ICSMA-18-058-01
Vulnerability from csaf_cisa - Published: 2018-02-27 00:00 - Updated: 2020-01-30 00:00Summary
Medtronic 2090 Carelink Programmer Vulnerabilities (Update C)
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities may allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network. These credentials grant access to the software deployment network, but access is limited to read-only versions of device software applications.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Ireland
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Recommended Practices
These vulnerabilities are not exploitable remotely.
{
"document": {
"acknowledgments": [
{
"names": [
"Billy Rios",
"Jonathan Butts"
],
"organization": "Whitescope LLC",
"summary": "identifying these vulnerabilities and reporting them to CISA"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities may allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network. These credentials grant access to the software deployment network, but access is limited to read-only versions of device software applications.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "Ireland",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "These vulnerabilities are not exploitable remotely.",
"title": "Recommended Practices"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-18-058-01 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2018/icsma-18-058-01.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-18-058-01 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "Medtronic 2090 Carelink Programmer Vulnerabilities (Update C)",
"tracking": {
"current_release_date": "2020-01-30T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-18-058-01",
"initial_release_date": "2018-02-27T00:00:00.000000Z",
"revision_history": [
{
"date": "2018-02-27T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-18-058-01 Medtronic 2090 CareLink Programmer Vulnerabilities"
},
{
"date": "2018-06-28T00:00:00.000000Z",
"legacy_version": "A",
"number": "2",
"summary": "ICSMA-18-058-01 Medtronic 2090 Carelink Programmer Vulnerabilities (Update A)"
},
{
"date": "2018-10-11T00:00:00.000000Z",
"legacy_version": "B",
"number": "3",
"summary": "ICSMA-18-058-01 Medtronic 2090 CareLink Programmer Vulnerabilities (Update B)"
},
{
"date": "2020-01-30T00:00:00.000000Z",
"legacy_version": "C",
"number": "4",
"summary": "ICSMA-18-058-01 Medtronic 2090 CareLink Programmer Vulnerabilities (Update C)"
}
],
"status": "final",
"version": "4"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version",
"name": "vers:all/*",
"product": {
"name": "29901 Encore Programmer: all versions",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "29901 Encore Programmer"
},
{
"branches": [
{
"category": "product_version",
"name": "vers:all/*",
"product": {
"name": "2090 CareLink Programmer: all versions",
"product_id": "CSAFPID-0002"
}
}
],
"category": "product_name",
"name": "2090 CareLink Programmer"
}
],
"category": "vendor",
"name": "Medtronic"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2018-5446",
"cwe": {
"id": "CWE-257",
"name": "Storing Passwords in a Recoverable Format"
},
"notes": [
{
"category": "summary",
"text": "The affected product uses a per-product username and password that is stored in a recoverable format.CVE-2018-5446 has been assigned to this vulnerability. A CVSS v3 base score of 4.9 has been calculated; the CVSS vector string is (AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-5446"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800 -638 -1991.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner.\nMedtronic has released a security bulletin for the 2090 CareLink Programmer.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.medtronic.com/security"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 4.9,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002"
]
}
]
},
{
"cve": "CVE-2018-5448",
"cwe": {
"id": "CWE-23",
"name": "Relative Path Traversal"
},
"notes": [
{
"category": "summary",
"text": "The affected product \u0027s software deployment network contains a directory traversal vulnerability that could allow an attacker to read files on the system.CVE-2018-5448 has been assigned to this vulnerability. A CVSS v3 base score of 4.8 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:L/UI:N/S:U/C:H/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-5448"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:L/UI:N/S:U/C:H/I:N/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800 -638 -1991.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner.\nMedtronic has released a security bulletin for the 2090 CareLink Programmer.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.medtronic.com/security"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 4.8,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:A/AC:H/PR:L/UI:N/S:U/C:H/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002"
]
}
]
},
{
"cve": "CVE-2018-10596",
"cwe": {
"id": "CWE-923",
"name": "Improper Restriction of Communication Channel to Intended Endpoints"
},
"notes": [
{
"category": "summary",
"text": "The affected product uses a virtual private network connection to securely download updates. The product does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.CVE-2018-10596 has been assigned to this vulnerability. A CVSS v3 base score of 7.1 has been calculated; the CVSS vector string is. (AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-10596"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800 -638 -1991.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner.\nMedtronic has released a security bulletin for the 2090 CareLink Programmer.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.medtronic.com/security"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.1,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…