GSD-2017-12725
Vulnerability from gsd - Updated: 2023-12-13 01:21Details
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
Aliases
Aliases
{
"GSD": {
"alias": "CVE-2017-12725",
"description": "A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.",
"id": "GSD-2017-12725"
},
"gsd": {
"metadata": {
"exploitCode": "unknown",
"remediation": "unknown",
"reportConfidence": "confirmed",
"type": "vulnerability"
},
"osvSchema": {
"aliases": [
"CVE-2017-12725"
],
"details": "A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.",
"id": "GSD-2017-12725",
"modified": "2023-12-13T01:21:03.330235Z",
"schema_version": "1.4.0"
}
},
"namespaces": {
"cve.org": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"ID": "CVE-2017-12725",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump",
"version": {
"version_data": [
{
"version_value": "Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump"
}
]
}
}
]
},
"vendor_name": "n/a"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Use of Hard-coded Credentials"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A"
},
{
"name": "100665",
"refsource": "BID",
"url": "http://www.securityfocus.com/bid/100665"
}
]
}
},
"nvd.nist.gov": {
"configurations": {
"CVE_data_version": "4.0",
"nodes": [
{
"children": [
{
"children": [],
"cpe_match": [
{
"cpe23Uri": "cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:*",
"cpe_name": [],
"vulnerable": true
},
{
"cpe23Uri": "cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.6:*:*:*:*:*:*:*",
"cpe_name": [],
"vulnerable": true
},
{
"cpe23Uri": "cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.5:*:*:*:*:*:*:*",
"cpe_name": [],
"vulnerable": true
}
],
"operator": "OR"
},
{
"children": [],
"cpe_match": [
{
"cpe23Uri": "cpe:2.3:h:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:-:*:*:*:*:*:*:*",
"cpe_name": [],
"vulnerable": false
}
],
"operator": "OR"
}
],
"cpe_match": [],
"operator": "AND"
}
]
},
"cve": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"ID": "CVE-2017-12725"
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "en",
"value": "A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "en",
"value": "CWE-798"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A",
"refsource": "MISC",
"tags": [
"Third Party Advisory",
"US Government Resource"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A"
},
{
"name": "100665",
"refsource": "BID",
"tags": [
"Third Party Advisory",
"VDB Entry"
],
"url": "http://www.securityfocus.com/bid/100665"
}
]
}
},
"impact": {
"baseMetricV2": {
"cvssV2": {
"accessComplexity": "MEDIUM",
"accessVector": "NETWORK",
"authentication": "NONE",
"availabilityImpact": "PARTIAL",
"baseScore": 6.8,
"confidentialityImpact": "PARTIAL",
"integrityImpact": "PARTIAL",
"vectorString": "AV:N/AC:M/Au:N/C:P/I:P/A:P",
"version": "2.0"
},
"exploitabilityScore": 8.6,
"impactScore": 6.4,
"obtainAllPrivilege": false,
"obtainOtherPrivilege": false,
"obtainUserPrivilege": false,
"severity": "MEDIUM",
"userInteractionRequired": false
},
"baseMetricV3": {
"cvssV3": {
"attackComplexity": "HIGH",
"attackVector": "NETWORK",
"availabilityImpact": "LOW",
"baseScore": 5.6,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L",
"version": "3.0"
},
"exploitabilityScore": 2.2,
"impactScore": 3.4
}
},
"lastModifiedDate": "2018-03-02T14:40Z",
"publishedDate": "2018-02-15T10:29Z"
}
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…