CVE-2025-54304 (GCVE-0-2025-54304)
Vulnerability from cvelistv5 – Published: 2025-12-04 00:00 – Updated: 2025-12-04 14:56
VLAI?
Summary
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
Severity ?
No CVSS data available.
CWE
- n/a
Assigner
References
{
"containers": {
"cna": {
"affected": [
{
"product": "n/a",
"vendor": "n/a",
"versions": [
{
"status": "affected",
"version": "n/a"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer."
}
],
"problemTypes": [
{
"descriptions": [
{
"description": "n/a",
"lang": "en",
"type": "text"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-12-04T14:56:05.054Z",
"orgId": "8254265b-2729-46b6-b9e3-3dfca2d5bfca",
"shortName": "mitre"
},
"references": [
{
"url": "https://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2Sys_UG.pdf"
},
{
"url": "https://www.thermofisher.com/order/catalog/product/4474779"
},
{
"url": "https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2_and_Torrent_Suite_Software.pdf"
}
]
}
},
"cveMetadata": {
"assignerOrgId": "8254265b-2729-46b6-b9e3-3dfca2d5bfca",
"assignerShortName": "mitre",
"cveId": "CVE-2025-54304",
"datePublished": "2025-12-04T00:00:00.000Z",
"dateReserved": "2025-07-18T00:00:00.000Z",
"dateUpdated": "2025-12-04T14:56:05.054Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"vulnerability-lookup:meta": {
"nvd": "{\"cve\":{\"id\":\"CVE-2025-54304\",\"sourceIdentifier\":\"cve@mitre.org\",\"published\":\"2025-12-04T15:15:58.773\",\"lastModified\":\"2025-12-04T17:15:08.283\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.\"}],\"metrics\":{},\"references\":[{\"url\":\"https://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2Sys_UG.pdf\",\"source\":\"cve@mitre.org\"},{\"url\":\"https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2_and_Torrent_Suite_Software.pdf\",\"source\":\"cve@mitre.org\"},{\"url\":\"https://www.thermofisher.com/order/catalog/product/4474779\",\"source\":\"cve@mitre.org\"}]}}"
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…