Vulnerability from fkie_nvd - Published: 2018-07-03 01:29 - Updated: 2025-05-22 18:15
Severity ?
Summary
Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
References
Impacted products
| Vendor | Product | Version | |
|---|---|---|---|
| medtronic | 2090_carelink_programmer_firmware | * | |
| medtronic | 2090_carelink_programmer | - |
{
"configurations": [
{
"nodes": [
{
"cpeMatch": [
{
"criteria": "cpe:2.3:o:medtronic:2090_carelink_programmer_firmware:*:*:*:*:*:*:*:*",
"matchCriteriaId": "884B08DC-B6F9-4B34-9679-61C256437DF9",
"vulnerable": true
}
],
"negate": false,
"operator": "OR"
},
{
"cpeMatch": [
{
"criteria": "cpe:2.3:h:medtronic:2090_carelink_programmer:-:*:*:*:*:*:*:*",
"matchCriteriaId": "215ED381-B433-4D05-97AD-4E7287E5820D",
"vulnerable": false
}
],
"negate": false,
"operator": "OR"
}
],
"operator": "AND"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "Medtronic 2090 CareLink Programmer \n\nuses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications."
},
{
"lang": "es",
"value": "Medtronic 2090 CareLink Programmer en todas sus versiones emplea una conexi\u00f3n de red privada virtual para descargar actualizaciones de forma segura. El producto no verifica si sigue conectado a esta red virtual privada antes de descargar actualizaciones. Un atacante con acceso de red local al programador podr\u00eda influenciar estas comunicaciones."
}
],
"id": "CVE-2018-10596",
"lastModified": "2025-05-22T18:15:21.490",
"metrics": {
"cvssMetricV2": [
{
"acInsufInfo": false,
"baseSeverity": "MEDIUM",
"cvssData": {
"accessComplexity": "LOW",
"accessVector": "ADJACENT_NETWORK",
"authentication": "SINGLE",
"availabilityImpact": "PARTIAL",
"baseScore": 5.2,
"confidentialityImpact": "PARTIAL",
"integrityImpact": "PARTIAL",
"vectorString": "AV:A/AC:L/Au:S/C:P/I:P/A:P",
"version": "2.0"
},
"exploitabilityScore": 5.1,
"impactScore": 6.4,
"obtainAllPrivilege": false,
"obtainOtherPrivilege": false,
"obtainUserPrivilege": false,
"source": "nvd@nist.gov",
"type": "Primary",
"userInteractionRequired": false
}
],
"cvssMetricV30": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 8.0,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "LOW",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"exploitabilityScore": 2.1,
"impactScore": 5.9,
"source": "nvd@nist.gov",
"type": "Primary"
}
],
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "HIGH",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "LOW",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"exploitabilityScore": 1.2,
"impactScore": 5.9,
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
}
]
},
"published": "2018-07-03T01:29:00.487",
"references": [
{
"source": "ics-cert@hq.dhs.gov",
"url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html"
},
{
"source": "ics-cert@hq.dhs.gov",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"vulnStatus": "Modified",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-923"
}
],
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"description": [
{
"lang": "en",
"value": "CWE-200"
}
],
"source": "nvd@nist.gov",
"type": "Primary"
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…