CVE-2024-48971
Vulnerability from cvelistv5
Published
2024-11-14 21:13
Modified
2024-11-15 21:06
Severity ?
EPSS score ?
Summary
Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
References
Impacted products
▼ | Vendor | Product |
---|---|---|
Baxter | Life2000 Ventilation System |
{ "containers": { "adp": [ { "affected": [ { "cpes": [ "cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*" ], "defaultStatus": "unaffected", "product": "life2000_ventilator_firmware", "vendor": "baxter", "versions": [ { "lessThanOrEqual": "06.08.00.00", "status": "affected", "version": "0", "versionType": "custom" } ] } ], "metrics": [ { "other": { "content": { "id": "CVE-2024-48971", "options": [ { "Exploitation": "none" }, { "Automatable": "no" }, { "Technical Impact": "total" } ], "role": "CISA Coordinator", "timestamp": "2024-11-15T21:05:32.910003Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2024-11-15T21:06:24.325Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "Life2000 Ventilation System", "vendor": "Baxter", "versions": [ { "status": "affected", "version": "06.08.00.00 and prior" } ] } ], "datePublic": "2024-11-14T21:47:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.\u003cbr\u003e" } ], "value": "The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges." } ], "impacts": [ { "capecId": "CAPEC-191", "descriptions": [ { "lang": "en", "value": "CAPEC-191 Read Sensitive Constants Within an Executable" } ] } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "LOCAL", "availabilityImpact": "HIGH", "baseScore": 9.3, "baseSeverity": "CRITICAL", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-798", "description": "CWE-798 Use of Hard-coded Credentials", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2024-11-14T21:49:12.240Z", "orgId": "dba971b9-eb30-4121-91e1-3b45611354aa", "shortName": "Baxter" }, "references": [ { "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01" } ], "source": { "discovery": "UNKNOWN" }, "title": "Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator", "x_generator": { "engine": "Vulnogram 0.2.0" } } }, "cveMetadata": { "assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa", "assignerShortName": "Baxter", "cveId": "CVE-2024-48971", "datePublished": "2024-11-14T21:13:36.036Z", "dateReserved": "2024-10-10T19:24:41.495Z", "dateUpdated": "2024-11-15T21:06:24.325Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1", "meta": { "nvd": "{\"cve\":{\"id\":\"CVE-2024-48971\",\"sourceIdentifier\":\"productsecurity@baxter.com\",\"published\":\"2024-11-14T22:15:18.327\",\"lastModified\":\"2024-11-15T13:58:08.913\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.\"},{\"lang\":\"es\",\"value\":\"La contrase\u00f1a del m\u00e9dico y el n\u00famero de serie est\u00e1n codificados en el respirador en formato de texto simple. Esto podr\u00eda permitir que un atacante obtenga la contrase\u00f1a del respirador y la use para obtener acceso no autorizado al dispositivo con privilegios de m\u00e9dico.\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\",\"attackVector\":\"LOCAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\",\"baseScore\":9.3,\"baseSeverity\":\"CRITICAL\"},\"exploitabilityScore\":2.5,\"impactScore\":6.0}]},\"weaknesses\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-798\"}]}],\"references\":[{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\",\"source\":\"productsecurity@baxter.com\"}]}}" } }
Loading...
Loading...
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.