ICSMA-20-296-01
Vulnerability from csaf_cisa - Published: 2020-10-22 00:00 - Updated: 2020-10-22 00:00Summary
B. Braun OnlineSuite
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could allow an attacker to escalate privileges, download and upload arbitrary files, and perform remote code execution.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Germany
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities.
{
"document": {
"acknowledgments": [
{
"names": [
"Julian Suleder",
"Nils Emmerich",
"Birk Kauer"
],
"organization": "ERNW Research GmbH",
"summary": "reporting these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices)"
},
{
"names": [
"Dr. Oliver Matula"
],
"organization": "ERNW Enno",
"summary": "reporting these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices)"
},
{
"organization": "Rey Netzwerke GmbH",
"summary": "reporting these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices)"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities could allow an attacker to escalate privileges, download and upload arbitrary files, and perform remote code execution.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "Germany",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "other",
"text": "No known public exploits specifically target these vulnerabilities.",
"title": "Exploitability"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-20-296-01 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2020/icsma-20-296-01.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-20-296-01 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-296-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/ics/alerts/ICS-ALERT-10-301-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "B. Braun OnlineSuite",
"tracking": {
"current_release_date": "2020-10-22T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-20-296-01",
"initial_release_date": "2020-10-22T00:00:00.000000Z",
"revision_history": [
{
"date": "2020-10-22T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-20-296-01 B.Braun OnlineSuite"
}
],
"status": "final",
"version": "1"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c= 3.0",
"product": {
"name": "AP: 3.0 and earlier",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "AP"
}
],
"category": "vendor",
"name": "B. Braun Medical"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2020-25172",
"cwe": {
"id": "CWE-23",
"name": "Relative Path Traversal"
},
"notes": [
{
"category": "summary",
"text": "A relative path traversal attack in the B. Braun OnlineSuite Version AP 3.0 and earlier allows unauthenticated attackers to upload or download arbitrary files.CVE-2020-25172 has been assigned to this vulnerability. A CVSS v3 base score of 8.6 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25172"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun recommends applying the following update:",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory.",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 8.6,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2020-25174",
"cwe": {
"id": "CWE-427",
"name": "Uncontrolled Search Path Element"
},
"notes": [
{
"category": "summary",
"text": "A DLL hijacking vulnerability in the B. Braun OnlineSuite Version AP 3.0 and earlier allows local attackers to execute code on the system as a high privileged user.CVE-2020-25174 has been assigned to this vulnerability. A CVSS v3 base score of 8.4 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25174"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun recommends applying the following update:",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory.",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 8.4,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2020-25170",
"cwe": {
"id": "CWE-1236",
"name": "Improper Neutralization of Formula Elements in a CSV File"
},
"notes": [
{
"category": "summary",
"text": "An Excel Macro Injection vulnerability exists in the export feature in the B. Braun OnlineSuite Version AP 3.0 and earlier via multiple input fields that are mishandled in an Excel export.CVE-2020-25170 has been assigned to this vulnerability. A CVSS v3 base score of 6.9 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:H/UI:R/S:C/C:L/I:H/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25170"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:H/UI:R/S:C/C:L/I:H/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun recommends applying the following update:",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory.",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.9,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:H/UI:R/S:C/C:L/I:H/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…