ICSMA-20-296-02
Vulnerability from csaf_cisa - Published: 2020-10-22 00:00 - Updated: 2020-10-22 00:00Summary
B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could allow an attacker to compromise the security of the Space or compactplus communication devices, allowing an attacker to escalate privileges, view sensitive information, upload arbitrary files, and perform remote code execution.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Germany
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should: Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet; Locate control system networks and remote devices behind firewalls and isolate them from the business network; When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities.
{
"document": {
"acknowledgments": [
{
"names": [
"Julian Suleder",
"Nils Emmerich",
"Birk Kauer"
],
"organization": "ERNW Research GmbH",
"summary": "reporting these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices)"
},
{
"names": [
"Dr. Oliver Matula"
],
"organization": "ERNW Enno",
"summary": "reporting these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices)"
},
{
"organization": "Rey Netzwerke GmbH",
"summary": "reporting these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices)"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities could allow an attacker to compromise the security of the Space or compactplus communication devices, allowing an attacker to escalate privileges, view sensitive information, upload arbitrary files, and perform remote code execution.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "Germany",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should: Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet; Locate control system networks and remote devices behind firewalls and isolate them from the business network; When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "other",
"text": "No known public exploits specifically target these vulnerabilities.",
"title": "Exploitability"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-20-296-02 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2020/icsma-20-296-02.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-20-296-02 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-296-02"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/ics/alerts/ICS-ALERT-10-301-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus",
"tracking": {
"current_release_date": "2020-10-22T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-20-296-02",
"initial_release_date": "2020-10-22T00:00:00.000000Z",
"revision_history": [
{
"date": "2020-10-22T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-20-296-02 B.Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus"
}
],
"status": "final",
"version": "1"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c= U61 |\u003c= L81",
"product": {
"name": "SpaceCom: software Versions U61 and earlier (United States) L81 and earlier (outside the United States)",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "SpaceCom"
},
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c= U61 |\u003c= L81",
"product": {
"name": "Battery pack with Wi-Fi: software Versions U61 and earlier (United States) L81 and earlier (outside the United States)",
"product_id": "CSAFPID-0002"
}
}
],
"category": "product_name",
"name": "Battery pack with Wi-Fi"
},
{
"branches": [
{
"category": "product_version",
"name": "A10 | A11",
"product": {
"name": "Data module compactplus: software Versions A10 and A11 (not distributed in the United States)",
"product_id": "CSAFPID-0003"
}
}
],
"category": "product_name",
"name": "Data module compactplus"
}
],
"category": "vendor",
"name": "B. Braun Medical"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2020-25158",
"cwe": {
"id": "CWE-79",
"name": "Improper Neutralization of Input During Web Page Generation (\u0027Cross-site Scripting\u0027)"
},
"notes": [
{
"category": "summary",
"text": "A reflected cross-site scripting (XSS) vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to inject arbitrary web script or HTML into various locations. CVE-2020-25158 has been assigned to this vulnerability. A CVSS v3 base score of 7.6 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:L/A:L).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25158"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:L/A:L"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.6,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:L/A:L",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25154",
"cwe": {
"id": "CWE-601",
"name": "URL Redirection to Untrusted Site (\u0027Open Redirect\u0027)"
},
"notes": [
{
"category": "summary",
"text": "An open redirect vulnerability in the administrative interface of the B. Braun Melsungen AG SpaceCom device Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to redirect users to malicious websites. CVE-2020-25154 has been assigned to this vulnerability. A CVSS v3 base score of 5.4 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25154"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 5.4,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25162",
"cwe": {
"id": "CWE-643",
"name": "Improper Neutralization of Data within XPath Expressions (\u0027XPath Injection\u0027)"
},
"notes": [
{
"category": "summary",
"text": "A XPath injection vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows unauthenticated remote attackers to access sensitive information and escalate privileges. CVE-2020-25162 has been assigned to this vulnerability. A CVSS v3 base score of 7.5 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25162"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.5,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25152",
"cwe": {
"id": "CWE-384",
"name": "Session Fixation"
},
"notes": [
{
"category": "summary",
"text": "A session fixation vulnerability in the B. Braun Melsungen AG SpaceCom administrative interface Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to hijack web sessions and escalate privileges. CVE-2020-25152 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25152"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25164",
"cwe": {
"id": "CWE-759",
"name": "Use of a One-Way Hash without a Salt"
},
"notes": [
{
"category": "summary",
"text": "A vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to recover user credentials of the administrative interface. CVE-2020-25164 has been assigned to this vulnerability. A CVSS v3 base score of 5.1 has been assigned; the CVSS vector string is (AV:L/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25164"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 5.1,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:L/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25150",
"cwe": {
"id": "CWE-23",
"name": "Relative Path Traversal"
},
"notes": [
{
"category": "summary",
"text": "A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading a specially crafted tar file an attacker can execute arbitrary commands. CVE-2020-25150 has been assigned to this vulnerability. A CVSS v3 base score of 7.6 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25150"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.6,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25166",
"cwe": {
"id": "CWE-347",
"name": "Improper Verification of Cryptographic Signature"
},
"notes": [
{
"category": "summary",
"text": "An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices. CVE-2020-25166 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been assigned; the CVSS vector string is (AV:P/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25166"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-16238",
"cwe": {
"id": "CWE-269",
"name": "Improper Privilege Management"
},
"notes": [
{
"category": "summary",
"text": "A vulnerability in the configuration import mechanism of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with command line access to the underlying Linux system to escalate privileges to the root user. CVE-2020-16238 has been assigned to this vulnerability. A CVSS v3 base score of 6.7 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-16238"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.7,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25168",
"cwe": {
"id": "CWE-798",
"name": "Use of Hard-coded Credentials"
},
"notes": [
{
"category": "summary",
"text": "Hard-coded credentials in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 enables attackers with command line access to access the device\u0027s Wi-Fi module. CVE-2020-25168 has been assigned to this vulnerability. A CVSS v3 base score of 3.3 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25168"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 3.3,
"baseSeverity": "LOW",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25156",
"cwe": {
"id": "CWE-489",
"name": "Active Debug Code"
},
"notes": [
{
"category": "summary",
"text": "Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root. CVE-2020-25156 has been assigned to this vulnerability. A CVSS v3 base score of 7.2 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25156"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.2,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2020-25160",
"cwe": {
"id": "CWE-284",
"name": "Improper Access Control"
},
"notes": [
{
"category": "summary",
"text": "Improper access controls in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 enables attackers to extract and tamper with the devices network configuration. CVE-2020-25160 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:N/UI:N/S:C/C:L/I:L/A:L).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2020-25160"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:C/C:L/I:L/A:L"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Within the United States and Canada: Battery pack SP with Wi-Fi, software 028U00093 (SN 138852 and lower); Battery pack SP with Wi-Fi, software 054U00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software Version 012U000093. For details on acquiring this software, see the B. Braun Advisory. Note: Facilities in Canada using \u201cU\u201d versions of software should follow the U.S. vulnerability disclosure. Facilities in Canada using non \u201cU\u201d versions (e.g. L) should follow the vulnerability disclosure for outside the U.S.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "vendor_fix",
"details": "Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "vendor_fix",
"details": "B. Braun has released software updates to mitigate the reported vulnerabilities. Outside the United States and Canada: Battery Pack SP with Wi-Fi, software 027L000093 (below SN 138853); Battery pack SP with Wi-Fi, software 053L00093 (SN 138853 and higher); SpaceStation with SpaceCom 2, software version 011L000093. For more information, see the B. Braun\u0027s Vulnerability Advisory.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
],
"url": "https://www.bbraun.com/en/products-and-solutions/temp/b--braun-coordinated-vulnerability-disclosure/security-advisory.html"
},
{
"category": "mitigation",
"details": "All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
{
"category": "mitigation",
"details": "Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:C/C:L/I:L/A:L",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…