ICSMA-19-311-01

Vulnerability from csaf_cisa - Published: 2019-11-07 00:00 - Updated: 2019-11-07 00:00
Summary
Medtronic Valleylab FT10 and LS10

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities may allow an attacker to connect inauthentic instruments to the affected products by spoofing RFID security mechanisms. This may lead to a loss of performance integrity and platform availability due to incorrect identification of instrument and associated parameters.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Ireland
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Recommended Practices
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies. Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Recommended Practices
These vulnerabilities are not exploitable remotely.

{
  "document": {
    "acknowledgments": [
      {
        "organization": "Medtronic",
        "summary": "reporting these vulnerabilities to CISA"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE",
        "url": "https://us-cert.cisa.gov/tlp/"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "legal_disclaimer",
        "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
        "title": "Legal Notice"
      },
      {
        "category": "summary",
        "text": "Successful exploitation of these vulnerabilities may allow an attacker to connect inauthentic instruments to the affected products by spoofing RFID security mechanisms. This may lead to a loss of performance integrity and platform availability due to incorrect identification of instrument and associated parameters.",
        "title": "Risk evaluation"
      },
      {
        "category": "other",
        "text": "Healthcare and Public Health",
        "title": "Critical infrastructure sectors"
      },
      {
        "category": "other",
        "text": "Worldwide",
        "title": "Countries/areas deployed"
      },
      {
        "category": "other",
        "text": "Ireland",
        "title": "Company headquarters location"
      },
      {
        "category": "general",
        "text": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "These vulnerabilities are not exploitable remotely.",
        "title": "Recommended Practices"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-19-311-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2019/icsma-19-311-01.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-19-311-01 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.fda.gov/medical-devices/digital-health/cybersecurity"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.us-cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.us-cert.gov/ics/tips/ICS-TIP-12-146-01B"
      }
    ],
    "title": "Medtronic Valleylab FT10 and LS10",
    "tracking": {
      "current_release_date": "2019-11-07T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA CSAF Generator",
          "version": "1.0.0"
        }
      },
      "id": "ICSMA-19-311-01",
      "initial_release_date": "2019-11-07T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2019-11-07T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSMA-19-311-01 Medtronic Valleylab FT10 and LS10"
        }
      ],
      "status": "final",
      "version": "1"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c= 1.20.2",
                "product": {
                  "name": "Valleylab FT10 Valleylab LS10: Version 1.20.2 and lower",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "Valleylab LS10 Energy Platform (VLLS10GEN)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c= 2.1.0",
                "product": {
                  "name": "Valleylab FT10 Energy Platform (VLFT10GEN): \u003c= 2.1.0",
                  "product_id": "CSAFPID-0002"
                }
              }
            ],
            "category": "product_name",
            "name": "Valleylab FT10 Energy Platform (VLFT10GEN)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c= 2.0.3",
                "product": {
                  "name": "Valleylab FT10 Valleylab LS10: Version 2.0.3 and lower",
                  "product_id": "CSAFPID-0003"
                }
              }
            ],
            "category": "product_name",
            "name": "Valleylab FT10 Energy Platform (VLFT10GEN)"
          }
        ],
        "category": "vendor",
        "name": "Medtronic"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2019-13531",
      "cwe": {
        "id": "CWE-287",
        "name": "Improper Authentication"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.CVE-2019-13531 has been assigned to this vulnerability. A CVSS v3 base score of 4.8 has been calculated; the CVSS vector string is (AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-13531"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact Medtronic to obtain the software patch.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "https://www.medtronic.com/covidien/en-us/support/software.html"
        },
        {
          "category": "vendor_fix",
          "details": "Medtronic has released additional patient focused information at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.medtronic.com/security",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "https://www.medtronic.com/security"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 4.8,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2019-13535",
      "cwe": {
        "id": "CWE-693",
        "name": "Protection Mechanism Failure"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.CVE-2019-13535 has been assigned to this vulnerability. A CVSS v3 base score of 4.6 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-13535"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact Medtronic to obtain the software patch.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "https://www.medtronic.com/covidien/en-us/support/software.html"
        },
        {
          "category": "vendor_fix",
          "details": "Medtronic has released additional patient focused information at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.medtronic.com/security",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "https://www.medtronic.com/security"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 4.6,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        }
      ]
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.


Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…