FKIE_CVE-2024-48966

Vulnerability from fkie_nvd - Published: 2024-11-14 22:15 - Updated: 2024-11-15 13:58
Severity ?
Summary
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Impacted products
Vendor Product Version

{
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "The software tools used by service personnel to test \u0026 calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator\u0027s settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance."
    },
    {
      "lang": "es",
      "value": "Las herramientas de software que utiliza el personal de servicio para probar y calibrar el respirador no admiten la autenticaci\u00f3n de usuarios. Un atacante con acceso a la PC de servicio donde est\u00e1n instaladas las herramientas podr\u00eda obtener informaci\u00f3n de diagn\u00f3stico a trav\u00e9s de la herramienta de prueba o manipular la configuraci\u00f3n del respirador y el software integrado a trav\u00e9s de la herramienta de calibraci\u00f3n, sin tener que autenticarse en ninguna de las herramientas. Esto podr\u00eda dar lugar a una divulgaci\u00f3n no autorizada de informaci\u00f3n o tener efectos no deseados en la configuraci\u00f3n y el rendimiento del dispositivo."
    }
  ],
  "id": "CVE-2024-48966",
  "lastModified": "2024-11-15T13:58:08.913",
  "metrics": {
    "cvssMetricV31": [
      {
        "cvssData": {
          "attackComplexity": "LOW",
          "attackVector": "NETWORK",
          "availabilityImpact": "HIGH",
          "baseScore": 10.0,
          "baseSeverity": "CRITICAL",
          "confidentialityImpact": "HIGH",
          "integrityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "scope": "CHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
          "version": "3.1"
        },
        "exploitabilityScore": 3.9,
        "impactScore": 6.0,
        "source": "productsecurity@baxter.com",
        "type": "Secondary"
      }
    ]
  },
  "published": "2024-11-14T22:15:17.727",
  "references": [
    {
      "source": "productsecurity@baxter.com",
      "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"
    }
  ],
  "sourceIdentifier": "productsecurity@baxter.com",
  "vulnStatus": "Awaiting Analysis",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-306"
        }
      ],
      "source": "productsecurity@baxter.com",
      "type": "Secondary"
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.


Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…