OS command injection vulnerabilities in GE HealthCare ultrasound devices
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "affected": [
          {
            "cpes": [
              "cpe:2.3:a:ge_healthcare:venue:*:*:*:*:*:*:*:*"
            ],
            "defaultStatus": "unknown",
            "product": "venue",
            "vendor": "ge_healthcare",
            "versions": [
              {
                "status": "affected",
                "version": "multiple versions"
              }
            ]
          },
          {
            "cpes": [
              "cpe:2.3:a:ge_healthcare:vivid:*:*:*:*:*:*:*:*"
            ],
            "defaultStatus": "unknown",
            "product": "vivid",
            "vendor": "ge_healthcare",
            "versions": [
              {
                "status": "affected",
                "version": "multiple versions"
              }
            ]
          },
          {
            "cpes": [
              "cpe:2.3:a:ge_healthcare:voluson:*:*:*:*:*:*:*:*"
            ],
            "defaultStatus": "unknown",
            "product": "voluson",
            "vendor": "ge_healthcare",
            "versions": [
              {
                "status": "affected",
                "version": "multiple versions"
              }
            ]
          },
          {
            "cpes": [
              "cpe:2.3:a:ge_healthcare:logiq:*:*:*:*:*:*:*:*"
            ],
            "defaultStatus": "unknown",
            "product": "logiq",
            "vendor": "ge_healthcare",
            "versions": [
              {
                "status": "affected",
                "version": "multiple versions"
              }
            ]
          }
        ],
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2024-1628",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2024-05-16T16:10:00.517201Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2024-06-04T18:01:30.939Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      },
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-01T18:48:21.580Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_transferred"
            ],
            "url": "https://securityupdate.gehealthcare.com/"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Venue",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "status": "affected",
              "version": "R1"
            },
            {
              "status": "affected",
              "version": "R2"
            },
            {
              "lessThanOrEqual": "R3.3",
              "status": "affected",
              "version": "R3",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "R4.3",
              "status": "affected",
              "version": "R4",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Venue Go",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "status": "affected",
              "version": "R2"
            },
            {
              "lessThanOrEqual": "R3.3",
              "status": "affected",
              "version": "R3",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "R4.3",
              "status": "affected",
              "version": "R4",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Venue Fit",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThanOrEqual": "R3.3",
              "status": "affected",
              "version": "R3",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "R4.3",
              "status": "affected",
              "version": "R4",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "LOGIQ e",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThanOrEqual": "R9.1.4",
              "status": "affected",
              "version": "R7",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "R10.1.3",
              "status": "affected",
              "version": "R8",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "R11.0.3",
              "status": "affected",
              "version": "R9",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "LOGIQ He",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThanOrEqual": "R9.3.1",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Vivid E",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThan": "206",
              "status": "affected",
              "version": "E95",
              "versionType": "custom"
            },
            {
              "lessThan": "206",
              "status": "affected",
              "version": "E90",
              "versionType": "custom"
            },
            {
              "lessThan": "206",
              "status": "affected",
              "version": "E80",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Vivid S",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThan": "206",
              "status": "affected",
              "version": "70N",
              "versionType": "custom"
            },
            {
              "lessThan": "206",
              "status": "affected",
              "version": "60N",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Vivid T",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThan": "206",
              "status": "affected",
              "version": "T8",
              "versionType": "custom"
            },
            {
              "lessThan": "206",
              "status": "affected",
              "version": "T9",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Vivid iq",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThan": "206",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Voluson Expert 16",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "status": "affected",
              "version": "0"
            },
            {
              "lessThan": "Ext1",
              "status": "affected",
              "version": "BT24",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Voluson Expert 18",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "status": "affected",
              "version": "0"
            },
            {
              "lessThan": "Ext1",
              "status": "affected",
              "version": "BT24",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Voluson Expert 22",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "status": "affected",
              "version": "0"
            },
            {
              "lessThan": "Ext1",
              "status": "affected",
              "version": "BT24",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Voluson SWIFT",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "status": "affected",
              "version": "0"
            },
            {
              "status": "affected",
              "version": "BT24"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Invenia ABUS 2.0",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThanOrEqual": "2.2.7",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "LOGIQ E10",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThan": "R3.2.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "LOGIQ E10s",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThan": "R3.2.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "LOGIQ Fortis",
          "vendor": "GE HealthCare",
          "versions": [
            {
              "lessThan": "R3.2.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Andrea Palanca and Gabriele Quagliarella of Nozomi Networks"
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "OS command injection vulnerabilities in GE HealthCare ultrasound devices"
            }
          ],
          "value": "OS command injection vulnerabilities in GE HealthCare ultrasound devices"
        }
      ],
      "impacts": [
        {
          "capecId": "CAPEC-6",
          "descriptions": [
            {
              "lang": "en",
              "value": "CAPEC-6 Argument Injection"
            }
          ]
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "LOCAL",
            "availabilityImpact": "HIGH",
            "baseScore": 8.4,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-78",
              "description": "CWE-78 Improper Neutralization of Special Elements used in an OS Command (\u0027OS Command Injection\u0027)",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2024-05-14T16:04:57.780Z",
        "orgId": "171caf72-b841-4e04-a68e-93493aff2b94",
        "shortName": "GEHC"
      },
      "references": [
        {
          "url": "https://securityupdate.gehealthcare.com/"
        }
      ],
      "source": {
        "discovery": "EXTERNAL"
      },
      "title": "OS command injection vulnerabilities in GE HealthCare ultrasound devices",
      "x_generator": {
        "engine": "Vulnogram 0.1.0-dev"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "171caf72-b841-4e04-a68e-93493aff2b94",
    "assignerShortName": "GEHC",
    "cveId": "CVE-2024-1628",
    "datePublished": "2024-05-14T16:04:57.780Z",
    "dateReserved": "2024-02-19T15:23:20.113Z",
    "dateUpdated": "2024-08-01T18:48:21.580Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "meta": {
    "nvd": "{\"cve\":{\"id\":\"CVE-2024-1628\",\"sourceIdentifier\":\"171caf72-b841-4e04-a68e-93493aff2b94\",\"published\":\"2024-05-14T16:16:00.470\",\"lastModified\":\"2024-05-14T19:18:31.490\",\"vulnStatus\":\"Awaiting Analysis\",\"descriptions\":[{\"lang\":\"en\",\"value\":\"OS command injection vulnerabilities in GE HealthCare ultrasound devices\"},{\"lang\":\"es\",\"value\":\"Vulnerabilidades de inyecci\u00f3n de comandos del sistema operativo en dispositivos de ultrasonido GE HealthCare\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"171caf72-b841-4e04-a68e-93493aff2b94\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H\",\"attackVector\":\"LOCAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\",\"baseScore\":8.4,\"baseSeverity\":\"HIGH\"},\"exploitabilityScore\":2.5,\"impactScore\":5.9}]},\"weaknesses\":[{\"source\":\"171caf72-b841-4e04-a68e-93493aff2b94\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-78\"}]}],\"references\":[{\"url\":\"https://securityupdate.gehealthcare.com/\",\"source\":\"171caf72-b841-4e04-a68e-93493aff2b94\"}]}}"
  }
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading...

Loading...

Loading...

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.