cve-2020-16228
Vulnerability from cvelistv5
Published
2020-09-11 12:58
Modified
2024-08-04 13:37
Severity ?
EPSS score ?
Summary
Philips Patient Monitoring Devices Improper Check for Certificate Revocation
References
▼ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 | Third Party Advisory, US Government Resource | |
ics-cert@hq.dhs.gov | https://www.philips.com/productsecurity |
Impacted products
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T13:37:54.179Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01" }, { "tags": [ "x_transferred" ], "url": "https://www.philips.com/productsecurity" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "Patient Information Center iX (PICiX)", "vendor": "Philips ", "versions": [ { "status": "affected", "version": "C.02" }, { "status": "affected", "version": "C.03" } ] }, { "defaultStatus": "unaffected", "product": "PerformanceBridge Focal Point", "vendor": "Philips ", "versions": [ { "status": "affected", "version": "A.01" } ] }, { "defaultStatus": "unaffected", "product": "IntelliVue patient monitors", "vendor": "Philips ", "versions": [ { "status": "affected", "version": "MX100" }, { "status": "affected", "version": "MX400-MX550" }, { "status": "affected", "version": "MX750" }, { "status": "affected", "version": "MX850" } ] }, { "defaultStatus": "unaffected", "product": "IntelliVue X3", "vendor": "Philips ", "versions": [ { "lessThanOrEqual": "N ", "status": "affected", "version": "0", "versionType": "custom" } ] } ], "credits": [ { "lang": "en", "type": "finder", "user": "00000000-0000-4000-9000-000000000000", "value": "Julian Suleder, Nils Emmerich, Birk Kauer of ERNW Research GmbH, Dr. Oliver Matula of ERNW Enno, and Rey Netzwerke GmbH reported these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices), which reported these to Philips." } ], "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003e\nIn Patient Information Center iX (PICiX) Versions C.02 and C.03, \nPerformanceBridge Focal Point Version A.01, IntelliVue patient monitors \nMX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and \nprior, the software does not check or incorrectly checks the revocation \nstatus of a certificate, which may cause it to use a compromised \ncertificate.\n\n\u003c/p\u003e" } ], "value": "In Patient Information Center iX (PICiX) Versions C.02 and C.03, \nPerformanceBridge Focal Point Version A.01, IntelliVue patient monitors \nMX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and \nprior, the software does not check or incorrectly checks the revocation \nstatus of a certificate, which may cause it to use a compromised \ncertificate.\n\n\n\n" } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-299", "description": "CWE-299 Improper Check for Certificate Revocation", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2023-12-12T20:52:52.755Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01" }, { "url": "https://www.philips.com/productsecurity" } ], "solutions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\n\u003cp\u003ePhilips released the following versions to remediate reported vulnerabilities:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003ePatient Information Center iX (PICiX) Version C.03\u003c/li\u003e\n\u003cli\u003ePerformanceBridge Focal Point\u003c/li\u003e\n\u003cli\u003eIntelliVue Patient Monitors Versions N.00 and N.01\u003c/li\u003e\n\u003cli\u003eIntelliVue Patient Monitors Version M.04: Contact a \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.usa.philips.com/healthcare/solutions/customer-service-solutions\"\u003ePhilips service support team\u003c/a\u003e for an upgrade path\u003c/li\u003e\n\u003cli\u003eCertificate revocation within the system was implemented for PIC iX \nand Performance Bridge FocalPoint in 2023. The implementation of the \nIntelliVue Patient Monitors will be completed in Q3 of 2024.\u003c/li\u003e\n\u003c/ul\u003e\n\n\u003cbr\u003e" } ], "value": "Philips released the following versions to remediate reported vulnerabilities:\n\n\n\n * Patient Information Center iX (PICiX) Version C.03\n\n * PerformanceBridge Focal Point\n\n * IntelliVue Patient Monitors Versions N.00 and N.01\n\n * IntelliVue Patient Monitors Version M.04: Contact a Philips service support team https://www.usa.philips.com/healthcare/solutions/customer-service-solutions for an upgrade path\n\n * Certificate revocation within the system was implemented for PIC iX \nand Performance Bridge FocalPoint in 2023. The implementation of the \nIntelliVue Patient Monitors will be completed in Q3 of 2024.\n\n\n\n\n\n\n" } ], "source": { "discovery": "EXTERNAL" }, "title": "Philips Patient Monitoring Devices Improper Check for Certificate Revocation", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\n\u003cp\u003eAs a mitigation to these vulnerabilities, Philips recommends the following:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003eThe Philips patient monitoring network is required to be physically \nor logically isolated from the hospital local area network (LAN). \nPhilips recommends using a firewall or routers that can implement access\n control lists restricting access in and out of the patient monitoring \nnetwork for only necessary ports and IP addresses. Refer to the Philips \nPatient Monitoring System Security for Clinical Networks guide for \nadditional information on \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://incenter.medical.philips.com/\"\u003eInCenter\u003c/a\u003e.\u003c/li\u003e\n\u003cli\u003eBy default, the simple certificate enrollment protocol (SCEP) \nservice is not running. When needed, the service is configured to run \nbased on the duration or the number of certificates to be assigned. One \ncertificate is default, but if a certificate is not issued, the service \nwill continue to run. Limit exposure by ensuring the SCEP service is not\n running unless it is actively being used to enroll new devices.\u003c/li\u003e\n\u003cli\u003eWhen enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits.\u003c/li\u003e\n\u003cli\u003eImplement physical security controls to prevent unauthorized login \nattempts on the PIC iX application. Servers should be kept in controlled\n locked data centers. Access to equipment at nurses\u2019 stations should be \ncontrolled and monitored.\u003c/li\u003e\n\u003cli\u003eOnly grant remote access to PIC iX servers on a must-have basis.\u003c/li\u003e\n\u003cli\u003eGrant login privileges to the bedside monitor and PIC iX application\n on a role-based, least-privilege basis, and only to trusted users.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eUsers with questions regarding their specific Philips Patient \nInformation Center (PIC iX) and/or IntelliVue patient monitor \ninstallations and new release eligibility should contact their local \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.usa.philips.com/healthcare/solutions/customer-service-solutions\"\u003ePhilips service support team, or regional service support\u003c/a\u003e, or call 1-800-722-9377.\u003c/p\u003e\n\u003cp\u003ePlease see the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/productsecurity\"\u003ePhilips product security website\u003c/a\u003e for the Philips advisory and the latest security information for Philips products.\u003c/p\u003e\n\n\u003cbr\u003e" } ], "value": "As a mitigation to these vulnerabilities, Philips recommends the following:\n\n\n\n * The Philips patient monitoring network is required to be physically \nor logically isolated from the hospital local area network (LAN). \nPhilips recommends using a firewall or routers that can implement access\n control lists restricting access in and out of the patient monitoring \nnetwork for only necessary ports and IP addresses. Refer to the Philips \nPatient Monitoring System Security for Clinical Networks guide for \nadditional information on InCenter https://incenter.medical.philips.com/ .\n\n * By default, the simple certificate enrollment protocol (SCEP) \nservice is not running. When needed, the service is configured to run \nbased on the duration or the number of certificates to be assigned. One \ncertificate is default, but if a certificate is not issued, the service \nwill continue to run. Limit exposure by ensuring the SCEP service is not\n running unless it is actively being used to enroll new devices.\n\n * When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits.\n\n * Implement physical security controls to prevent unauthorized login \nattempts on the PIC iX application. Servers should be kept in controlled\n locked data centers. Access to equipment at nurses\u2019 stations should be \ncontrolled and monitored.\n\n * Only grant remote access to PIC iX servers on a must-have basis.\n\n * Grant login privileges to the bedside monitor and PIC iX application\n on a role-based, least-privilege basis, and only to trusted users.\n\n\n\n\nUsers with questions regarding their specific Philips Patient \nInformation Center (PIC iX) and/or IntelliVue patient monitor \ninstallations and new release eligibility should contact their local Philips service support team, or regional service support https://www.usa.philips.com/healthcare/solutions/customer-service-solutions , or call 1-800-722-9377.\n\n\nPlease see the Philips product security website https://www.philips.com/productsecurity for the Philips advisory and the latest security information for Philips products.\n\n\n\n\n" } ], "x_generator": { "engine": "Vulnogram 0.1.0-dev" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-16228", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Philips Patient Information Center iX (PICiX), PerformanceBridge Focal Point, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90, IntelliVue X3 and X2.", "version": { "version_data": [ { "version_value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior." } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. The software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "IMPROPER CHECK FOR CERTIFICATE REVOCATION CWE-299" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-16228", "datePublished": "2020-09-11T12:58:13", "dateReserved": "2020-07-31T00:00:00", "dateUpdated": "2024-08-04T13:37:54.179Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1", "meta": { "nvd": "{\"cve\":{\"id\":\"CVE-2020-16228\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2020-09-11T13:15:11.377\",\"lastModified\":\"2023-12-12T21:15:07.970\",\"vulnStatus\":\"Modified\",\"descriptions\":[{\"lang\":\"en\",\"value\":\"In Patient Information Center iX (PICiX) Versions C.02 and C.03, \\nPerformanceBridge Focal Point Version A.01, IntelliVue patient monitors \\nMX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and \\nprior, the software does not check or incorrectly checks the revocation \\nstatus of a certificate, which may cause it to use a compromised \\ncertificate.\\n\\n\\n\\n\"},{\"lang\":\"es\",\"value\":\"Patient Information Center iX (PICiX) Versiones B.02, C.02, C.03, PerformanceBridge Focal Point Versi\u00f3n A.01, Monitores de paciente IntelliVue MX100, MX400-MX850 y MP2-MP90 Versiones N y anteriores, IntelliVue X3 y X2 Versiones N y anteriores.\u0026#xa0;El software no comprueba o comprueba inapropiadamente el estado de revocaci\u00f3n de un certificado, lo que puede causar que use un certificado comprometido\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:A/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:L\",\"attackVector\":\"ADJACENT_NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"HIGH\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"LOW\",\"baseScore\":6.4,\"baseSeverity\":\"MEDIUM\"},\"exploitabilityScore\":0.9,\"impactScore\":5.5}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:A/AC:L/Au:S/C:P/I:P/A:P\",\"accessVector\":\"ADJACENT_NETWORK\",\"accessComplexity\":\"LOW\",\"authentication\":\"SINGLE\",\"confidentialityImpact\":\"PARTIAL\",\"integrityImpact\":\"PARTIAL\",\"availabilityImpact\":\"PARTIAL\",\"baseScore\":5.2},\"baseSeverity\":\"MEDIUM\",\"exploitabilityScore\":5.1,\"impactScore\":6.4,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":false}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-299\"}]}],\"configurations\":[{\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:a:philips:patient_information_center_ix:b.02:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"C4FA1FEC-5139-48C1-856C-8062436AE6C6\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:a:philips:patient_information_center_ix:c.02:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F3220AB2-AC0D-4AC2-90D8-76C02FC693EB\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:a:philips:patient_information_center_ix:c.03:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"AE1FB3E9-E269-434A-B1C2-D54C40F437BE\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:a:philips:performancebridge_focal_point:a.01:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"3B4CF59B-32DC-4F48-88C5-77B96E937E93\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mp2-mp90_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"4BDDAC8D-53B7-4B52-8EE2-510E6FE215D8\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mp2-mp90:n:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"A9B11FBA-4FCF-4B0A-ADA9-6BB59686A8DE\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mx100_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"8D7C4F77-6005-4AAA-83CA-CA9F60043A7A\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mx100:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"0D2A5BBF-C360-4281-AD34-C0941831DA64\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mx400_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"58F21718-A0DE-419E-B82D-447ADC337E30\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mx400:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"27B8D0E2-316A-45E5-8FC1-AE70F07DB7FA\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mx850_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"77650C8F-73D8-4DBC-8673-40502748C3CE\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mx850:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"A1F44D95-5744-4475-9312-63E1F422B236\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_x2_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"E42604C2-7CB6-42AC-9272-CFABFC4DC85A\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_x2:n:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"DA10D49B-DBB0-4E4D-B0C9-17F91D22E5AB\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_x3_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"708EB7F4-9747-4B0C-937D-F1ED07300FAE\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_x3:n:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"221DDFB6-3F7D-4611-9A87-2D216DF79A2B\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mx800_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"E9C6A5F3-C845-4561-A451-DC3C966E8D8C\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mx800:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"C38B732B-8629-40FE-A333-E69A7F2CFB18\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mx750_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"B60F7F6F-995F-4251-9DF8-897F42025B2C\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mx750:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F77F9328-BFE3-42C9-A487-6295B05274ED\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mx700_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"43C2466F-A371-4386-B098-F724D5CD14CB\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mx700:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"DDF21FA3-9650-442F-8E13-281A9975C47B\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mx600_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"B4394A2A-C9F4-4E56-866E-EB49F6B32C3D\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mx600:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"186C49C9-BDA8-4083-A3E9-606BBA027CAB\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:intellivue_mx550_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"6D72DF93-DF02-411D-80BE-85286F553B1E\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:philips:intellivue_mx550:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"8B03E7D9-01DC-492F-BA52-9165E78BE498\"}]}]}],\"references\":[{\"url\":\"https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"https://www.philips.com/productsecurity\",\"source\":\"ics-cert@hq.dhs.gov\"}]}}" } }
Loading...
Loading...
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.