ICSMA-17-292-01

Vulnerability from csaf_cisa - Published: 2017-10-19 00:00 - Updated: 2017-10-19 00:00
Summary
ICSMA-17-292-01_Boston Scientific ZOOM LATITUDE PRM Vulnerabilities

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Exploitability
No known public exploits specifically target these vulnerabilities.

{
  "document": {
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "other",
        "text": "No known public exploits specifically target these vulnerabilities.",
        "title": "Exploitability"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "CISAservicedesk@cisa.dhs.gov",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-17-292-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2017/icsma-17-292-01.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-17-292-01 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-292-01"
      }
    ],
    "title": "ICSMA-17-292-01_Boston Scientific ZOOM LATITUDE PRM Vulnerabilities",
    "tracking": {
      "current_release_date": "2017-10-19T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA USCert CSAF Generator",
          "version": "1"
        }
      },
      "id": "ICSMA-17-292-01",
      "initial_release_date": "2017-10-19T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2017-10-19T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSMA-17-292-01 Boston Scientific Zoom Latitude PRM Vulnerabilities"
        }
      ],
      "status": "final",
      "version": "1"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version",
                "name": "vers:all/*",
                "product": {
                  "name": "ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) - Model 3120: all versions",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) - Model 3120"
          }
        ],
        "category": "vendor",
        "name": "Boston Scientific"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2017-14014",
      "cwe": {
        "id": "CWE-321",
        "name": "Use of Hard-coded Cryptographic Key"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected device uses a hard-coded cryptographic key to encrypt PHI prior to having it transferred to removable media.",
          "title": "Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Boston Scientific will not be issuing a product update to address the identified vulnerabilities in the ZOOM LATITUDE PRM - Model 3120. Boston Scientific has identified compensating controls to reduce the risk of exploitation and recommends that users implement the following measures:",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "http://ics-cert.us-cert.gov/content/recommended-practices"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 4.6,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ],
      "title": "CVE-2017-14014"
    },
    {
      "cve": "CVE-2017-14012",
      "cwe": {
        "id": "CWE-311",
        "name": "Missing Encryption of Sensitive Data"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected device does not encrypt PHI at rest.",
          "title": "Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Boston Scientific will not be issuing a product update to address the identified vulnerabilities in the ZOOM LATITUDE PRM - Model 3120. Boston Scientific has identified compensating controls to reduce the risk of exploitation and recommends that users implement the following measures:",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "http://ics-cert.us-cert.gov/content/recommended-practices"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 4.6,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ],
      "title": "CVE-2017-14012"
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.


Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…