ICSMA-19-311-02
Vulnerability from csaf_cisa - Published: 2019-11-07 00:00 - Updated: 2019-11-07 00:00Summary
Medtronic Valleylab FT10 and FX8
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities may allow an attacker to overwrite files or remotely execute code, resulting in a remote, non-root shell on the affected products. By default, the network connections on these devices are disabled. Additionally, the Ethernet port is disabled upon reboot. However, it is known that network connectivity is often enabled.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Ireland
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Recommended Practices
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities.
{
"document": {
"acknowledgments": [
{
"organization": "Medtronic",
"summary": "reporting these vulnerabilities to CISA"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities may allow an attacker to overwrite files or remotely execute code, resulting in a remote, non-root shell on the affected products. By default, the network connections on these devices are disabled. Additionally, the Ethernet port is disabled upon reboot. However, it is known that network connectivity is often enabled.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "Ireland",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "other",
"text": "No known public exploits specifically target these vulnerabilities.",
"title": "Exploitability"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-19-311-02 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2019/icsma-19-311-02.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-19-311-02 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-02"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.fda.gov/medical-devices/digital-health/cybersecurity"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-02"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "Medtronic Valleylab FT10 and FX8",
"tracking": {
"current_release_date": "2019-11-07T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-19-311-02",
"initial_release_date": "2019-11-07T00:00:00.000000Z",
"revision_history": [
{
"date": "2019-11-07T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-19-311-02 Medtronic Valleylab FT10 and FX8"
}
],
"status": "final",
"version": "1"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c= 1.0.0",
"product": {
"name": "Valleylab FX8 Energy Platform (VLFX8GEN) software: Version 1.1.0 and below",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "Valleylab FX8 Energy Platform (VLFX8GEN) software"
},
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c= 4.0.0",
"product": {
"name": "Valleylab FT10 Energy Platform (VLFT10GEN) software: Version 4.0.0 and below",
"product_id": "CSAFPID-0002"
}
}
],
"category": "product_name",
"name": "Valleylab FT10 Energy Platform (VLFT10GEN) software"
},
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c= 3.4",
"product": {
"name": "Valleylab Exchange Client: Version 3.4 and below",
"product_id": "CSAFPID-0003"
}
}
],
"category": "product_name",
"name": "Valleylab Exchange Client"
}
],
"category": "vendor",
"name": "Medtronic"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2019-13543",
"cwe": {
"id": "CWE-798",
"name": "Use of Hard-coded Credentials"
},
"notes": [
{
"category": "summary",
"text": "The affected products use multiple sets of hard-coded credentials. If discovered, they can be used to read files on the device.CVE-2019-13543 has been assigned to this vulnerability. A CVSS v3 base score of 5.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-13543"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:N/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Software patches are currently available for the FT10 platform and will be available in early 2020 for the FX8 platform. Until these updates can be applied, Medtronic recommends to either disconnect affected products from IP networks or to segregate those networks, such that the devices are not accessible from an untrusted network (e.g., Internet). Patches can be downloaded at the following location:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "https://www.medtronic.com/covidien/en-us/support/software.html",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.medtronic.com/covidien/en-us/support/software.html"
},
{
"category": "vendor_fix",
"details": "Medtronic has released additional patient focused information, at the following location:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "https://www.medtronic.com/security",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.medtronic.com/security"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 5.8,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2019-13539",
"cwe": {
"id": "CWE-328",
"name": "Use of Weak Hash"
},
"notes": [
{
"category": "summary",
"text": "The affected products use the descrypt algorithm for OS password hashing. While interactive, network-based logons are disabled, and attackers can use the other vulnerabilities within this report to obtain local shell access and access these hashes.CVE-2019-13539 has been assigned to this vulnerability. A CVSS v3 base score of 7.0 has been calculated; the CVSS vector string is (AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-13539"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Software patches are currently available for the FT10 platform and will be available in early 2020 for the FX8 platform. Until these updates can be applied, Medtronic recommends to either disconnect affected products from IP networks or to segregate those networks, such that the devices are not accessible from an untrusted network (e.g., Internet). Patches can be downloaded at the following location:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "https://www.medtronic.com/covidien/en-us/support/software.html",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.medtronic.com/covidien/en-us/support/software.html"
},
{
"category": "vendor_fix",
"details": "Medtronic has released additional patient focused information, at the following location:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "https://www.medtronic.com/security",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.medtronic.com/security"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.0,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2019-3464",
"cwe": {
"id": "CWE-20",
"name": "Improper Input Validation"
},
"notes": [
{
"category": "summary",
"text": "The affected products use a vulnerable version of the rssh utility in order to facilitate file uploads. This could provide an attacker with administrative access to files or the ability to execute arbitrary code.CVE-2019-3464, and CVE-2019-3463 have been assigned to this vulnerability. A CVSS v3 base score of 9.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-3464"
},
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-3463"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Software patches are currently available for the FT10 platform and will be available in early 2020 for the FX8 platform. Until these updates can be applied, Medtronic recommends to either disconnect affected products from IP networks or to segregate those networks, such that the devices are not accessible from an untrusted network (e.g., Internet). Patches can be downloaded at the following location:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "https://www.medtronic.com/covidien/en-us/support/software.html",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.medtronic.com/covidien/en-us/support/software.html"
},
{
"category": "vendor_fix",
"details": "Medtronic has released additional patient focused information, at the following location:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "https://www.medtronic.com/security",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.medtronic.com/security"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 9.8,
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…