GCVE Workshop - 22 September 2026 (14:00-18:00), Luxembourg Before The Vulnopticon Conference - Registration

CVE-2026-18164 (GCVE-0-2026-18164)

Vulnerability from cvelistv5 – Published: 2026-08-13 19:13 – Updated: 2026-08-14 16:37
VLAI
Title
Flow Neuroscience FL-100 Use of Hard-coded Credentials
Summary
An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state.
SSVC
Exploitation: none Automatable: no Technical Impact: partial
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2026-08-14 16:36 UTC
CWE
Impacted products
Vendor Product Version CPE status
Flow Neuroscience FL-100 Affected: 0 , < July 2026 (custom)
guessed Create a notification for this product.
Halo Neuroscience FL-100 Affected: 0 , < July 2026 (custom)
guessed Create a notification for this product.
Date Public
2026-08-13 17:00
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2026-18164",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2026-08-14T16:36:43.830245Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2026-08-14T16:37:03.741Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "FL-100",
          "vendor": "Flow Neuroscience",
          "versions": [
            {
              "lessThan": "July 2026",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "FL-100",
          "vendor": "Halo Neuroscience",
          "versions": [
            {
              "lessThan": "July 2026",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "A.C. Buglione reported this vulnerability to CISA."
        }
      ],
      "datePublic": "2026-08-13T17:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state."
            }
          ],
          "value": "An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "ADJACENT",
            "baseScore": 7.2,
            "baseSeverity": "HIGH",
            "exploitMaturity": "NOT_DEFINED",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:H/VI:N/VA:H/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "HIGH",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "NONE",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "ADJACENT_NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 8.1,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "NONE",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-798",
              "description": "CWE-798",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2026-08-13T19:13:04.494Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-225-01"
        },
        {
          "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-225-01.json"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "Users are encouraged to install the latest firmware updates provided by Flow Neuroscience via the Flow app."
            }
          ],
          "value": "Users are encouraged to install the latest firmware updates provided by Flow Neuroscience via the Flow app."
        }
      ],
      "source": {
        "advisory": "ICSMA-26-225-01",
        "discovery": "EXTERNAL"
      },
      "title": "Flow Neuroscience FL-100 Use of Hard-coded Credentials",
      "x_generator": {
        "engine": "Vulnogram 1.0.4"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2026-18164",
    "datePublished": "2026-08-13T19:13:04.494Z",
    "dateReserved": "2026-07-28T20:16:40.248Z",
    "dateUpdated": "2026-08-14T16:37:03.741Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.2",
  "vulnerability-lookup:meta": {
    "epss": {
      "cve": "CVE-2026-18164",
      "date": "2026-09-17",
      "epss": "0.00229",
      "percentile": "0.13846"
    },
    "nvd": "{\"cve\":{\"id\":\"CVE-2026-18164\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2026-08-13T20:17:19.440\",\"lastModified\":\"2026-09-08T19:30:43.093\",\"vulnStatus\":\"Deferred\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state.\"}],\"affected\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"affectedData\":[{\"vendor\":\"Flow Neuroscience\",\"product\":\"FL-100\",\"defaultStatus\":\"unaffected\",\"versions\":[{\"version\":\"0\",\"lessThan\":\"July 2026\",\"versionType\":\"custom\",\"status\":\"affected\"}]},{\"vendor\":\"Halo Neuroscience\",\"product\":\"FL-100\",\"defaultStatus\":\"unaffected\",\"versions\":[{\"version\":\"0\",\"lessThan\":\"July 2026\",\"versionType\":\"custom\",\"status\":\"affected\"}]}]}],\"metrics\":{\"cvssMetricV40\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"4.0\",\"vectorString\":\"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:H/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X\",\"baseScore\":7.2,\"baseSeverity\":\"HIGH\",\"attackVector\":\"ADJACENT\",\"attackComplexity\":\"LOW\",\"attackRequirements\":\"NONE\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"vulnConfidentialityImpact\":\"HIGH\",\"vulnIntegrityImpact\":\"NONE\",\"vulnAvailabilityImpact\":\"HIGH\",\"subConfidentialityImpact\":\"NONE\",\"subIntegrityImpact\":\"NONE\",\"subAvailabilityImpact\":\"NONE\",\"exploitMaturity\":\"NOT_DEFINED\",\"confidentialityRequirement\":\"NOT_DEFINED\",\"integrityRequirement\":\"NOT_DEFINED\",\"availabilityRequirement\":\"NOT_DEFINED\",\"modifiedAttackVector\":\"NOT_DEFINED\",\"modifiedAttackComplexity\":\"NOT_DEFINED\",\"modifiedAttackRequirements\":\"NOT_DEFINED\",\"modifiedPrivilegesRequired\":\"NOT_DEFINED\",\"modifiedUserInteraction\":\"NOT_DEFINED\",\"modifiedVulnConfidentialityImpact\":\"NOT_DEFINED\",\"modifiedVulnIntegrityImpact\":\"NOT_DEFINED\",\"modifiedVulnAvailabilityImpact\":\"NOT_DEFINED\",\"modifiedSubConfidentialityImpact\":\"NOT_DEFINED\",\"modifiedSubIntegrityImpact\":\"NOT_DEFINED\",\"modifiedSubAvailabilityImpact\":\"NOT_DEFINED\",\"Safety\":\"NOT_DEFINED\",\"Automatable\":\"NOT_DEFINED\",\"Recovery\":\"NOT_DEFINED\",\"valueDensity\":\"NOT_DEFINED\",\"vulnerabilityResponseEffort\":\"NOT_DEFINED\",\"providerUrgency\":\"NOT_DEFINED\"}}],\"cvssMetricV31\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H\",\"baseScore\":8.1,\"baseSeverity\":\"HIGH\",\"attackVector\":\"ADJACENT_NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":2.8,\"impactScore\":5.2}],\"ssvcV203\":[{\"source\":\"134c704f-9b21-4f2e-91b3-4a467353bcc0\",\"ssvcData\":{\"timestamp\":\"2026-08-14T16:36:43.830245Z\",\"id\":\"CVE-2026-18164\",\"options\":[{\"exploitation\":\"none\"},{\"automatable\":\"no\"},{\"technicalImpact\":\"partial\"}],\"role\":\"CISA Coordinator\",\"version\":\"2.0.3\"}}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-798\"}]}],\"references\":[{\"url\":\"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-225-01.json\",\"source\":\"ics-cert@hq.dhs.gov\"},{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-225-01\",\"source\":\"ics-cert@hq.dhs.gov\"}]}}"
  }
}



Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.

Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…

Related by attack behaviour

Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.


Loading…