CVE-2024-48973 (GCVE-0-2024-48973)
Vulnerability from cvelistv5 – Published: 2024-11-14 21:24 – Updated: 2024-11-18 15:33
VLAI?
Summary
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Severity ?
9.3 (Critical)
CWE
- CWE-1263 - Improper Physical Access Control
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Baxter | Life2000 Ventilation System |
Affected:
06.08.00.00 and prior
|
{
"containers": {
"adp": [
{
"affected": [
{
"cpes": [
"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"
],
"defaultStatus": "unaffected",
"product": "life2000_ventilator_firmware",
"vendor": "baxter",
"versions": [
{
"lessThanOrEqual": "06.08.00.00",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"metrics": [
{
"other": {
"content": {
"id": "CVE-2024-48973",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2024-11-18T15:33:07.996857Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2024-11-18T15:33:44.924Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "Life2000 Ventilation System",
"vendor": "Baxter",
"versions": [
{
"status": "affected",
"version": "06.08.00.00 and prior"
}
]
}
],
"datePublic": "2024-11-14T21:21:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "The debug port on the ventilator\u0027s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.\u003cbr\u003e"
}
],
"value": "The debug port on the ventilator\u0027s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance."
}
],
"impacts": [
{
"capecId": "CAPEC-117",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-117 Interception"
}
]
},
{
"capecId": "CAPEC-441",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-441 Malicious Logic Insertion"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 9.3,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "CHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-1263",
"description": "CWE-1263 Improper Physical Access Control",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2024-11-14T21:49:44.756Z",
"orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
"shortName": "Baxter"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "Debug port on Life2000 Ventilator serial interface is enabled by default",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
"assignerShortName": "Baxter",
"cveId": "CVE-2024-48973",
"datePublished": "2024-11-14T21:24:14.099Z",
"dateReserved": "2024-10-10T19:24:41.495Z",
"dateUpdated": "2024-11-18T15:33:44.924Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1",
"vulnerability-lookup:meta": {
"fkie_nvd": {
"descriptions": "[{\"lang\": \"en\", \"value\": \"The debug port on the ventilator\u0027s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.\"}, {\"lang\": \"es\", \"value\": \"El puerto de depuraci\\u00f3n de la interfaz serial del respirador est\\u00e1 habilitado de manera predeterminada. Esto podr\\u00eda permitir que un atacante env\\u00ede y reciba mensajes a trav\\u00e9s del puerto de depuraci\\u00f3n (que no est\\u00e1n cifrados; consulte 3.2.1) que resulten en la divulgaci\\u00f3n no autorizada de informaci\\u00f3n o tengan impactos no deseados en la configuraci\\u00f3n y el rendimiento del dispositivo.\"}]",
"id": "CVE-2024-48973",
"lastModified": "2024-11-15T13:58:08.913",
"metrics": "{\"cvssMetricV31\": [{\"source\": \"productsecurity@baxter.com\", \"type\": \"Secondary\", \"cvssData\": {\"version\": \"3.1\", \"vectorString\": \"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\", \"baseScore\": 9.3, \"baseSeverity\": \"CRITICAL\", \"attackVector\": \"LOCAL\", \"attackComplexity\": \"LOW\", \"privilegesRequired\": \"NONE\", \"userInteraction\": \"NONE\", \"scope\": \"CHANGED\", \"confidentialityImpact\": \"HIGH\", \"integrityImpact\": \"HIGH\", \"availabilityImpact\": \"HIGH\"}, \"exploitabilityScore\": 2.5, \"impactScore\": 6.0}]}",
"published": "2024-11-14T22:15:18.510",
"references": "[{\"url\": \"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\", \"source\": \"productsecurity@baxter.com\"}]",
"sourceIdentifier": "productsecurity@baxter.com",
"vulnStatus": "Awaiting Analysis",
"weaknesses": "[{\"source\": \"productsecurity@baxter.com\", \"type\": \"Secondary\", \"description\": [{\"lang\": \"en\", \"value\": \"CWE-1263\"}]}]"
},
"nvd": "{\"cve\":{\"id\":\"CVE-2024-48973\",\"sourceIdentifier\":\"productsecurity@baxter.com\",\"published\":\"2024-11-14T22:15:18.510\",\"lastModified\":\"2024-11-15T13:58:08.913\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"The debug port on the ventilator\u0027s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.\"},{\"lang\":\"es\",\"value\":\"El puerto de depuraci\u00f3n de la interfaz serial del respirador est\u00e1 habilitado de manera predeterminada. Esto podr\u00eda permitir que un atacante env\u00ede y reciba mensajes a trav\u00e9s del puerto de depuraci\u00f3n (que no est\u00e1n cifrados; consulte 3.2.1) que resulten en la divulgaci\u00f3n no autorizada de informaci\u00f3n o tengan impactos no deseados en la configuraci\u00f3n y el rendimiento del dispositivo.\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\",\"baseScore\":9.3,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"LOCAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":2.5,\"impactScore\":6.0}]},\"weaknesses\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-1263\"}]}],\"references\":[{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\",\"source\":\"productsecurity@baxter.com\"}]}}",
"vulnrichment": {
"containers": "{\"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2024-48973\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"no\"}, {\"Technical Impact\": \"total\"}], \"version\": \"2.0.3\", \"timestamp\": \"2024-11-18T15:33:07.996857Z\"}}}], \"affected\": [{\"cpes\": [\"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*\"], \"vendor\": \"baxter\", \"product\": \"life2000_ventilator_firmware\", \"versions\": [{\"status\": \"affected\", \"version\": \"0\", \"versionType\": \"custom\", \"lessThanOrEqual\": \"06.08.00.00\"}], \"defaultStatus\": \"unaffected\"}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2024-11-18T15:33:39.812Z\"}}], \"cna\": {\"title\": \"Debug port on Life2000 Ventilator serial interface is enabled by default\", \"source\": {\"discovery\": \"UNKNOWN\"}, \"impacts\": [{\"capecId\": \"CAPEC-117\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-117 Interception\"}]}, {\"capecId\": \"CAPEC-441\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-441 Malicious Logic Insertion\"}]}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"CHANGED\", \"version\": \"3.1\", \"baseScore\": 9.3, \"attackVector\": \"LOCAL\", \"baseSeverity\": \"CRITICAL\", \"vectorString\": \"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\", \"integrityImpact\": \"HIGH\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"HIGH\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Baxter\", \"product\": \"Life2000 Ventilation System\", \"versions\": [{\"status\": \"affected\", \"version\": \"06.08.00.00 and prior\"}], \"defaultStatus\": \"unaffected\"}], \"datePublic\": \"2024-11-14T21:21:00.000Z\", \"references\": [{\"url\": \"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\"}], \"x_generator\": {\"engine\": \"Vulnogram 0.2.0\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"The debug port on the ventilator\u0027s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"The debug port on the ventilator\u0027s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.\u003cbr\u003e\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-1263\", \"description\": \"CWE-1263 Improper Physical Access Control\"}]}], \"providerMetadata\": {\"orgId\": \"dba971b9-eb30-4121-91e1-3b45611354aa\", \"shortName\": \"Baxter\", \"dateUpdated\": \"2024-11-14T21:49:44.756Z\"}}}",
"cveMetadata": "{\"cveId\": \"CVE-2024-48973\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2024-11-18T15:33:44.924Z\", \"dateReserved\": \"2024-10-10T19:24:41.495Z\", \"assignerOrgId\": \"dba971b9-eb30-4121-91e1-3b45611354aa\", \"datePublished\": \"2024-11-14T21:24:14.099Z\", \"assignerShortName\": \"Baxter\"}",
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…