FKIE_CVE-2024-48973
Vulnerability from fkie_nvd - Published: 2024-11-14 22:15 - Updated: 2026-04-15 00:35
Severity
Summary
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
References
Impacted products
| Vendor | Product | Version |
|---|
{
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "The debug port on the ventilator\u0027s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance."
},
{
"lang": "es",
"value": "El puerto de depuraci\u00f3n de la interfaz serial del respirador est\u00e1 habilitado de manera predeterminada. Esto podr\u00eda permitir que un atacante env\u00ede y reciba mensajes a trav\u00e9s del puerto de depuraci\u00f3n (que no est\u00e1n cifrados; consulte 3.2.1) que resulten en la divulgaci\u00f3n no autorizada de informaci\u00f3n o tengan impactos no deseados en la configuraci\u00f3n y el rendimiento del dispositivo."
}
],
"id": "CVE-2024-48973",
"lastModified": "2026-04-15T00:35:42.020",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 9.3,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "CHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
"version": "3.1"
},
"exploitabilityScore": 2.5,
"impactScore": 6.0,
"source": "productsecurity@baxter.com",
"type": "Secondary"
}
]
},
"published": "2024-11-14T22:15:18.510",
"references": [
{
"source": "productsecurity@baxter.com",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"
}
],
"sourceIdentifier": "productsecurity@baxter.com",
"vulnStatus": "Deferred",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-1263"
}
],
"source": "productsecurity@baxter.com",
"type": "Secondary"
}
]
}
Loading…
Loading…
Experimental. This forecast is provided for visualization only and may change without notice. Do not use it for operational decisions.
Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.
Sightings
| Author | Source | Type | Date | Other |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…