CVE-2024-48967 (GCVE-0-2024-48967)
Vulnerability from cvelistv5 – Published: 2024-11-14 21:40 – Updated: 2024-11-15 14:28
VLAI?
Summary
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
Severity ?
10 (Critical)
CWE
- CWE-778 - Insufficient Logging
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Baxter | Life2000 Ventilation System |
Affected:
06.08.00.00 and prior
|
{
"containers": {
"adp": [
{
"affected": [
{
"cpes": [
"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"
],
"defaultStatus": "unknown",
"product": "life2000_ventilator_firmware",
"vendor": "baxter",
"versions": [
{
"lessThan": "6.08.00.00",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"metrics": [
{
"other": {
"content": {
"id": "CVE-2024-48967",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2024-11-15T14:28:10.676566Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2024-11-15T14:28:13.605Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "Life2000 Ventilation System",
"vendor": "Baxter",
"versions": [
{
"status": "affected",
"version": "06.08.00.00 and prior"
}
]
}
],
"datePublic": "2024-11-14T21:38:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance."
}
],
"value": "The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance."
}
],
"impacts": [
{
"capecId": "CAPEC-117",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-117 Interception"
}
]
},
{
"capecId": "CAPEC-441",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-441 Malicious Logic Insertion"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 10,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "CHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-778",
"description": "CWE-778: Insufficient Logging",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2024-11-14T21:52:23.702Z",
"orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
"shortName": "Baxter"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "Life2000 ventilator and Service PC lack sufficient audit logging capabilities",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
"assignerShortName": "Baxter",
"cveId": "CVE-2024-48967",
"datePublished": "2024-11-14T21:40:44.904Z",
"dateReserved": "2024-10-10T19:24:34.436Z",
"dateUpdated": "2024-11-15T14:28:13.605Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1",
"vulnerability-lookup:meta": {
"fkie_nvd": {
"descriptions": "[{\"lang\": \"en\", \"value\": \"The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.\"}, {\"lang\": \"es\", \"value\": \"El respirador y la PC de servicio carecen de capacidades de registro de auditor\\u00eda suficientes para permitir la detecci\\u00f3n de actividad maliciosa y el posterior examen forense. Un atacante con acceso al respirador o a la PC de servicio podr\\u00eda, sin ser detectado, realizar cambios no autorizados en la configuraci\\u00f3n del respirador que resulten en la divulgaci\\u00f3n no autorizada de informaci\\u00f3n o tengan impactos no deseados en el rendimiento del dispositivo.\"}]",
"id": "CVE-2024-48967",
"lastModified": "2024-11-15T13:58:08.913",
"metrics": "{\"cvssMetricV31\": [{\"source\": \"productsecurity@baxter.com\", \"type\": \"Secondary\", \"cvssData\": {\"version\": \"3.1\", \"vectorString\": \"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\", \"baseScore\": 10.0, \"baseSeverity\": \"CRITICAL\", \"attackVector\": \"NETWORK\", \"attackComplexity\": \"LOW\", \"privilegesRequired\": \"NONE\", \"userInteraction\": \"NONE\", \"scope\": \"CHANGED\", \"confidentialityImpact\": \"HIGH\", \"integrityImpact\": \"HIGH\", \"availabilityImpact\": \"HIGH\"}, \"exploitabilityScore\": 3.9, \"impactScore\": 6.0}]}",
"published": "2024-11-14T22:15:17.927",
"references": "[{\"url\": \"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\", \"source\": \"productsecurity@baxter.com\"}]",
"sourceIdentifier": "productsecurity@baxter.com",
"vulnStatus": "Awaiting Analysis",
"weaknesses": "[{\"source\": \"productsecurity@baxter.com\", \"type\": \"Secondary\", \"description\": [{\"lang\": \"en\", \"value\": \"CWE-778\"}]}]"
},
"nvd": "{\"cve\":{\"id\":\"CVE-2024-48967\",\"sourceIdentifier\":\"productsecurity@baxter.com\",\"published\":\"2024-11-14T22:15:17.927\",\"lastModified\":\"2024-11-15T13:58:08.913\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.\"},{\"lang\":\"es\",\"value\":\"El respirador y la PC de servicio carecen de capacidades de registro de auditor\u00eda suficientes para permitir la detecci\u00f3n de actividad maliciosa y el posterior examen forense. Un atacante con acceso al respirador o a la PC de servicio podr\u00eda, sin ser detectado, realizar cambios no autorizados en la configuraci\u00f3n del respirador que resulten en la divulgaci\u00f3n no autorizada de informaci\u00f3n o tengan impactos no deseados en el rendimiento del dispositivo.\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\",\"baseScore\":10.0,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":3.9,\"impactScore\":6.0}]},\"weaknesses\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-778\"}]}],\"references\":[{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\",\"source\":\"productsecurity@baxter.com\"}]}}",
"vulnrichment": {
"containers": "{\"cna\": {\"title\": \"Life2000 ventilator and Service PC lack sufficient audit logging capabilities\", \"source\": {\"discovery\": \"UNKNOWN\"}, \"impacts\": [{\"capecId\": \"CAPEC-117\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-117 Interception\"}]}, {\"capecId\": \"CAPEC-441\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-441 Malicious Logic Insertion\"}]}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"CHANGED\", \"version\": \"3.1\", \"baseScore\": 10, \"attackVector\": \"NETWORK\", \"baseSeverity\": \"CRITICAL\", \"vectorString\": \"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\", \"integrityImpact\": \"HIGH\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"HIGH\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Baxter\", \"product\": \"Life2000 Ventilation System\", \"versions\": [{\"status\": \"affected\", \"version\": \"06.08.00.00 and prior\"}], \"defaultStatus\": \"unaffected\"}], \"datePublic\": \"2024-11-14T21:38:00.000Z\", \"references\": [{\"url\": \"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\"}], \"x_generator\": {\"engine\": \"Vulnogram 0.2.0\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-778\", \"description\": \"CWE-778: Insufficient Logging\"}]}], \"providerMetadata\": {\"orgId\": \"dba971b9-eb30-4121-91e1-3b45611354aa\", \"shortName\": \"Baxter\", \"dateUpdated\": \"2024-11-14T21:52:23.702Z\"}}, \"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2024-48967\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"yes\"}, {\"Technical Impact\": \"total\"}], \"version\": \"2.0.3\", \"timestamp\": \"2024-11-15T14:28:10.676566Z\"}}}], \"affected\": [{\"cpes\": [\"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*\"], \"vendor\": \"baxter\", \"product\": \"life2000_ventilator_firmware\", \"versions\": [{\"status\": \"affected\", \"version\": \"0\", \"lessThan\": \"6.08.00.00\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}], \"providerMetadata\": {\"shortName\": \"CISA-ADP\", \"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"dateUpdated\": \"2024-11-15T14:27:46.856Z\"}}]}",
"cveMetadata": "{\"cveId\": \"CVE-2024-48967\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2024-11-14T21:52:23.702Z\", \"dateReserved\": \"2024-10-10T19:24:34.436Z\", \"assignerOrgId\": \"dba971b9-eb30-4121-91e1-3b45611354aa\", \"datePublished\": \"2024-11-14T21:40:44.904Z\", \"assignerShortName\": \"Baxter\"}",
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…