CVE-2024-1275
Vulnerability from cvelistv5
Published
2024-05-31 17:23
Modified
2024-08-01 18:33
Summary
Vulnerability in Baxter Welch Allyn Connex Spot Monitor
Impacted products
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "affected": [
          {
            "cpes": [
              "cpe:2.3:o:baxter:welch_allyn_connex_spot_monitor:*:*:*:*:*:*:*:*"
            ],
            "defaultStatus": "unaffected",
            "product": "welch_allyn_connex_spot_monitor",
            "vendor": "baxter",
            "versions": [
              {
                "lessThanOrEqual": "1.52",
                "status": "affected",
                "version": "0",
                "versionType": "custom"
              }
            ]
          }
        ],
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2024-1275",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2024-06-05T14:39:19.332683Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2024-06-05T15:01:00.230Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      },
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-01T18:33:25.355Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_transferred"
            ],
            "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-151-02"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Welch Allyn Connex Spot Monitor",
          "vendor": "Baxter",
          "versions": [
            {
              "lessThanOrEqual": "1.52",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "reporter",
          "value": "Maarten Boone and Edwin Van Andel (CTO of Zerocopter) reported this vulnerability to Baxter."
        }
      ],
      "datePublic": "2024-05-30T17:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.\u003cp\u003eThis issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.\u003c/p\u003e"
            }
          ],
          "value": "Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52."
        }
      ],
      "impacts": [
        {
          "capecId": "CAPEC-176",
          "descriptions": [
            {
              "lang": "en",
              "value": "CAPEC-176 Configuration/Environment Manipulation"
            }
          ]
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "HIGH",
            "attackRequirements": "NONE",
            "attackVector": "NETWORK",
            "baseScore": 9.1,
            "baseSeverity": "CRITICAL",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "NONE",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "HIGH",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-1394",
              "description": "CWE-1394 Use of Default Cryptographic Key",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2024-06-05T14:42:55.386Z",
        "orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
        "shortName": "Baxter"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-151-02"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eBaxter has released a software update for all impacted devices and software to address this vulnerability. A new version of the product that mitigates the vulnerability is available as follows:\u003c/p\u003e\u003cul\u003e\u003cli\u003eWelch Allyn Connex Spot Monitor: Version 1.52.01 (available October 16, 2023)\u003c/li\u003e\u003c/ul\u003e\u003cp\u003eBaxter recommends users upgrade to the latest versions of their products. Information on how to update products to their new versions can be found on the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.baxter.com/product-security\"\u003eBaxter disclosure page\u003c/a\u003e\u0026nbsp;or the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.hillrom.com/en/responsible-disclosures/\"\u003eHillrom disclosure page\u003c/a\u003e.\u003c/p\u003e\u003cp\u003eBaxter recommends the following workarounds to help reduce risk:\u003c/p\u003e\u003cul\u003e\u003cli\u003eApply proper network and physical security controls.\u003c/li\u003e\u003cli\u003eEnsure a unique encryption key is configured and applied to the product (as described in the Connex Spot Monitor Service Manual).\u003c/li\u003e\u003c/ul\u003e"
            }
          ],
          "value": "Baxter has released a software update for all impacted devices and software to address this vulnerability. A new version of the product that mitigates the vulnerability is available as follows:\n\n  *  Welch Allyn Connex Spot Monitor: Version 1.52.01 (available October 16, 2023)\n\n\nBaxter recommends users upgrade to the latest versions of their products. Information on how to update products to their new versions can be found on the  Baxter disclosure page https://www.baxter.com/product-security \u00a0or the  Hillrom disclosure page https://www.hillrom.com/en/responsible-disclosures/ .\n\nBaxter recommends the following workarounds to help reduce risk:\n\n  *  Apply proper network and physical security controls.\n  *  Ensure a unique encryption key is configured and applied to the product (as described in the Connex Spot Monitor Service Manual)."
        }
      ],
      "source": {
        "advisory": "ICSMA-24-151-02",
        "discovery": "UNKNOWN"
      },
      "title": "Vulnerability in Baxter Welch Allyn Connex Spot Monitor",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
    "assignerShortName": "Baxter",
    "cveId": "CVE-2024-1275",
    "datePublished": "2024-05-31T17:23:19.207Z",
    "dateReserved": "2024-02-06T14:20:33.446Z",
    "dateUpdated": "2024-08-01T18:33:25.355Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "meta": {
    "nvd": "{\"cve\":{\"id\":\"CVE-2024-1275\",\"sourceIdentifier\":\"productsecurity@baxter.com\",\"published\":\"2024-05-31T18:15:10.140\",\"lastModified\":\"2024-06-05T15:15:10.990\",\"vulnStatus\":\"Awaiting Analysis\",\"descriptions\":[{\"lang\":\"en\",\"value\":\"Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.\"},{\"lang\":\"es\",\"value\":\"El uso de la vulnerabilidad de clave criptogr\u00e1fica predeterminada en Baxter Welch Ally Connex Spot Monitor puede permitir la manipulaci\u00f3n de la configuraci\u00f3n/entorno. Este problema afecta a Welch Ally Connex Spot Monitor en todas las versiones anteriores a la 1.52.\"}],\"metrics\":{},\"weaknesses\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-1394\"}]}],\"references\":[{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-151-02\",\"source\":\"productsecurity@baxter.com\"}]}}"
  }
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading...

Loading...

Loading...

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.