ICSMA-19-113-01
Vulnerability from csaf_cisa - Published: 2019-04-23 00:00 - Updated: 2019-04-23 00:00Summary
Fujifilm FCR Capsula X/Carbon X
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could result in a denial-of-service condition in affected cassette reader units, causing potential image loss or device unavailability. Attackers could gain unauthorized access to the underlying operating system, allowing arbitrary code execution.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Japan
Recommended Practices
NCCIC recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Recommended Practices
NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS-CERT website in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities.
{
"document": {
"acknowledgments": [
{
"names": [
"Marc Ruef",
"Rocco Gagliardi"
],
"organization": "Scip AG",
"summary": "reporting these vulnerabilities to NCCIC"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities could result in a denial-of-service condition in affected cassette reader units, causing potential image loss or device unavailability. Attackers could gain unauthorized access to the underlying operating system, allowing arbitrary code execution.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "Japan",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "NCCIC recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS-CERT website in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "other",
"text": "No known public exploits specifically target these vulnerabilities.",
"title": "Exploitability"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-19-113-01 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2019/icsma-19-113-01.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-19-113-01 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-113-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/ics/alerts/ICS-ALERT-10-301-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "Fujifilm FCR Capsula X/Carbon X",
"tracking": {
"current_release_date": "2019-04-23T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-19-113-01",
"initial_release_date": "2019-04-23T00:00:00.000000Z",
"revision_history": [
{
"date": "2019-04-23T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-19-113-01 Fujifilm FCR Capsula X/Carbon X"
}
],
"status": "final",
"version": "1"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version",
"name": "CR-IR 357",
"product": {
"name": "FCR XC-2: CR-IR 357",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "FCR XC-2"
},
{
"branches": [
{
"category": "product_version",
"name": "CR-IR 357",
"product": {
"name": "FCR Capsula X: CR-IR 357",
"product_id": "CSAFPID-0002"
}
}
],
"category": "product_name",
"name": "FCR Capsula X"
},
{
"branches": [
{
"category": "product_version",
"name": "CR-IR 357",
"product": {
"name": "FCR Carbon X: CR-IR 357",
"product_id": "CSAFPID-0003"
}
}
],
"category": "product_name",
"name": "FCR Carbon X"
}
],
"category": "vendor",
"name": "Fujifilm"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2019-10948",
"cwe": {
"id": "CWE-400",
"name": "Uncontrolled Resource Consumption"
},
"notes": [
{
"category": "summary",
"text": "The device is susceptible to a denial-of-service condition as a result of an overflow of TCP packets, which requires the device to be manually rebooted.CVE-2019-10948 has been assigned to this vulnerability. A CVSS v3 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-10948"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Fujifilm has stated the CR-IR 357 system can be configured with what they call Secure Host functionality. This configuration of the software instructs CR-IR 357 to ignore all network traffic other than from the IP address of the Fujifilm image acquisition console. However, this configuration prevents more than one image acquisition console to share the CR-IR 357 Reader Unit. If the user has not implemented Reader Unit sharing, they may contact Fujifilm to request Secure Host functionality be enabled. If the user has implemented Reader Unit sharing, they should contact Fujifilm to discuss available options. Fujifilm can be contacted at 888-FUJI-MED (888-385-4633). Users outside the United States should contact their Fujifilm contact.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "Fujifilm recommends that the first line of defense should be a compensating control of securing the user \u0027s network. Measures should be taken to ensure only authorized devices and personnel have access to the network. Public or guest networks should be segmented, or users should use a VLAN to segregate public traffic from the private network. Administrative and technical controls should also be implemented.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.5,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2019-10950",
"cwe": {
"id": "CWE-284",
"name": "Improper Access Control"
},
"notes": [
{
"category": "summary",
"text": "The device provides insecure telnet services that lack authentication requirements. An attacker who successfully exploits this vulnerability may be able to access the underlying operating system.CVE-2019-10950 has been assigned to this vulnerability. A CVSS v3 base score of 9.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-10950"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Fujifilm has stated the CR-IR 357 system can be configured with what they call Secure Host functionality. This configuration of the software instructs CR-IR 357 to ignore all network traffic other than from the IP address of the Fujifilm image acquisition console. However, this configuration prevents more than one image acquisition console to share the CR-IR 357 Reader Unit. If the user has not implemented Reader Unit sharing, they may contact Fujifilm to request Secure Host functionality be enabled. If the user has implemented Reader Unit sharing, they should contact Fujifilm to discuss available options. Fujifilm can be contacted at 888-FUJI-MED (888-385-4633). Users outside the United States should contact their Fujifilm contact.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "Fujifilm recommends that the first line of defense should be a compensating control of securing the user \u0027s network. Measures should be taken to ensure only authorized devices and personnel have access to the network. Public or guest networks should be segmented, or users should use a VLAN to segregate public traffic from the private network. Administrative and technical controls should also be implemented.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 9.8,
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…