GHSA-83WF-R644-8PQQ
Vulnerability from github – Published: 2026-06-02 21:30 – Updated: 2026-06-03 18:33Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
{
"affected": [],
"aliases": [
"CVE-2019-25721"
],
"database_specific": {
"cwe_ids": [
"CWE-400"
],
"github_reviewed": false,
"github_reviewed_at": null,
"nvd_published_at": "2026-06-02T20:16:23Z",
"severity": "HIGH"
},
"details": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.",
"id": "GHSA-83wf-r644-8pqq",
"modified": "2026-06-03T18:33:08Z",
"published": "2026-06-02T21:30:41Z",
"references": [
{
"type": "ADVISORY",
"url": "https://nvd.nist.gov/vuln/detail/CVE-2019-25721"
},
{
"type": "WEB",
"url": "https://static.draeger.com/security"
},
{
"type": "WEB",
"url": "https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf"
},
{
"type": "WEB",
"url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service"
}
],
"schema_version": "1.4.0",
"severity": [
{
"score": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"type": "CVSS_V3"
},
{
"score": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X",
"type": "CVSS_V4"
}
]
}
Sightings
| Author | Source | Type | Date | Other |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
The approach is described in our paper Mapping CVEs to MITRE ATT&CK Techniques: A Curated Gold-Set Classifier and the Limits of LLM-Assisted Label Expansion.
Browse all ATT&CK techniques and the vulnerabilities related to each.
Related by attack behaviour
Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.