnerability from fkie_nvd - Published: 2022-09-13 15:15 - Updated: 2024-11-21 07:15
Severity ?
7.5 (High) - CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
7.5 (High) - CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
7.5 (High) - CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Summary
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
References
| URL | Tags | ||
|---|---|---|---|
| ics-cert@hq.dhs.gov | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource | |
| af854a3a-2127-422b-91ae-364da2661108 | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource |
Impacted products
| Vendor | Product | Version | |
|---|---|---|---|
| contechealth | cms8000_firmware | - | |
| contechealth | cms8000 | - |
{
"configurations": [
{
"nodes": [
{
"cpeMatch": [
{
"criteria": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
"matchCriteriaId": "2C197D62-6F35-4B87-A721-BDB696EA240F",
"vulnerable": true
}
],
"negate": false,
"operator": "OR"
},
{
"cpeMatch": [
{
"criteria": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
"matchCriteriaId": "3A0CD9FA-68D7-4EEE-93A5-97275D84E2D3",
"vulnerable": false
}
],
"negate": false,
"operator": "OR"
}
],
"operator": "AND"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network."
},
{
"lang": "es",
"value": "El dispositivo CMS800 falla mientras intenta analizar los datos de red malformados enviados por un actor de la amenaza. Un actor de la amenaza con acceso a la red puede emitir de forma remota una petici\u00f3n UDP con un formato especial que causar\u00e1 el bloqueo de todo el dispositivo y requerir\u00e1 un reinicio f\u00edsico. Puede enviarse una petici\u00f3n de difusi\u00f3n UDP que cause un ataque de denegaci\u00f3n de servicio masivo en todos los dispositivos CME8000 conectados a la misma red"
}
],
"id": "CVE-2022-38100",
"lastModified": "2024-11-21T07:15:47.503",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"exploitabilityScore": 3.9,
"impactScore": 3.6,
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"exploitabilityScore": 3.9,
"impactScore": 3.6,
"source": "nvd@nist.gov",
"type": "Primary"
}
]
},
"published": "2022-09-13T15:15:08.843",
"references": [
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"vulnStatus": "Modified",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-400"
}
],
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"description": [
{
"lang": "en",
"value": "NVD-CWE-Other"
}
],
"source": "nvd@nist.gov",
"type": "Primary"
}
]
}
FKIE_CVE-2022-38453
Vulnerability from fkie_nvd - Published: 2022-09-13 15:15 - Updated: 2024-11-21 07:16
Severity ?
3.0 (Low) - CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N
4.4 (Medium) - CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:N
4.4 (Medium) - CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:N
Summary
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.
References
| URL | Tags | ||
|---|---|---|---|
| ics-cert@hq.dhs.gov | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource | |
| af854a3a-2127-422b-91ae-364da2661108 | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource |
Impacted products
| Vendor | Product | Version | |
|---|---|---|---|
| contechealth | cms8000_firmware | - | |
| contechealth | cms8000 | - |
{
"configurations": [
{
"nodes": [
{
"cpeMatch": [
{
"criteria": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
"matchCriteriaId": "2C197D62-6F35-4B87-A721-BDB696EA240F",
"vulnerable": true
}
],
"negate": false,
"operator": "OR"
},
{
"cpeMatch": [
{
"criteria": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
"matchCriteriaId": "3A0CD9FA-68D7-4EEE-93A5-97275D84E2D3",
"vulnerable": false
}
],
"negate": false,
"operator": "OR"
}
],
"operator": "AND"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "Multiple binary application files on the CMS8000 device are compiled with \u0027not stripped\u0027 and \u0027debug_info\u0027 compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities."
},
{
"lang": "es",
"value": "M\u00faltiples archivos de aplicaciones binarias en el dispositivo CMS8000 se compilan con las configuraciones de compilaci\u00f3n \"not stripped\" y \"debug_info\". Estas configuraciones de compilaci\u00f3n disminuyen en gran medida el nivel de esfuerzo para que un actor de la amenaza realice ingenier\u00eda inversa del c\u00f3digo confidencial e identifique vulnerabilidades adicionales"
}
],
"id": "CVE-2022-38453",
"lastModified": "2024-11-21T07:16:30.503",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "HIGH",
"attackVector": "LOCAL",
"availabilityImpact": "NONE",
"baseScore": 3.0,
"baseSeverity": "LOW",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "HIGH",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N",
"version": "3.1"
},
"exploitabilityScore": 0.5,
"impactScore": 2.5,
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "NONE",
"baseScore": 4.4,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "LOW",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:N",
"version": "3.1"
},
"exploitabilityScore": 1.8,
"impactScore": 2.5,
"source": "nvd@nist.gov",
"type": "Primary"
}
]
},
"published": "2022-09-13T15:15:08.907",
"references": [
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"vulnStatus": "Modified",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-489"
}
],
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"description": [
{
"lang": "en",
"value": "NVD-CWE-Other"
}
],
"source": "nvd@nist.gov",
"type": "Primary"
}
]
}
FKIE_CVE-2022-38069
Vulnerability from fkie_nvd - Published: 2022-09-13 15:15 - Updated: 2024-11-21 07:15
Severity ?
4.3 (Medium) - CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L
6.1 (Medium) - CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
6.1 (Medium) - CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Summary
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
References
| URL | Tags | ||
|---|---|---|---|
| ics-cert@hq.dhs.gov | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource | |
| af854a3a-2127-422b-91ae-364da2661108 | https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 | Mitigation, Third Party Advisory, US Government Resource |
Impacted products
| Vendor | Product | Version | |
|---|---|---|---|
| contechealth | cms8000_firmware | - | |
| contechealth | cms8000 | - |
{
"configurations": [
{
"nodes": [
{
"cpeMatch": [
{
"criteria": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
"matchCriteriaId": "2C197D62-6F35-4B87-A721-BDB696EA240F",
"vulnerable": true
}
],
"negate": false,
"operator": "OR"
},
{
"cpeMatch": [
{
"criteria": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
"matchCriteriaId": "3A0CD9FA-68D7-4EEE-93A5-97275D84E2D3",
"vulnerable": false
}
],
"negate": false,
"operator": "OR"
}
],
"operator": "AND"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters"
},
{
"lang": "es",
"value": "Se presentan m\u00faltiples credenciales predeterminadas globales en todos los dispositivos CMS8000 que, una vez expuestas, permiten a un actor de amenazas con acceso f\u00edsico moment\u00e1neo conseguir acceso privilegiado a cualquier dispositivo. El acceso con credenciales privilegiadas permite una extracci\u00f3n de informaci\u00f3n confidencial del paciente o la modificaci\u00f3n de los par\u00e1metros del dispositivo"
}
],
"id": "CVE-2022-38069",
"lastModified": "2024-11-21T07:15:43.047",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 4.3,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
"version": "3.1"
},
"exploitabilityScore": 0.9,
"impactScore": 3.4,
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "NONE",
"baseScore": 6.1,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
"version": "3.1"
},
"exploitabilityScore": 0.9,
"impactScore": 5.2,
"source": "nvd@nist.gov",
"type": "Primary"
}
]
},
"published": "2022-09-13T15:15:08.777",
"references": [
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"vulnStatus": "Modified",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-798"
}
],
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"description": [
{
"lang": "en",
"value": "CWE-798"
}
],
"source": "nvd@nist.gov",
"type": "Primary"
}
]
}
CVE-2022-38453 (GCVE-0-2022-38453)
Vulnerability from cvelistv5 – Published: 2022-09-13 14:55 – Updated: 2025-04-16 16:09
VLAI?
Summary
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.
Severity ?
CWE
- CWE-489 - Leftover Debug Code
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T10:54:03.773Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-38453",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T15:49:56.819949Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T16:09:53.917Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "Multiple binary application files on the CMS8000 device are compiled with \u0027not stripped\u0027 and \u0027debug_info\u0027 compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "HIGH",
"attackVector": "LOCAL",
"availabilityImpact": "NONE",
"baseScore": 3,
"baseSeverity": "LOW",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "HIGH",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-489",
"description": "CWE-489 Leftover Debug Code",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:55:03.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-38453",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Multiple binary application files on the CMS8000 device are compiled with \u0027not stripped\u0027 and \u0027debug_info\u0027 compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "HIGH",
"attackVector": "LOCAL",
"availabilityImpact": "NONE",
"baseScore": 3,
"baseSeverity": "LOW",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "HIGH",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-489 Leftover Debug Code"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-38453",
"datePublished": "2022-09-13T14:55:03.842Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T16:09:53.917Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-3027 (GCVE-0-2022-3027)
Vulnerability from cvelistv5 – Published: 2022-09-13 14:54 – Updated: 2025-04-16 16:10
VLAI?
Summary
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
Severity ?
5.7 (Medium)
CWE
- CWE-284 - Improper Access Control
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T00:53:00.651Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-3027",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T15:50:01.916127Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T16:10:02.402Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 5.7,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-284",
"description": "CWE-284 Improper Access Control",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:54:58.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-3027",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 5.7,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-284 Improper Access Control"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-3027",
"datePublished": "2022-09-13T14:54:58.631Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T16:10:02.402Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-38069 (GCVE-0-2022-38069)
Vulnerability from cvelistv5 – Published: 2022-09-13 14:54 – Updated: 2025-04-16 16:10
VLAI?
Summary
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
Severity ?
4.3 (Medium)
CWE
- CWE-798 - Use of Hard-coded Credentials
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T10:45:52.456Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-38069",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T15:50:05.918933Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T16:10:11.963Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters"
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 4.3,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798 Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:54:54.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-38069",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters"
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 4.3,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-798 Use of Hard-coded Credentials"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-38069",
"datePublished": "2022-09-13T14:54:54.113Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T16:10:11.963Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-38100 (GCVE-0-2022-38100)
Vulnerability from cvelistv5 – Published: 2022-09-13 14:54 – Updated: 2025-04-16 16:10
VLAI?
Summary
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Severity ?
7.5 (High)
CWE
- CWE-400 - Uncontrolled Resource Consumption
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T10:45:52.412Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-38100",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T15:54:20.725304Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T16:10:19.613Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-400",
"description": "CWE-400 Uncontrolled Resource Consumption",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:54:50.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-38100",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-400 Uncontrolled Resource Consumption"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-38100",
"datePublished": "2022-09-13T14:54:50.356Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T16:10:19.613Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-36385 (GCVE-0-2022-36385)
Vulnerability from cvelistv5 – Published: 2022-09-13 14:54 – Updated: 2025-04-16 17:47
VLAI?
Summary
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
Severity ?
6.8 (Medium)
CWE
- CWE-284 - Improper Access Control
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T10:00:04.453Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-36385",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T17:27:16.252781Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T17:47:37.455Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-284",
"description": "CWE-284 Improper Access Control",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:54:35.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-36385",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-284 Improper Access Control"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-36385",
"datePublished": "2022-09-13T14:54:35.692Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T17:47:37.455Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-38453 (GCVE-0-2022-38453)
Vulnerability from nvd – Published: 2022-09-13 14:55 – Updated: 2025-04-16 16:09
VLAI?
Summary
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.
Severity ?
CWE
- CWE-489 - Leftover Debug Code
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T10:54:03.773Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-38453",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T15:49:56.819949Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T16:09:53.917Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "Multiple binary application files on the CMS8000 device are compiled with \u0027not stripped\u0027 and \u0027debug_info\u0027 compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "HIGH",
"attackVector": "LOCAL",
"availabilityImpact": "NONE",
"baseScore": 3,
"baseSeverity": "LOW",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "HIGH",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-489",
"description": "CWE-489 Leftover Debug Code",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:55:03.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-38453",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Multiple binary application files on the CMS8000 device are compiled with \u0027not stripped\u0027 and \u0027debug_info\u0027 compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "HIGH",
"attackVector": "LOCAL",
"availabilityImpact": "NONE",
"baseScore": 3,
"baseSeverity": "LOW",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "HIGH",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-489 Leftover Debug Code"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-38453",
"datePublished": "2022-09-13T14:55:03.842Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T16:09:53.917Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-3027 (GCVE-0-2022-3027)
Vulnerability from nvd – Published: 2022-09-13 14:54 – Updated: 2025-04-16 16:10
VLAI?
Summary
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
Severity ?
5.7 (Medium)
CWE
- CWE-284 - Improper Access Control
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T00:53:00.651Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-3027",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T15:50:01.916127Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T16:10:02.402Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 5.7,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-284",
"description": "CWE-284 Improper Access Control",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:54:58.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-3027",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 5.7,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-284 Improper Access Control"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-3027",
"datePublished": "2022-09-13T14:54:58.631Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T16:10:02.402Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-38069 (GCVE-0-2022-38069)
Vulnerability from nvd – Published: 2022-09-13 14:54 – Updated: 2025-04-16 16:10
VLAI?
Summary
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
Severity ?
4.3 (Medium)
CWE
- CWE-798 - Use of Hard-coded Credentials
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T10:45:52.456Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-38069",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T15:50:05.918933Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T16:10:11.963Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters"
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 4.3,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798 Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:54:54.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-38069",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters"
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 4.3,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-798 Use of Hard-coded Credentials"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-38069",
"datePublished": "2022-09-13T14:54:54.113Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T16:10:11.963Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-38100 (GCVE-0-2022-38100)
Vulnerability from nvd – Published: 2022-09-13 14:54 – Updated: 2025-04-16 16:10
VLAI?
Summary
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Severity ?
7.5 (High)
CWE
- CWE-400 - Uncontrolled Resource Consumption
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T10:45:52.412Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-38100",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T15:54:20.725304Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T16:10:19.613Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-400",
"description": "CWE-400 Uncontrolled Resource Consumption",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:54:50.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-38100",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-400 Uncontrolled Resource Consumption"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-38100",
"datePublished": "2022-09-13T14:54:50.356Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T16:10:19.613Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2022-36385 (GCVE-0-2022-36385)
Vulnerability from nvd – Published: 2022-09-13 14:54 – Updated: 2025-04-16 17:47
VLAI?
Summary
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
Severity ?
6.8 (Medium)
CWE
- CWE-284 - Improper Access Control
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
Affected:
All
|
Credits
Level Nine reported these vulnerabilities to CISA.
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-03T10:00:04.453Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2022-36385",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-04-16T17:27:16.252781Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-04-16T17:47:37.455Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"datePublic": "2022-09-01T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
}
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-284",
"description": "CWE-284 Improper Access Control",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2022-09-13T14:54:35.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
],
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"title": "Contec Health CMS8000",
"workarounds": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
],
"x_generator": {
"engine": "Vulnogram 0.0.9"
},
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-36385",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-284 Improper Access Control"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "en",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2022-36385",
"datePublished": "2022-09-13T14:54:35.692Z",
"dateReserved": "2022-08-29T00:00:00.000Z",
"dateUpdated": "2025-04-16T17:47:37.455Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}