CVE-2025-14096 (GCVE-0-2025-14096)
Vulnerability from cvelistv5 – Published: 2025-12-17 12:19 – Updated: 2025-12-17 14:43
VLAI?
Title
Credential Disclosure vulnerability in Radiometer Products
Summary
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Severity ?
8.4 (High)
Assigner
References
Impacted products
| Vendor | Product | Version | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers |
Affected:
Windows 7 Operating system
(Radiometer Customized Windows Operating System)
Affected: Windows XP Operating system (Radiometer Customized Windows Operating System) |
||||||||||||
|
||||||||||||||
Credits
Florian Hauser and Fabian Weber from CODE WHITE GmbH
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-14096",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-12-17T14:42:57.843796Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T14:43:16.523Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"platforms": [
"Windows XP",
"Windows 7"
],
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "AQT90 FLEX Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "ABL800 BASIC and ABL800 FLEX analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
}
],
"configurations": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Requires physical access to the Analyzer.\n\n\u003cbr\u003e"
}
],
"value": "Requires physical access to the Analyzer."
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Florian Hauser and Fabian Weber from CODE WHITE GmbH"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\u003cbr\u003e\u003cbr\u003eOther related CVE\u0027s are CVE-2025-14095 \u0026amp; CVE-2025-14097.\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eRequired Configuration for Exposure:\u003cbr\u003e\u003c/span\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAttacker requires physical access to the analyzer.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003eTemporary work Around:\u003cbr\u003eOnly authorized people can physically access the analyzer.\u003cbr\u003e\u003cbr\u003ePermanent solution:\u003cbr\u003eLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\u003cbr\u003e\u003cbr\u003eExploit Status:\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e"
}
],
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"exploits": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\n\n\u003cbr\u003e"
}
],
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"impacts": [
{
"capecId": "CAPEC-69",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-69 \u2014 Target Programs with Elevated Privileges"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 8.4,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798: Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
},
{
"descriptions": [
{
"cweId": "CWE-250",
"description": "CWE-250: Execution with Unnecessary Privileges",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T13:03:05.696Z",
"orgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"shortName": "Radiometer"
},
"references": [
{
"url": "https://www.radiometer.com/myradiometer"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution."
}
],
"source": {
"discovery": "EXTERNAL"
},
"timeline": [
{
"lang": "en",
"time": "2025-12-17T13:01:00.000Z",
"value": "CVE Published"
}
],
"title": "Credential Disclosure vulnerability in Radiometer Products",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Only authorized people can physically access the analyzer.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Only authorized people can physically access the analyzer."
}
],
"x_generator": {
"engine": "Vulnogram 0.5.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"assignerShortName": "Radiometer",
"cveId": "CVE-2025-14096",
"datePublished": "2025-12-17T12:19:27.489Z",
"dateReserved": "2025-12-05T10:50:00.566Z",
"dateUpdated": "2025-12-17T14:43:16.523Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"vulnerability-lookup:meta": {
"nvd": "{\"cve\":{\"id\":\"CVE-2025-14096\",\"sourceIdentifier\":\"46b595e9-1acc-41cb-9398-adaf98d37a9b\",\"published\":\"2025-12-17T13:15:57.890\",\"lastModified\":\"2025-12-18T15:08:06.237\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\\n\\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\\n\\n\\n\\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\\n\\n\\n\\n\\n\\nRequired Configuration for Exposure:\\n\\n\\nAttacker requires physical access to the analyzer.\\n\\n\\n\\nTemporary work Around:\\nOnly authorized people can physically access the analyzer.\\n\\nPermanent solution:\\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\\n\\nExploit Status:\\n\\n\\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"46b595e9-1acc-41cb-9398-adaf98d37a9b\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H\",\"baseScore\":8.4,\"baseSeverity\":\"HIGH\",\"attackVector\":\"LOCAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":2.5,\"impactScore\":5.9}]},\"weaknesses\":[{\"source\":\"46b595e9-1acc-41cb-9398-adaf98d37a9b\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-250\"},{\"lang\":\"en\",\"value\":\"CWE-798\"}]}],\"references\":[{\"url\":\"https://www.radiometer.com/myradiometer\",\"source\":\"46b595e9-1acc-41cb-9398-adaf98d37a9b\"}]}}",
"vulnrichment": {
"containers": "{\"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2025-14096\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"no\"}, {\"Technical Impact\": \"total\"}], \"version\": \"2.0.3\", \"timestamp\": \"2025-12-17T14:42:57.843796Z\"}}}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2025-12-17T14:43:12.844Z\"}}], \"cna\": {\"title\": \"Credential Disclosure vulnerability in Radiometer Products\", \"source\": {\"discovery\": \"EXTERNAL\"}, \"credits\": [{\"lang\": \"en\", \"type\": \"reporter\", \"value\": \"Florian Hauser and Fabian Weber from CODE WHITE GmbH\"}], \"impacts\": [{\"capecId\": \"CAPEC-69\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-69 \\u2014 Target Programs with Elevated Privileges\"}]}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"UNCHANGED\", \"version\": \"3.1\", \"baseScore\": 8.4, \"attackVector\": \"LOCAL\", \"baseSeverity\": \"HIGH\", \"vectorString\": \"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H\", \"integrityImpact\": \"HIGH\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"HIGH\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Radiometer Medical Aps\", \"product\": \"ABL90 FLEX and ABL90 FLEX PLUS Analyzers\", \"versions\": [{\"status\": \"affected\", \"version\": \"Windows 7 Operating system\", \"versionType\": \"Radiometer Customized Windows Operating System\"}, {\"status\": \"affected\", \"version\": \"Windows XP Operating system\", \"versionType\": \"Radiometer Customized Windows Operating System\"}], \"platforms\": [\"Windows XP\", \"Windows 7\"], \"defaultStatus\": \"unaffected\"}, {\"vendor\": \"Radiometer Medical Aps\", \"product\": \"AQT90 FLEX Analyzers\", \"versions\": [{\"status\": \"affected\", \"version\": \"Windows 7 Operating system\", \"versionType\": \"Radiometer Customized Windows Operating System\"}, {\"status\": \"affected\", \"version\": \"Windows XP Operating system\", \"versionType\": \"Radiometer Customized Windows Operating System\"}], \"platforms\": [\"Windows 7\", \"Windows XP\"], \"defaultStatus\": \"unaffected\"}, {\"vendor\": \"Radiometer Medical Aps\", \"product\": \"ABL800 BASIC and ABL800 FLEX analyzers\", \"versions\": [{\"status\": \"affected\", \"version\": \"Windows 7 Operating system\", \"versionType\": \"Radiometer Customized Windows Operating System\"}, {\"status\": \"affected\", \"version\": \"Windows XP Operating system\", \"versionType\": \"Radiometer Customized Windows Operating System\"}], \"platforms\": [\"Windows 7\", \"Windows XP\"], \"defaultStatus\": \"unaffected\"}], \"exploits\": [{\"lang\": \"en\", \"value\": \"Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\\n\\n\u003cbr\u003e\", \"base64\": false}]}], \"timeline\": [{\"lang\": \"en\", \"time\": \"2025-12-17T13:01:00.000Z\", \"value\": \"CVE Published\"}], \"solutions\": [{\"lang\": \"en\", \"value\": \"Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\\n\\n\\n\\n\u003cbr\u003e\", \"base64\": false}]}], \"references\": [{\"url\": \"https://www.radiometer.com/myradiometer\"}], \"workarounds\": [{\"lang\": \"en\", \"value\": \"Only authorized people can physically access the analyzer.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"Only authorized people can physically access the analyzer.\\n\\n\\n\\n\u003cbr\u003e\", \"base64\": false}]}], \"x_generator\": {\"engine\": \"Vulnogram 0.5.0\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\\n\\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\\n\\n\\n\\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\\n\\n\\n\\n\\n\\nRequired Configuration for Exposure:\\n\\n\\nAttacker requires physical access to the analyzer.\\n\\n\\n\\nTemporary work Around:\\nOnly authorized people can physically access the analyzer.\\n\\nPermanent solution:\\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\\n\\nExploit Status:\\n\\n\\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\u003cbr\u003e\u003cbr\u003eOther related CVE\u0027s are CVE-2025-14095 \u0026amp; CVE-2025-14097.\u003cbr\u003e\u003cbr\u003e\\n\\n\u003cspan style=\\\"background-color: rgb(255, 255, 255);\\\"\u003eAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\u003c/span\u003e\\n\\n\u003cbr\u003e\u003cbr\u003e\\n\\n\u003cspan style=\\\"background-color: rgb(255, 255, 255);\\\"\u003eRequired Configuration for Exposure:\u003cbr\u003e\u003c/span\u003e\\n\\n\u003cspan style=\\\"background-color: rgb(255, 255, 255);\\\"\u003eAttacker requires physical access to the analyzer.\u003c/span\u003e\\n\\n\u003cbr\u003e\u003cbr\u003eTemporary work Around:\u003cbr\u003eOnly authorized people can physically access the analyzer.\u003cbr\u003e\u003cbr\u003ePermanent solution:\u003cbr\u003eLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\u003cbr\u003e\u003cbr\u003eExploit Status:\u003cbr\u003e\\n\\n\u003cspan style=\\\"background-color: rgb(255, 255, 255);\\\"\u003eResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\u003c/span\u003e\\n\\n\u003cbr\u003e\u003cbr\u003e\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-798\", \"description\": \"CWE-798: Use of Hard-coded Credentials\"}]}, {\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-250\", \"description\": \"CWE-250: Execution with Unnecessary Privileges\"}]}], \"configurations\": [{\"lang\": \"en\", \"value\": \"Requires physical access to the Analyzer.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"Requires physical access to the Analyzer.\\n\\n\u003cbr\u003e\", \"base64\": false}]}], \"providerMetadata\": {\"orgId\": \"46b595e9-1acc-41cb-9398-adaf98d37a9b\", \"shortName\": \"Radiometer\", \"dateUpdated\": \"2025-12-17T13:03:05.696Z\"}}}",
"cveMetadata": "{\"cveId\": \"CVE-2025-14096\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2025-12-17T14:43:16.523Z\", \"dateReserved\": \"2025-12-05T10:50:00.566Z\", \"assignerOrgId\": \"46b595e9-1acc-41cb-9398-adaf98d37a9b\", \"datePublished\": \"2025-12-17T12:19:27.489Z\", \"assignerShortName\": \"Radiometer\"}",
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…