CWE-920
Improper Restriction of Power Consumption
The product operates in an environment in which power is a limited resource that cannot be automatically replenished, but the product does not properly restrict the amount of power that its operation consumes.
CVE-2017-12714 (GCVE-0-2017-12714)
Vulnerability from cvelistv5 – Published: 2018-04-25 13:00 – Updated: 2024-09-17 00:16
VLAI?
Summary
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Severity ?
No CVSS data available.
CWE
- CWE-920 - Improper Restriction of power consumption CWE-920
Assigner
References
| URL | Tags | |||||||
|---|---|---|---|---|---|---|---|---|
|
||||||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Abbott Laboratories | Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI. |
Affected:
All versions of pacemakers manufactured prior to August 28, 2017
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T18:43:56.537Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"tags": [
"vdb-entry",
"x_refsource_BID",
"x_transferred"
],
"url": "http://www.securityfocus.com/bid/100523"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
"vendor": "Abbott Laboratories",
"versions": [
{
"status": "affected",
"version": "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
],
"datePublic": "2017-08-29T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted \"RF wake-up\" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-920",
"description": "Improper Restriction of power consumption CWE-920",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-04-26T09:57:01",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"tags": [
"vdb-entry",
"x_refsource_BID"
],
"url": "http://www.securityfocus.com/bid/100523"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2017-08-29T00:00:00",
"ID": "CVE-2017-12714",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
"version": {
"version_data": [
{
"version_value": "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
}
]
},
"vendor_name": "Abbott Laboratories"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted \"RF wake-up\" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Improper Restriction of power consumption CWE-920"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"refsource": "BID",
"url": "http://www.securityfocus.com/bid/100523"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2017-12714",
"datePublished": "2018-04-25T13:00:00Z",
"dateReserved": "2017-08-09T00:00:00",
"dateUpdated": "2024-09-17T00:16:50.163Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
No mitigation information available for this CWE.
No CAPEC attack patterns related to this CWE.