ICSMA-24-200-01
Vulnerability from csaf_cisa
Published
2024-07-18 06:00
Modified
2024-11-21 07:00
Summary
Philips Vue PACS (Update A)

Notes

Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could allow an attacker to gain access to the database, which could impact system availability and data integrity or cause a denial-of-service condition.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Netherlands
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:
Recommended Practices
Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
Recommended Practices
Locate control system networks and remote devices behind firewalls and isolating them from business networks.
Recommended Practices
When remote access is required, use more secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Recommended Practices
CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Recommended Practices
Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.
Recommended Practices
No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time. These vulnerabilities are not exploitable remotely.



{
  "document": {
    "acknowledgments": [
      {
        "organization": "TAS Health NZ",
        "summary": "reporting these vulnerabilities to Philips"
      },
      {
        "organization": "Camiel van Es",
        "summary": "reporting these vulnerabilities to Philips"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE",
        "url": "https://us-cert.cisa.gov/tlp/"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "legal_disclaimer",
        "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
        "title": "Legal Notice"
      },
      {
        "category": "summary",
        "text": "Successful exploitation of these vulnerabilities could allow an attacker to gain access to the database, which could impact system availability and data integrity or cause a denial-of-service condition.",
        "title": "Risk evaluation"
      },
      {
        "category": "other",
        "text": "Healthcare and Public Health",
        "title": "Critical infrastructure sectors"
      },
      {
        "category": "other",
        "text": "Worldwide",
        "title": "Countries/areas deployed"
      },
      {
        "category": "other",
        "text": "Netherlands",
        "title": "Company headquarters location"
      },
      {
        "category": "general",
        "text": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Locate control system networks and remote devices behind firewalls and isolating them from business networks.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "When remote access is required, use more secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time. These vulnerabilities are not exploitable remotely.",
        "title": "Recommended Practices"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "central@cisa.dhs.gov",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-24-200-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2024/icsma-24-200-01.json"
      },
      {
        "category": "self",
        "summary": "ICSA Advisory ICSMA-24-200-01 - Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-200-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/ics/alerts/ICS-ALERT-10-301-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/resources-tools/resources/ics-recommended-practices"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/sites/default/files/publications/Cybersecurity_Best_Practices_for_Industrial_Control_Systems.pdf"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/topics/industrial-control-systems"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/ics/tips/ICS-TIP-12-146-01B"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/sites/default/files/publications/emailscams0905.pdf"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/ncas/tips/ST04-014"
      }
    ],
    "title": "Philips Vue PACS (Update A)",
    "tracking": {
      "current_release_date": "2024-11-21T07:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA CSAF Generator",
          "version": "1.0.0"
        }
      },
      "id": "ICSMA-24-200-01",
      "initial_release_date": "2024-07-18T06:00:00.000000Z",
      "revision_history": [
        {
          "date": "2024-07-18T06:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "Initial Publication"
        },
        {
          "date": "2024-11-21T07:00:00.000000Z",
          "legacy_version": "Update A",
          "number": "2",
          "summary": "Update A - Vulnerabilities removed due to further analysis by Philips showing they did not affect the device or had no security impact."
        }
      ],
      "status": "final",
      "version": "2"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c12.2.8.410",
                "product": {
                  "name": "Philips Vue PACS: \u003c12.2.8.410",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "Vue PACS"
          }
        ],
        "category": "vendor",
        "name": "Philips"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2021-28165",
      "cwe": {
        "id": "CWE-770",
        "name": "Allocation of Resources Without Limits or Throttling"
      },
      "notes": [
        {
          "category": "summary",
          "text": "Attackers can exploit this vulnerability by making numerous requests or sending large amounts of data to the application, leading to resource exhaustion (e.g., memory, CPU), which can cause the application to crash or become unresponsive. This vulnerability does not expose patient data or allow for its modification. It allows an attacker, with access to the hospital\u0027s private network, which is protected by security controls (e.g., firewalls, VPNs), to send messages to the server, leading to potential CPU overload and a denial-of-service (DoS) condition. No response is sent back to the attacker, and patient information remains secure.",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "www.cve.org",
          "url": "https://www.cve.org/CVERecord?id=CVE-2021-28165"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.1#CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H"
        },
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "https://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-28165"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/4.0#CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Philips recommends the following mitigations:",
          "product_ids": [
            "CSAFPID-0001"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For CVE-2021-28165, Philips recommends configuring the Vue PACS environment per D000763414 \u2013 Vue_PACS_12_Ports_Protocols_Services_Guide available on Incenter. Vue PACS version 12.2.8.410* released in October 2023 prevents this vulnerability.",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://incenter.medical.philips.com/Default.aspx?tabid=867"
        },
        {
          "category": "vendor_fix",
          "details": "For managed services customers, new releases will be made available upon resource availability. Releases are subject to country specific regulations. Users with questions regarding their specific Philips Vue PACS installations and new release eligibility should contact their local Philips sales representative or submit a request in the Philips Informatics Support portal.",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.informatics.support.philips.com/csm"
        },
        {
          "category": "mitigation",
          "details": "Refer to the Philips advisory for more details.",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.philips.com/productsecurity"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
            "version": "3.1"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2023-40704",
      "cwe": {
        "id": "CWE-1392",
        "name": "Use of Default Credentials"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The product does not require unique and complex passwords to be created during installation. Using Philips\u0027s default password could jeopardize the PACS system if the password was hacked or leaked. An attacker could gain access to the database impacting system availability and data integrity.",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "www.cve.org",
          "url": "https://www.cve.org/CVERecord?id=CVE-2023-40704"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.1#CVSS:3.1/AV:A/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H"
        },
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "https://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2023-40704"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/4.0#CVSS:4.0/AV:L/AC:H/AT:N/PR:H/UI:N/VC:H/VI:L/VA:H/SC:N/SI:N/SA:N"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Philips recommends the following mitigations:",
          "product_ids": [
            "CSAFPID-0001"
          ]
        },
        {
          "category": "mitigation",
          "details": "For CVE-2023-40704, Philips recommends no action needed due to low risk of exploitability, but customers can request that Philips update database password(s).",
          "product_ids": [
            "CSAFPID-0001"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For managed services customers, new releases will be made available upon resource availability. Releases are subject to country specific regulations. Users with questions regarding their specific Philips Vue PACS installations and new release eligibility should contact their local Philips sales representative or submit a request in the Philips Informatics Support portal.",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.informatics.support.philips.com/csm"
        },
        {
          "category": "mitigation",
          "details": "Refer to the Philips advisory for more details.",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.philips.com/productsecurity"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 6.8,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.1/AV:A/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H",
            "version": "3.1"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ]
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.