Search criteria
3 vulnerabilities by Abbott Laboratories
CVE-2017-12716 (GCVE-0-2017-12716)
Vulnerability from cvelistv5 – Published: 2018-04-25 13:00 – Updated: 2024-09-17 01:55
VLAI?
Summary
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Severity ?
No CVSS data available.
CWE
- CWE-311 - Missing encryption of sensitive data CWE-311
Assigner
References
| URL | Tags | |||||||
|---|---|---|---|---|---|---|---|---|
|
||||||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Abbott Laboratories | Accent and Anthem |
Affected:
All versions of pacemakers manufactured prior to August 28, 2017
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T18:43:56.589Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"tags": [
"vdb-entry",
"x_refsource_BID",
"x_transferred"
],
"url": "http://www.securityfocus.com/bid/100523"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "Accent and Anthem",
"vendor": "Abbott Laboratories",
"versions": [
{
"status": "affected",
"version": "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
],
"datePublic": "2017-08-29T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-311",
"description": "Missing encryption of sensitive data CWE-311",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-04-26T09:57:01",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"tags": [
"vdb-entry",
"x_refsource_BID"
],
"url": "http://www.securityfocus.com/bid/100523"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2017-08-29T00:00:00",
"ID": "CVE-2017-12716",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "Accent and Anthem",
"version": {
"version_data": [
{
"version_value": "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
}
]
},
"vendor_name": "Abbott Laboratories"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Missing encryption of sensitive data CWE-311"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"refsource": "BID",
"url": "http://www.securityfocus.com/bid/100523"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2017-12716",
"datePublished": "2018-04-25T13:00:00Z",
"dateReserved": "2017-08-09T00:00:00",
"dateUpdated": "2024-09-17T01:55:36.387Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2017-12714 (GCVE-0-2017-12714)
Vulnerability from cvelistv5 – Published: 2018-04-25 13:00 – Updated: 2024-09-17 00:16
VLAI?
Summary
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Severity ?
No CVSS data available.
CWE
- CWE-920 - Improper Restriction of power consumption CWE-920
Assigner
References
| URL | Tags | |||||||
|---|---|---|---|---|---|---|---|---|
|
||||||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Abbott Laboratories | Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI. |
Affected:
All versions of pacemakers manufactured prior to August 28, 2017
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T18:43:56.537Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"tags": [
"vdb-entry",
"x_refsource_BID",
"x_transferred"
],
"url": "http://www.securityfocus.com/bid/100523"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
"vendor": "Abbott Laboratories",
"versions": [
{
"status": "affected",
"version": "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
],
"datePublic": "2017-08-29T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted \"RF wake-up\" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-920",
"description": "Improper Restriction of power consumption CWE-920",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-04-26T09:57:01",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"tags": [
"vdb-entry",
"x_refsource_BID"
],
"url": "http://www.securityfocus.com/bid/100523"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2017-08-29T00:00:00",
"ID": "CVE-2017-12714",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
"version": {
"version_data": [
{
"version_value": "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
}
]
},
"vendor_name": "Abbott Laboratories"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted \"RF wake-up\" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Improper Restriction of power consumption CWE-920"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"refsource": "BID",
"url": "http://www.securityfocus.com/bid/100523"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2017-12714",
"datePublished": "2018-04-25T13:00:00Z",
"dateReserved": "2017-08-09T00:00:00",
"dateUpdated": "2024-09-17T00:16:50.163Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2017-12712 (GCVE-0-2017-12712)
Vulnerability from cvelistv5 – Published: 2018-04-25 13:00 – Updated: 2024-09-17 03:48
VLAI?
Summary
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Severity ?
No CVSS data available.
CWE
- CWE-287 - Improper authentication CWE-287
Assigner
References
| URL | Tags | |||||||
|---|---|---|---|---|---|---|---|---|
|
||||||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Abbott Laboratories | Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI. |
Affected:
All versions of pacemakers manufactured prior to August 28, 2017
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T18:43:56.613Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"tags": [
"vdb-entry",
"x_refsource_BID",
"x_transferred"
],
"url": "http://www.securityfocus.com/bid/100523"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
"vendor": "Abbott Laboratories",
"versions": [
{
"status": "affected",
"version": "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
],
"datePublic": "2017-08-29T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-287",
"description": "Improper authentication CWE-287",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-04-26T09:57:01",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"tags": [
"vdb-entry",
"x_refsource_BID"
],
"url": "http://www.securityfocus.com/bid/100523"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2017-08-29T00:00:00",
"ID": "CVE-2017-12712",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
"version": {
"version_data": [
{
"version_value": "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
}
]
},
"vendor_name": "Abbott Laboratories"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Improper authentication CWE-287"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name": "100523",
"refsource": "BID",
"url": "http://www.securityfocus.com/bid/100523"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2017-12712",
"datePublished": "2018-04-25T13:00:00Z",
"dateReserved": "2017-08-09T00:00:00",
"dateUpdated": "2024-09-17T03:48:37.996Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}