GHSA-RXFJ-GVQJ-6MMM
Vulnerability from github – Published: 2025-07-25 15:30 – Updated: 2025-07-25 15:30
VLAI?
Details
Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify system functionality.
This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
Severity ?
6.8 (Medium)
{
"affected": [],
"aliases": [
"CVE-2025-4395"
],
"database_specific": {
"cwe_ids": [
"CWE-258"
],
"github_reviewed": false,
"github_reviewed_at": null,
"nvd_published_at": "2025-07-24T07:15:53Z",
"severity": "MODERATE"
},
"details": "Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify system functionality. \n\nThis issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025",
"id": "GHSA-rxfj-gvqj-6mmm",
"modified": "2025-07-25T15:30:42Z",
"published": "2025-07-25T15:30:41Z",
"references": [
{
"type": "ADVISORY",
"url": "https://nvd.nist.gov/vuln/detail/CVE-2025-4395"
},
{
"type": "WEB",
"url": "https://global.medtronic.com/xg-en/product-security/security-bulletins.html"
}
],
"schema_version": "1.4.0",
"severity": [
{
"score": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"type": "CVSS_V3"
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…