ICSA-21-350-16

Vulnerability from csaf_cisa - Published: 2021-12-16 00:00 - Updated: 2022-02-10 00:00
Summary
Siemens Healthineers syngo fastView (Update A)

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could lead to a crash of the application or arbitrary code execution.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Germany
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures. CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies. Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.

{
  "document": {
    "acknowledgments": [
      {
        "names": [
          "Tran Van Khang - khangkito"
        ],
        "organization": "VinCSS",
        "summary": "reporting these vulnerabilities to Siemens and CISA"
      },
      {
        "organization": "Trend Micro Zero Day Initiative",
        "summary": "reporting these vulnerabilities to Siemens and CISA"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE",
        "url": "https://us-cert.cisa.gov/tlp/"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "legal_disclaimer",
        "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
        "title": "Legal Notice"
      },
      {
        "category": "summary",
        "text": "Successful exploitation of these vulnerabilities could lead to a crash of the application or arbitrary code execution.",
        "title": "Risk evaluation"
      },
      {
        "category": "other",
        "text": "Healthcare and Public Health",
        "title": "Critical infrastructure sectors"
      },
      {
        "category": "other",
        "text": "Worldwide",
        "title": "Countries/areas deployed"
      },
      {
        "category": "other",
        "text": "Germany",
        "title": "Company headquarters location"
      },
      {
        "category": "general",
        "text": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
        "title": "Recommended Practices"
      },
      {
        "category": "other",
        "text": "No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.",
        "title": "Exploitability"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSA-21-350-16 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2021/icsa-21-350-16.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSA-21-350-16 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-advisories/icsa-21-350-16"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/ics/tips/ICS-TIP-12-146-01B"
      }
    ],
    "title": "Siemens Healthineers syngo fastView (Update A)",
    "tracking": {
      "current_release_date": "2022-02-10T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA CSAF Generator",
          "version": "1.0.0"
        }
      },
      "id": "ICSA-21-350-16",
      "initial_release_date": "2021-12-16T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2021-12-16T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSA-21-350-16 Siemens Healthineers syngo fastView"
        },
        {
          "date": "2022-02-10T00:00:00.000000Z",
          "legacy_version": "A",
          "number": "2",
          "summary": "ICSA-21-350-16 Siemens Healthineers syngo fastView (Update A)"
        }
      ],
      "status": "final",
      "version": "2"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version",
                "name": "vers:all/*",
                "product": {
                  "name": "Syngo fastView: All versions",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "Syngo fastView"
          }
        ],
        "category": "vendor",
        "name": "Siemens Healthineers, Siemens"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2021-40367",
      "cwe": {
        "id": "CWE-787",
        "name": "Out-of-bounds Write"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected application lacks proper validation of user-supplied data when parsing DICOM files, which could result in an out-of-bounds write past the end of an allocated structure. An attacker could leverage this vulnerability to execute code in the context of the current process.CVE-2021-40367 has been assigned to this vulnerability. A CVSS v3 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-40367"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Siemens Healthineers has identi\ufb01ed the following speci\ufb01c workarounds and mitigations that customers can apply to reduce the risk:",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/en-us/medical-imaging-it/syngo-special-topics/syngo-fastview"
        },
        {
          "category": "vendor_fix",
          "details": "For additional information, please refer to Siemens Healthineers \u0027 Security Advisory SHSA-688797",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/support-documentation/cybersecurity/SHSA-688797"
        },
        {
          "category": "vendor_fix",
          "details": "In addition, Siemens Healthineers recommends the following:",
          "product_ids": [
            "CSAFPID-0001"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For further inquiries on security vulnerabilities in Siemens Healthineers products and solutions, please contact Siemens Healthineers.",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/cybersecurity"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 7.8,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2021-42028",
      "cwe": {
        "id": "CWE-787",
        "name": "Out-of-bounds Write"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected application lacks proper validation of user-supplied data when parsing BMP files, which could result in an out-of-bounds write past the end of an allocated structure. An attacker could leverage this vulnerability to execute code in the context of the current process. CVE-2021-42028 has been assigned to this vulnerability. A CVSS v3 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).. --------- Begin Update A Part 2 of 2 --------CVE-2021-42028 has been assigned to this vulnerability. A CVSS v3 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-42028"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Siemens Healthineers has identi\ufb01ed the following speci\ufb01c workarounds and mitigations that customers can apply to reduce the risk:",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/en-us/medical-imaging-it/syngo-special-topics/syngo-fastview"
        },
        {
          "category": "vendor_fix",
          "details": "For additional information, please refer to Siemens Healthineers \u0027 Security Advisory SHSA-688797",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/support-documentation/cybersecurity/SHSA-688797"
        },
        {
          "category": "vendor_fix",
          "details": "In addition, Siemens Healthineers recommends the following:",
          "product_ids": [
            "CSAFPID-0001"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For further inquiries on security vulnerabilities in Siemens Healthineers products and solutions, please contact Siemens Healthineers.",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/cybersecurity"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 7.8,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2021-45465",
      "cwe": {
        "id": "CWE-123",
        "name": "Write-what-where Condition"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected application lacks proper validation of user-supplied data when parsing BMP files. This could result in a write-what-where condition and an attacker could leverage this vulnerability to execute code in the context of the current process. CVE-2021-45465 has been assigned to this vulnerability. A CVSS v3 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).. --------- End Update A Part 2 of 2 --------CVE-2021-45465 has been assigned to this vulnerability. A CVSS v3 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-45465"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Siemens Healthineers has identi\ufb01ed the following speci\ufb01c workarounds and mitigations that customers can apply to reduce the risk:",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/en-us/medical-imaging-it/syngo-special-topics/syngo-fastview"
        },
        {
          "category": "vendor_fix",
          "details": "For additional information, please refer to Siemens Healthineers \u0027 Security Advisory SHSA-688797",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/support-documentation/cybersecurity/SHSA-688797"
        },
        {
          "category": "vendor_fix",
          "details": "In addition, Siemens Healthineers recommends the following:",
          "product_ids": [
            "CSAFPID-0001"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For further inquiries on security vulnerabilities in Siemens Healthineers products and solutions, please contact Siemens Healthineers.",
          "product_ids": [
            "CSAFPID-0001"
          ],
          "url": "https://www.siemens-healthineers.com/cybersecurity"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 7.8,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ]
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.


Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…