GHSA-2VHP-V3RH-J2QR
Vulnerability from github – Published: 2022-05-13 01:38 – Updated: 2022-05-13 01:38
VLAI?
Details
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Severity ?
9.8 (Critical)
{
"affected": [],
"aliases": [
"CVE-2015-3956"
],
"database_specific": {
"cwe_ids": [
"CWE-345"
],
"github_reviewed": false,
"github_reviewed_at": null,
"nvd_published_at": "2019-03-25T18:29:00Z",
"severity": "CRITICAL"
},
"details": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.",
"id": "GHSA-2vhp-v3rh-j2qr",
"modified": "2022-05-13T01:38:53Z",
"published": "2022-05-13T01:38:53Z",
"references": [
{
"type": "ADVISORY",
"url": "https://nvd.nist.gov/vuln/detail/CVE-2015-3956"
},
{
"type": "WEB",
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
}
],
"schema_version": "1.4.0",
"severity": [
{
"score": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"type": "CVSS_V3"
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…