GSD-2022-38069
Vulnerability from gsd - Updated: 2023-12-13 01:19Details
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
Aliases
Aliases
{
"GSD": {
"alias": "CVE-2022-38069",
"description": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters",
"id": "GSD-2022-38069"
},
"gsd": {
"metadata": {
"exploitCode": "unknown",
"remediation": "unknown",
"reportConfidence": "confirmed",
"type": "vulnerability"
},
"osvSchema": {
"aliases": [
"CVE-2022-38069"
],
"details": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters",
"id": "GSD-2022-38069",
"modified": "2023-12-13T01:19:22.144497Z",
"schema_version": "1.4.0"
}
},
"namespaces": {
"cve.org": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2022-09-01T17:00:00.000Z",
"ID": "CVE-2022-38069",
"STATE": "PUBLIC",
"TITLE": "Contec Health CMS8000"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_name": "All",
"version_value": "All"
}
]
}
}
]
},
"vendor_name": "Contec Health"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Level Nine reported these vulnerabilities to CISA."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters"
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 4.3,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-798 Use of Hard-coded Credentials"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
},
"source": {
"advisory": "ICSMA-22-244-01",
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "eng",
"value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened"
}
]
},
"nvd.nist.gov": {
"configurations": {
"CVE_data_version": "4.0",
"nodes": [
{
"children": [
{
"children": [],
"cpe_match": [
{
"cpe23Uri": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
"cpe_name": [],
"vulnerable": true
}
],
"operator": "OR"
},
{
"children": [],
"cpe_match": [
{
"cpe23Uri": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
"cpe_name": [],
"vulnerable": false
}
],
"operator": "OR"
}
],
"cpe_match": [],
"operator": "AND"
}
]
},
"cve": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"ID": "CVE-2022-38069"
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "en",
"value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters"
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "en",
"value": "CWE-798"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01",
"refsource": "MISC",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01"
}
]
}
},
"impact": {
"baseMetricV3": {
"cvssV3": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "NONE",
"baseScore": 6.1,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
"version": "3.1"
},
"exploitabilityScore": 0.9,
"impactScore": 5.2
}
},
"lastModifiedDate": "2022-09-14T22:51Z",
"publishedDate": "2022-09-13T15:15Z"
}
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…