ICSMA-20-023-01
Vulnerability from csaf_cisa
Published
2020-01-23 00:00
Modified
2020-01-23 00:00
Summary
GE CARESCAPE, ApexPro, and Clinical Information Center systems

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could occur when an attacker gains access to the mission critical (MC) and/or information exchange (IX) networks due to improper configuration or physical access to devices. An exploit could result in a loss of monitoring and/or loss of alarms during active patient monitoring. These vulnerabilities, if exploited, may allow an attacker to obtain PHI data, make changes at the operating system level of the device, with effects such as rendering the device unusable, otherwise interfering with the function of the device and/or making certain changes to alarm settings on connected patient monitors, and/or utilizing services used for remote viewing and control of devices on the network to access the clinical user interface and make changes to device settings and alarm limits, which could result in missed or unnecessary alarms or silencing of some alarms.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
United States
Recommended Practices
At the time of the publication of this advisory, GE was not aware of any reported incidences of a cyber attack in a clinical use or any reported injuries associated with any of these vulnerabilities. In addition, GE is developing software updates/patches with additional security enhancements. In accordance with GE 's continual cybersecurity hygiene process, users can access GE 's security website to receive the most up-to-date information and subscribe to receive notifications when new updates/patches are available. CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures. CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies. Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities.



{
  "document": {
    "acknowledgments": [
      {
        "names": [
          "Elad Luz"
        ],
        "organization": "CyberMDX",
        "summary": "reporting these vulnerabilities to CISA"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE",
        "url": "https://us-cert.cisa.gov/tlp/"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "legal_disclaimer",
        "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
        "title": "Legal Notice"
      },
      {
        "category": "summary",
        "text": "Successful exploitation of these vulnerabilities could occur when an attacker gains access to the mission critical (MC) and/or information exchange (IX) networks due to improper configuration or physical access to devices. An exploit could result in a loss of monitoring and/or loss of alarms during active patient monitoring. These vulnerabilities, if exploited, may allow an attacker to obtain PHI data, make changes at the operating system level of the device, with effects such as rendering the device unusable, otherwise interfering with the function of the device and/or making certain changes to alarm settings on connected patient monitors, and/or utilizing services used for remote viewing and control of devices on the network to access the clinical user interface and make changes to device settings and alarm limits, which could result in missed or unnecessary alarms or silencing of some alarms.",
        "title": "Risk evaluation"
      },
      {
        "category": "other",
        "text": "Healthcare and Public Health",
        "title": "Critical infrastructure sectors"
      },
      {
        "category": "other",
        "text": "Worldwide",
        "title": "Countries/areas deployed"
      },
      {
        "category": "other",
        "text": "United States",
        "title": "Company headquarters location"
      },
      {
        "category": "general",
        "text": "At the time of the publication of this advisory, GE was not aware of any reported incidences of a cyber attack in a clinical use or any reported injuries associated with any of these vulnerabilities.\nIn addition, GE is developing software updates/patches with additional security enhancements. In accordance with GE \u0027s continual cybersecurity hygiene process, users can access GE \u0027s security website to receive the most up-to-date information and subscribe to receive notifications when new updates/patches are available.\nCISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
        "title": "Recommended Practices"
      },
      {
        "category": "other",
        "text": "No known public exploits specifically target these vulnerabilities.",
        "title": "Exploitability"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-20-023-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2020/icsma-20-023-01.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-20-023-01 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-023-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://securityupdate.gehealthcare.com/"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.us-cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.us-cert.gov/ics/tips/ICS-TIP-12-146-01B"
      }
    ],
    "title": "GE CARESCAPE, ApexPro, and Clinical Information Center systems",
    "tracking": {
      "current_release_date": "2020-01-23T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA CSAF Generator",
          "version": "1.0.0"
        }
      },
      "id": "ICSMA-20-023-01",
      "initial_release_date": "2020-01-23T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2020-01-23T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSMA-20-023-01 GE CARESCAPE, ApexPro, Clinical Information Center systems"
        }
      ],
      "status": "final",
      "version": "1"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version",
                "name": "2.X",
                "product": {
                  "name": "B450: Version 2.X (Impacted by CVE-2020- 6962 and CVE-2020-6965)",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "B450"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "2.X",
                "product": {
                  "name": "B850: Version 2.X (Impacted by CVE-2020- 6962 and CVE-2020-6965)",
                  "product_id": "CSAFPID-0002"
                }
              }
            ],
            "category": "product_name",
            "name": "B850"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "1.X",
                "product": {
                  "name": "B650: Version 1.X (Impacted by CVE-2020- 6962 and CVE-2020-6965)",
                  "product_id": "CSAFPID-0003"
                }
              }
            ],
            "category": "product_name",
            "name": "B650"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "2.X",
                "product": {
                  "name": "CARESCAPE Central Station (CSCS): Versions 2.X (Impacted by CVE-2020- 6962 and CVE-2020-6964)",
                  "product_id": "CSAFPID-0004"
                }
              }
            ],
            "category": "product_name",
            "name": "CARESCAPE Central Station (CSCS)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c= 4.2",
                "product": {
                  "name": "ApexPro Telemetry Server: Versions 4.2 and prior",
                  "product_id": "CSAFPID-0005"
                }
              }
            ],
            "category": "product_name",
            "name": "ApexPro Telemetry Server"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "4.3",
                "product": {
                  "name": "CARESCAPE Telemetry Server: Version 4.3 (Impacted by CVE-2020- 6962 and CVE-2020-6961)",
                  "product_id": "CSAFPID-0006"
                }
              }
            ],
            "category": "product_name",
            "name": "CARESCAPE Telemetry Server"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "1.X",
                "product": {
                  "name": "CARESCAPE Central Station (CSCS): Versions 1.X",
                  "product_id": "CSAFPID-0007"
                }
              }
            ],
            "category": "product_name",
            "name": "CARESCAPE Central Station (CSCS)"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "1.X",
                "product": {
                  "name": "B850: Version 1.X (Impacted by CVE-2020- 6962 and CVE-2020-6965)",
                  "product_id": "CSAFPID-0008"
                }
              }
            ],
            "category": "product_name",
            "name": "B850"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c= 4.2",
                "product": {
                  "name": "CARESCAPE Telemetry Server: Versions 4.2 and prior",
                  "product_id": "CSAFPID-0009"
                }
              }
            ],
            "category": "product_name",
            "name": "CARESCAPE Telemetry Server"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "2.X",
                "product": {
                  "name": "B650: Version 2.X (Impacted by CVE-2020- 6962 and CVE-2020-6965)",
                  "product_id": "CSAFPID-00010"
                }
              }
            ],
            "category": "product_name",
            "name": "B650"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "4.X | 5.X",
                "product": {
                  "name": "Clinical Information Center (CIC): Versions 4.X and 5.X",
                  "product_id": "CSAFPID-00011"
                }
              }
            ],
            "category": "product_name",
            "name": "Clinical Information Center (CIC)"
          }
        ],
        "category": "vendor",
        "name": "General Electric (GE)"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2020-6961",
      "cwe": {
        "id": "CWE-256",
        "name": "Plaintext Storage of a Password"
      },
      "notes": [
        {
          "category": "summary",
          "text": "A vulnerability exists in the affected products that could allow an attacker to obtain access to the SSH private key in configuration files.CVE-2020-6961 has been assigned to this vulnerability. A CVSS v3 base score of 10 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H). To see how this vulnerability affects each unique configuration, please apply the environmental portion of CVSS v3.0 score.",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009",
          "CSAFPID-00010",
          "CSAFPID-00011"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-6961"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "GE recommends users confirm the proper configuration of the MC and IX networks to ensure that the isolation and configuration meet the requirements listed in the Patient Monitoring Network Configuration Guide, CARESCAPE Network Configuration Guide, and product technical and service manuals. These can be obtained by contacting GE via the customer support portal with a valid support account. A properly isolated network requires an attacker to gain physical access in order to carry out an exploit.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.gehealthcare.com/support/support-documentation-library"
        },
        {
          "category": "vendor_fix",
          "details": "GE recommends that, in addition to applying network management best practices, users ensure:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 10.0,
            "baseSeverity": "CRITICAL",
            "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2020-6962",
      "cwe": {
        "id": "CWE-20",
        "name": "Improper Input Validation"
      },
      "notes": [
        {
          "category": "summary",
          "text": "An input validation vulnerability exists in the web-based system configuration utility that could allow an attacker to obtain arbitrary remote code execution.CVE-2020-6962 has been assigned to this vulnerability. A CVSS v3 base score of 10 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H). To see how this vulnerability affects each unique configuration, please apply the environmental portion of CVSS v3.0 score",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009",
          "CSAFPID-00010",
          "CSAFPID-00011"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-6962"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "GE recommends users confirm the proper configuration of the MC and IX networks to ensure that the isolation and configuration meet the requirements listed in the Patient Monitoring Network Configuration Guide, CARESCAPE Network Configuration Guide, and product technical and service manuals. These can be obtained by contacting GE via the customer support portal with a valid support account. A properly isolated network requires an attacker to gain physical access in order to carry out an exploit.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.gehealthcare.com/support/support-documentation-library"
        },
        {
          "category": "vendor_fix",
          "details": "GE recommends that, in addition to applying network management best practices, users ensure:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 10.0,
            "baseSeverity": "CRITICAL",
            "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2020-6963",
      "cwe": {
        "id": "CWE-798",
        "name": "Use of Hard-coded Credentials"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.CVE-2020-6963 has been assigned to this vulnerability. A CVSS v3 base score of 10 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H). To see how this vulnerability affects each unique configuration, please apply the environmental portion of CVSS v3.0 score",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009",
          "CSAFPID-00010",
          "CSAFPID-00011"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-6963"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "GE recommends users confirm the proper configuration of the MC and IX networks to ensure that the isolation and configuration meet the requirements listed in the Patient Monitoring Network Configuration Guide, CARESCAPE Network Configuration Guide, and product technical and service manuals. These can be obtained by contacting GE via the customer support portal with a valid support account. A properly isolated network requires an attacker to gain physical access in order to carry out an exploit.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.gehealthcare.com/support/support-documentation-library"
        },
        {
          "category": "vendor_fix",
          "details": "GE recommends that, in addition to applying network management best practices, users ensure:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 10.0,
            "baseSeverity": "CRITICAL",
            "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2020-6964",
      "cwe": {
        "id": "CWE-306",
        "name": "Missing Authentication for Critical Function"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.CVE-2020-6964 has been assigned to this vulnerability. A CVSS v3 base score of 10 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H). To see how this vulnerability affects each unique configuration, please apply the environmental portion of CVSS v3.0 score",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009",
          "CSAFPID-00010",
          "CSAFPID-00011"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-6964"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "GE recommends users confirm the proper configuration of the MC and IX networks to ensure that the isolation and configuration meet the requirements listed in the Patient Monitoring Network Configuration Guide, CARESCAPE Network Configuration Guide, and product technical and service manuals. These can be obtained by contacting GE via the customer support portal with a valid support account. A properly isolated network requires an attacker to gain physical access in order to carry out an exploit.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.gehealthcare.com/support/support-documentation-library"
        },
        {
          "category": "vendor_fix",
          "details": "GE recommends that, in addition to applying network management best practices, users ensure:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 10.0,
            "baseSeverity": "CRITICAL",
            "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2020-6965",
      "cwe": {
        "id": "CWE-434",
        "name": "Unrestricted Upload of File with Dangerous Type"
      },
      "notes": [
        {
          "category": "summary",
          "text": "A vulnerability in the software update mechanism allows an authenticated attacker to upload arbitrary files on the system through a crafted update package.CVE-2020-6965 has been assigned to this vulnerability. A CVSS v3 base score of 8.5 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:L/UI:N/S:C/C:H/I:H/A:H). To see how this vulnerability affects each unique configuration, please apply the environmental portion of CVSS v3.0 score",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009",
          "CSAFPID-00010",
          "CSAFPID-00011"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-6965"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:H/PR:L/UI:N/S:C/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "GE recommends users confirm the proper configuration of the MC and IX networks to ensure that the isolation and configuration meet the requirements listed in the Patient Monitoring Network Configuration Guide, CARESCAPE Network Configuration Guide, and product technical and service manuals. These can be obtained by contacting GE via the customer support portal with a valid support account. A properly isolated network requires an attacker to gain physical access in order to carry out an exploit.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.gehealthcare.com/support/support-documentation-library"
        },
        {
          "category": "vendor_fix",
          "details": "GE recommends that, in addition to applying network management best practices, users ensure:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 8.5,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:N/AC:H/PR:L/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2020-6966",
      "cwe": {
        "id": "CWE-326",
        "name": "Inadequate Encryption Strength"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected products utilize a weak encryption scheme for remote desktop control, which may allow an attacker to obtain remote code execution of devices on the network.CVE-2020-6966 has been assigned to this vulnerability. A CVSS v3 base score of 10 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H). To see how this vulnerability affects each unique configuration, please apply the environmental portion of CVSS v3.0 score",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009",
          "CSAFPID-00010",
          "CSAFPID-00011"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-6966"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "GE recommends users confirm the proper configuration of the MC and IX networks to ensure that the isolation and configuration meet the requirements listed in the Patient Monitoring Network Configuration Guide, CARESCAPE Network Configuration Guide, and product technical and service manuals. These can be obtained by contacting GE via the customer support portal with a valid support account. A properly isolated network requires an attacker to gain physical access in order to carry out an exploit.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.gehealthcare.com/support/support-documentation-library"
        },
        {
          "category": "vendor_fix",
          "details": "GE recommends that, in addition to applying network management best practices, users ensure:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 10.0,
            "baseSeverity": "CRITICAL",
            "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ]
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.