CVE-2019-13531 (GCVE-0-2019-13531)

Vulnerability from cvelistv5 – Published: 2019-11-08 19:46 – Updated: 2025-05-22 18:37
VLAI?
Summary
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.
CWE
  • CWE-287 - Improper Authentication
Assigner
Impacted products
Credits
Medtronic reported these vulnerabilities to CISA.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T23:57:39.240Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Valleylab FT10 Energy Platform (VLFT10GEN)",
          "vendor": "Medtronic",
          "versions": [
            {
              "lessThanOrEqual": "2.1.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "2.0.3",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States)",
          "vendor": "Medtronic",
          "versions": [
            {
              "lessThanOrEqual": "1.20.2",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Medtronic reported these vulnerabilities to CISA."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eIn Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.\u003c/p\u003e"
            }
          ],
          "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 4.8,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "NONE",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-287",
              "description": "CWE-287 Improper Authentication",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-05-22T18:37:04.526Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html"
        },
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eA software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/covidien/en-us/support/software.html\"\u003eMedtronic\u003c/a\u003e\u0026nbsp;to obtain the software patch.\u003c/p\u003e\u003cp\u003eMedtronic has released additional patient focused information at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact  https://www.medtronic.com/security"
        }
      ],
      "source": {
        "advisory": "ICSMA-19-311-01",
        "discovery": "INTERNAL"
      },
      "title": "Medtronic Valleylab FT10 and LS10 Improper Authentication",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "ID": "CVE-2019-13531",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Valleylab FT10 Energy Platform (VLFT10GEN)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "version 2.1.0 and lower"
                          },
                          {
                            "version_value": "version 2.0.3 and lower"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "version 1.20.2 and lower"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Medtronic"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "IMPROPER AUTHENTICATION CWE-287"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01",
              "refsource": "MISC",
              "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2019-13531",
    "datePublished": "2019-11-08T19:46:45",
    "dateReserved": "2019-07-11T00:00:00",
    "dateUpdated": "2025-05-22T18:37:04.526Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "vulnerability-lookup:meta": {
    "fkie_nvd": {
      "configurations": "[{\"operator\": \"AND\", \"nodes\": [{\"operator\": \"OR\", \"negate\": false, \"cpeMatch\": [{\"vulnerable\": true, \"criteria\": \"cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.0.3:*:*:*:*:*:*:*\", \"matchCriteriaId\": \"89DAFB01-536A-4966-8EAC-5D6CED32CC63\"}, {\"vulnerable\": true, \"criteria\": \"cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.1.0:*:*:*:*:*:*:*\", \"matchCriteriaId\": \"31DF6D35-968D-4E8D-A8F5-578B4B9C2BE9\"}]}, {\"operator\": \"OR\", \"negate\": false, \"cpeMatch\": [{\"vulnerable\": false, \"criteria\": \"cpe:2.3:h:medtronic:valleylab_ft10_energy_platform:-:*:*:*:*:*:*:*\", \"matchCriteriaId\": \"164230CB-E2BF-447F-8537-C9401FA0CC09\"}]}]}, {\"operator\": \"AND\", \"nodes\": [{\"operator\": \"OR\", \"negate\": false, \"cpeMatch\": [{\"vulnerable\": true, \"criteria\": \"cpe:2.3:o:medtronic:valleylab_ls10_energy_platform_firmware:*:*:*:*:*:*:*:*\", \"versionEndIncluding\": \"1.20.2\", \"matchCriteriaId\": \"EEAE74E9-49AC-4B77-8831-11EA09B204CB\"}]}, {\"operator\": \"OR\", \"negate\": false, \"cpeMatch\": [{\"vulnerable\": false, \"criteria\": \"cpe:2.3:h:medtronic:valleylab_ls10_energy_platform:-:*:*:*:*:*:*:*\", \"matchCriteriaId\": \"82294F1E-0BD0-4F55-A732-09B68AA37624\"}]}]}]",
      "descriptions": "[{\"lang\": \"en\", \"value\": \"In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.\"}, {\"lang\": \"es\", \"value\": \"En Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) versi\\u00f3n 2.1.0 y versiones anteriores y versi\\u00f3n 2.0.3 y versiones atneriores, y Valleylab LS10 Energy Platform (VLLS10GEN \\u2014no disponible en los Estados Unidos) versi\\u00f3n 1.20.2 y versiones anteriores, se utiliz\\u00f3 el mecanismo de seguridad RFID para la autenticaci\\u00f3n entre la plataforma de energ\\u00eda FT10 / LS10 y los instrumentos se pueden omitir, lo que permite que los instrumentos no aut\\u00e9nticos se conecten al generador.\"}]",
      "id": "CVE-2019-13531",
      "lastModified": "2024-11-21T04:25:05.343",
      "metrics": "{\"cvssMetricV31\": [{\"source\": \"nvd@nist.gov\", \"type\": \"Primary\", \"cvssData\": {\"version\": \"3.1\", \"vectorString\": \"CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N\", \"baseScore\": 4.6, \"baseSeverity\": \"MEDIUM\", \"attackVector\": \"PHYSICAL\", \"attackComplexity\": \"LOW\", \"privilegesRequired\": \"NONE\", \"userInteraction\": \"NONE\", \"scope\": \"UNCHANGED\", \"confidentialityImpact\": \"NONE\", \"integrityImpact\": \"HIGH\", \"availabilityImpact\": \"NONE\"}, \"exploitabilityScore\": 0.9, \"impactScore\": 3.6}], \"cvssMetricV2\": [{\"source\": \"nvd@nist.gov\", \"type\": \"Primary\", \"cvssData\": {\"version\": \"2.0\", \"vectorString\": \"AV:L/AC:L/Au:N/C:N/I:P/A:N\", \"baseScore\": 2.1, \"accessVector\": \"LOCAL\", \"accessComplexity\": \"LOW\", \"authentication\": \"NONE\", \"confidentialityImpact\": \"NONE\", \"integrityImpact\": \"PARTIAL\", \"availabilityImpact\": \"NONE\"}, \"baseSeverity\": \"LOW\", \"exploitabilityScore\": 3.9, \"impactScore\": 2.9, \"acInsufInfo\": false, \"obtainAllPrivilege\": false, \"obtainUserPrivilege\": false, \"obtainOtherPrivilege\": false, \"userInteractionRequired\": false}]}",
      "published": "2019-11-08T20:15:10.570",
      "references": "[{\"url\": \"https://www.us-cert.gov/ics/advisories/icsma-19-311-01\", \"source\": \"ics-cert@hq.dhs.gov\", \"tags\": [\"Third Party Advisory\", \"US Government Resource\"]}, {\"url\": \"https://www.us-cert.gov/ics/advisories/icsma-19-311-01\", \"source\": \"af854a3a-2127-422b-91ae-364da2661108\", \"tags\": [\"Third Party Advisory\", \"US Government Resource\"]}]",
      "sourceIdentifier": "ics-cert@hq.dhs.gov",
      "vulnStatus": "Modified",
      "weaknesses": "[{\"source\": \"ics-cert@hq.dhs.gov\", \"type\": \"Secondary\", \"description\": [{\"lang\": \"en\", \"value\": \"CWE-287\"}]}, {\"source\": \"nvd@nist.gov\", \"type\": \"Primary\", \"description\": [{\"lang\": \"en\", \"value\": \"NVD-CWE-noinfo\"}]}]"
    },
    "nvd": "{\"cve\":{\"id\":\"CVE-2019-13531\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2019-11-08T20:15:10.570\",\"lastModified\":\"2025-05-22T19:15:22.673\",\"vulnStatus\":\"Modified\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.\"},{\"lang\":\"es\",\"value\":\"En Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) versi\u00f3n 2.1.0 y versiones anteriores y versi\u00f3n 2.0.3 y versiones atneriores, y Valleylab LS10 Energy Platform (VLLS10GEN \u2014no disponible en los Estados Unidos) versi\u00f3n 1.20.2 y versiones anteriores, se utiliz\u00f3 el mecanismo de seguridad RFID para la autenticaci\u00f3n entre la plataforma de energ\u00eda FT10 / LS10 y los instrumentos se pueden omitir, lo que permite que los instrumentos no aut\u00e9nticos se conecten al generador.\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L\",\"baseScore\":4.8,\"baseSeverity\":\"MEDIUM\",\"attackVector\":\"PHYSICAL\",\"attackComplexity\":\"HIGH\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"LOW\"},\"exploitabilityScore\":0.5,\"impactScore\":4.2},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N\",\"baseScore\":4.6,\"baseSeverity\":\"MEDIUM\",\"attackVector\":\"PHYSICAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"NONE\"},\"exploitabilityScore\":0.9,\"impactScore\":3.6}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:L/AC:L/Au:N/C:N/I:P/A:N\",\"baseScore\":2.1,\"accessVector\":\"LOCAL\",\"accessComplexity\":\"LOW\",\"authentication\":\"NONE\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"PARTIAL\",\"availabilityImpact\":\"NONE\"},\"baseSeverity\":\"LOW\",\"exploitabilityScore\":3.9,\"impactScore\":2.9,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":false}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-287\"}]},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"NVD-CWE-noinfo\"}]}],\"configurations\":[{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.0.3:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"89DAFB01-536A-4966-8EAC-5D6CED32CC63\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.1.0:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"31DF6D35-968D-4E8D-A8F5-578B4B9C2BE9\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:valleylab_ft10_energy_platform:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"164230CB-E2BF-447F-8537-C9401FA0CC09\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:valleylab_ls10_energy_platform_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"1.20.2\",\"matchCriteriaId\":\"EEAE74E9-49AC-4B77-8831-11EA09B204CB\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:valleylab_ls10_energy_platform:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"82294F1E-0BD0-4F55-A732-09B68AA37624\"}]}]}],\"references\":[{\"url\":\"https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html\",\"source\":\"ics-cert@hq.dhs.gov\"},{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01\",\"source\":\"ics-cert@hq.dhs.gov\"},{\"url\":\"https://www.us-cert.gov/ics/advisories/icsma-19-311-01\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Third Party Advisory\",\"US Government Resource\"]}]}}"
  }
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.


Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…