icsma-18-156-01
Vulnerability from csaf_cisa
Published
2018-06-05 00:00
Modified
2018-06-05 00:00
Summary
Philips' IntelliVue Patient and Avalon Fetal Monitors

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation may allow an attacker to read/write memory, and/or induce a denial of service through a system restart, thus potentially leading to a delay in diagnosis and treatment of patients.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Netherlands
Recommended Practices
NCCIC recommends that users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Recommended Practices
NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Recommended Practices
NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Recommended Practices
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS-CERT website in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Exploitability
No known public exploits specifically target these vulnerabilities. These vulnerabilities are exploitable from within the same local device subnet. High skill level is needed to exploit.



{
  "document": {
    "acknowledgments": [
      {
        "names": [
          "Oran Avraham"
        ],
        "organization": "Medigate",
        "summary": "reporting these vulnerabilities to NCCIC"
      },
      {
        "organization": "Philips",
        "summary": "reporting these vulnerabilities to NCCIC"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE",
        "url": "https://us-cert.cisa.gov/tlp/"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "legal_disclaimer",
        "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
        "title": "Legal Notice"
      },
      {
        "category": "summary",
        "text": "Successful exploitation may allow an attacker to read/write memory, and/or induce a denial of service through a system restart, thus potentially leading to a delay in diagnosis and treatment of patients.",
        "title": "Risk evaluation"
      },
      {
        "category": "other",
        "text": "Healthcare and Public Health",
        "title": "Critical infrastructure sectors"
      },
      {
        "category": "other",
        "text": "Worldwide",
        "title": "Countries/areas deployed"
      },
      {
        "category": "other",
        "text": "Netherlands",
        "title": "Company headquarters location"
      },
      {
        "category": "general",
        "text": "NCCIC recommends that users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Additional mitigation guidance and recommended practices are publicly available on the ICS-CERT website in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "other",
        "text": "No known public exploits specifically target these vulnerabilities. These vulnerabilities are exploitable from within the same local device subnet. High skill level is needed to exploit.",
        "title": "Exploitability"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-18-156-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2018/icsma-18-156-01.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-18-156-01 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-156-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/ics/alerts/ICS-ALERT-10-301-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-156-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/ics/content/recommended-practices"
      }
    ],
    "title": "Philips\u0027 IntelliVue Patient and Avalon Fetal Monitors",
    "tracking": {
      "current_release_date": "2018-06-05T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA CSAF Generator",
          "version": "1.0.0"
        }
      },
      "id": "ICSMA-18-156-01",
      "initial_release_date": "2018-06-05T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2018-06-05T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSMA-18-156-01 Philips\u0027 IntelliVue Patient and Avalon Fetal Monitors"
        },
        {
          "date": "2018-06-05T00:00:00.000000Z",
          "legacy_version": "A",
          "number": "2",
          "summary": "ICSMA-18-156-01 Philips\u0027 IntelliVue Patient and Avalon Fetal Monitors (Update A)"
        }
      ],
      "status": "final",
      "version": "2"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version",
                "name": "F.0.G.0 | J.3",
                "product": {
                  "name": "Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50: with software Revisions F.0 G.0 and J.3",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "Rev J-M | (X3/MX100 for Rev M only)",
                "product": {
                  "name": "IntelliVue Patient Monitors MX (MX400-550): Rev J-M and (X3/MX100 for Rev M only)",
                  "product_id": "CSAFPID-0002"
                }
              }
            ],
            "category": "product_name",
            "name": "IntelliVue Patient Monitors MX (MX400-550)"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "Rev B-M",
                "product": {
                  "name": "IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800): Rev B-M",
                  "product_id": "CSAFPID-0003"
                }
              }
            ],
            "category": "product_name",
            "name": "IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800)"
          }
        ],
        "category": "vendor",
        "name": "Philips"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2018-10597",
      "cwe": {
        "id": "CWE-287",
        "name": "Improper Authentication"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The vulnerability allows an unauthenticated attacker to access memory (write-what-where) from an attacker-chosen device address within the same subnet.CVE-2018-10597 has been assigned to this vulnerability. A CVSS v3 base score of 8.3 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-10597"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Philips will provide a remediation patch for specific supported versions, as well as an upgrade path for all versions. Philips will communicate service options to all affected install-base users.Philips recommends users obtain associated field change and service bulletin information from Philips by accessing their InCenter account at this location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "http://incenter.medical.philips.com (link is external).",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "http://incenter.medical.philips.com"
        },
        {
          "category": "vendor_fix",
          "details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.philips.com/productsecurity (link is external).",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "https://www.philips.com/productsecurity"
        },
        {
          "category": "vendor_fix",
          "details": "Philips provides the following mitigations for these vulnerabilities:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 8.3,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2018-10599",
      "cwe": {
        "id": "CWE-200",
        "name": "Exposure of Sensitive Information to an Unauthorized Actor"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The vulnerability allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet.CVE-2018-10599 has been assigned to this vulnerability. A CVSS v3 base score of 6.4 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:L/A:L).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-10599"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:L/A:L"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Philips will provide a remediation patch for specific supported versions, as well as an upgrade path for all versions. Philips will communicate service options to all affected install-base users.Philips recommends users obtain associated field change and service bulletin information from Philips by accessing their InCenter account at this location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "http://incenter.medical.philips.com (link is external).",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "http://incenter.medical.philips.com"
        },
        {
          "category": "vendor_fix",
          "details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.philips.com/productsecurity (link is external).",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "https://www.philips.com/productsecurity"
        },
        {
          "category": "vendor_fix",
          "details": "Philips provides the following mitigations for these vulnerabilities:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 6.4,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:L/A:L",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2018-10601",
      "cwe": {
        "id": "CWE-121",
        "name": "Stack-based Buffer Overflow"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The vulnerability exposes an echo service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.CVE-2018-10601 has been assigned to this vulnerability. A CVSS v3 base score of 8.2 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-10601"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Philips will provide a remediation patch for specific supported versions, as well as an upgrade path for all versions. Philips will communicate service options to all affected install-base users.Philips recommends users obtain associated field change and service bulletin information from Philips by accessing their InCenter account at this location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "http://incenter.medical.philips.com (link is external).",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "http://incenter.medical.philips.com"
        },
        {
          "category": "vendor_fix",
          "details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.philips.com/productsecurity (link is external).",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ],
          "url": "https://www.philips.com/productsecurity"
        },
        {
          "category": "vendor_fix",
          "details": "Philips provides the following mitigations for these vulnerabilities:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 8.2,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003"
          ]
        }
      ]
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading...

Loading...

Loading...

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.