ICSMA-18-123-01
Vulnerability from csaf_cisa - Published: 2018-05-03 00:00 - Updated: 2019-10-10 00:00Summary
Philips Brilliance Computed Tomography (CT) System (Update A)
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities may allow an attacker to attain elevated privileges and access unauthorized system resources, including access to execute software or to view/update files including patient health information (PHI), directories, or system configuration. This could impact system confidentiality, system integrity, or system availability. Philips has received no reports of exploitation or incidents from clinical associated with these vulnerabilities.
Critical infrastructure sectors
Healthcare and public Health
Countries/areas deployed
Worldwide
Company headquarters location
U.S.
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should follow these measures:
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.
{
"document": {
"acknowledgments": [
{
"organization": "Philips",
"summary": "reporting these vulnerabilities to NCCIC"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities may allow an attacker to attain elevated privileges and access unauthorized system resources, including access to execute software or to view/update files including patient health information (PHI), directories, or system configuration. This could impact system confidentiality, system integrity, or system availability. Philips has received no reports of exploitation or incidents from clinical associated with these vulnerabilities.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "U.S.",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should follow these measures:",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "other",
"text": "No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.",
"title": "Exploitability"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-18-123-01 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2018/icsma-18-123-01.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-18-123-01 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-123-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/cas/tips/ST04-014.html"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "Philips Brilliance Computed Tomography (CT) System (Update A)",
"tracking": {
"current_release_date": "2019-10-10T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-18-123-01",
"initial_release_date": "2018-05-03T00:00:00.000000Z",
"revision_history": [
{
"date": "2018-05-03T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-18-123-01 Philips Brilliance Computed Tomography (CT) System"
},
{
"date": "2019-10-10T00:00:00.000000Z",
"legacy_version": "A",
"number": "2",
"summary": "ICSMA-18-123-01 Philips Brilliance Computed Tomography (CT) System (Update A)"
}
],
"status": "final",
"version": "2"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c=2.3.5",
"product": {
"name": "Brilliance CT Big Bore: Versions 2.3.5 and below",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "Brilliance CT Big Bore"
},
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c=4.1.6",
"product": {
"name": "Brilliance iCT: Versions 4.1.6 and below",
"product_id": "CSAFPID-0002"
}
}
],
"category": "product_name",
"name": "Brilliance iCT"
},
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c=3.2.4",
"product": {
"name": "Brilliance iCT: SP Versions 3.2.4 and below",
"product_id": "CSAFPID-0003"
}
}
],
"category": "product_name",
"name": "Brilliance iCT"
},
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c=2.6.2",
"product": {
"name": "Brilliance 64: Versions 2.6.2 and below",
"product_id": "CSAFPID-0004"
}
}
],
"category": "product_name",
"name": "Brilliance 64"
}
],
"category": "vendor",
"name": "Philips"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2018-8853",
"cwe": {
"id": "CWE-250",
"name": "Execution with Unnecessary Privileges"
},
"notes": [
{
"category": "summary",
"text": "Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system.CVE-2018-8853 has been assigned to this vulnerability. A CVSS v3 base score of 6.1 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8853"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Philips has identified the following guidance and controlling risk mitigations:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Philips has also remediated hard-coded credential vulnerabilities for Brilliance iCT 4.x and above versions. The Philips iCT-iPatient (v4.x) family Instructions for Use (IFU) refers to the ability to manage credentials and is accessible to entitled users from the Philips InCenter.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "https://incenter.medical.philips.com"
},
{
"category": "vendor_fix",
"details": "As the MX8000 Dual EXP has been out of support since 2017, Philips recommends a replacement based on user need. Please contact a local sales organization.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Philips will be further assessing options for remediation with future product introductions and/or upgrades across the CT \u0026 AMI modalities to address identified security vulnerabilities. Users with questions about their specific Brilliance CT or MX8000 Dual EXP products should contact a Philips service support team.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "http://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
},
{
"category": "vendor_fix",
"details": "The Philips \u0027 advisory is available at the following URL:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "http://www.philips.com/productsecurity (link is external)",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "http://www.philips.com/productsecurity"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.1,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
}
]
},
{
"cve": "CVE-2018-8861",
"cwe": {
"id": "CWE-668",
"name": "Exposure of Resource to Wrong Sphere"
},
"notes": [
{
"category": "summary",
"text": "Vulnerabilities within the Brilliance CT kiosk environment could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system.CVE-2018-8861 has been assigned to this vulnerability. A CVSS v3 base score of 6.1 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8861"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Philips has identified the following guidance and controlling risk mitigations:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Philips has also remediated hard-coded credential vulnerabilities for Brilliance iCT 4.x and above versions. The Philips iCT-iPatient (v4.x) family Instructions for Use (IFU) refers to the ability to manage credentials and is accessible to entitled users from the Philips InCenter.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "https://incenter.medical.philips.com"
},
{
"category": "vendor_fix",
"details": "As the MX8000 Dual EXP has been out of support since 2017, Philips recommends a replacement based on user need. Please contact a local sales organization.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Philips will be further assessing options for remediation with future product introductions and/or upgrades across the CT \u0026 AMI modalities to address identified security vulnerabilities. Users with questions about their specific Brilliance CT or MX8000 Dual EXP products should contact a Philips service support team.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "http://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
},
{
"category": "vendor_fix",
"details": "The Philips \u0027 advisory is available at the following URL:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "http://www.philips.com/productsecurity (link is external)",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "http://www.philips.com/productsecurity"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.1,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
}
]
},
{
"cve": "CVE-2018-8857",
"cwe": {
"id": "CWE-798",
"name": "Use of Hard-coded Credentials"
},
"notes": [
{
"category": "summary",
"text": "The software contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system.CVE-2018-8857 has been assigned to this vulnerability. A CVSS v3 base score of 8.4 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8857"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Philips has identified the following guidance and controlling risk mitigations:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Philips has also remediated hard-coded credential vulnerabilities for Brilliance iCT 4.x and above versions. The Philips iCT-iPatient (v4.x) family Instructions for Use (IFU) refers to the ability to manage credentials and is accessible to entitled users from the Philips InCenter.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "https://incenter.medical.philips.com"
},
{
"category": "vendor_fix",
"details": "As the MX8000 Dual EXP has been out of support since 2017, Philips recommends a replacement based on user need. Please contact a local sales organization.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Philips will be further assessing options for remediation with future product introductions and/or upgrades across the CT \u0026 AMI modalities to address identified security vulnerabilities. Users with questions about their specific Brilliance CT or MX8000 Dual EXP products should contact a Philips service support team.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "http://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
},
{
"category": "vendor_fix",
"details": "The Philips \u0027 advisory is available at the following URL:",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "http://www.philips.com/productsecurity (link is external)",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "http://www.philips.com/productsecurity"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 8.4,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…