VAR-201804-0780
Vulnerability from variot - Updated: 2023-12-18 13:08Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point. Philips IntelliVue MX40 Contains a data processing vulnerability.Service operation interruption (DoS) There is a possibility of being put into a state. The MX40 Patient Worn Monitor is primarily used as a traditional telemetry medical device as part of a surveillance and alarm system. Philips IntelliView MX40 Patient Worn Monitor is prone to multiple denial-of-service vulnerabilities. Successful exploits may allow attackers to crash the affected application, resulting in denial-of-service conditions. Versions prior to Philips IntelliView MX40 Patient Worn Monitor B.06.18 are vulnerable
Show details on source website{
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/VARIoTentry#",
"affected_products": {
"@id": "https://www.variotdbs.pl/ref/affected_products"
},
"configurations": {
"@id": "https://www.variotdbs.pl/ref/configurations"
},
"credits": {
"@id": "https://www.variotdbs.pl/ref/credits"
},
"cvss": {
"@id": "https://www.variotdbs.pl/ref/cvss/"
},
"description": {
"@id": "https://www.variotdbs.pl/ref/description/"
},
"exploit_availability": {
"@id": "https://www.variotdbs.pl/ref/exploit_availability/"
},
"external_ids": {
"@id": "https://www.variotdbs.pl/ref/external_ids/"
},
"iot": {
"@id": "https://www.variotdbs.pl/ref/iot/"
},
"iot_taxonomy": {
"@id": "https://www.variotdbs.pl/ref/iot_taxonomy/"
},
"patch": {
"@id": "https://www.variotdbs.pl/ref/patch/"
},
"problemtype_data": {
"@id": "https://www.variotdbs.pl/ref/problemtype_data/"
},
"references": {
"@id": "https://www.variotdbs.pl/ref/references/"
},
"sources": {
"@id": "https://www.variotdbs.pl/ref/sources/"
},
"sources_release_date": {
"@id": "https://www.variotdbs.pl/ref/sources_release_date/"
},
"sources_update_date": {
"@id": "https://www.variotdbs.pl/ref/sources_update_date/"
},
"threat_type": {
"@id": "https://www.variotdbs.pl/ref/threat_type/"
},
"title": {
"@id": "https://www.variotdbs.pl/ref/title/"
},
"type": {
"@id": "https://www.variotdbs.pl/ref/type/"
}
},
"@id": "https://www.variotdbs.pl/vuln/VAR-201804-0780",
"affected_products": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/affected_products#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
},
"@id": "https://www.variotdbs.pl/ref/sources"
}
},
"data": [
{
"model": "intellivue mx40",
"scope": "lt",
"trust": 1.0,
"vendor": "philips",
"version": "b.06.18"
},
{
"model": "intellivue mx40",
"scope": "eq",
"trust": 0.8,
"vendor": "philips",
"version": "b.06.18"
},
{
"model": "intellivue mx40 patient worn monitor \u003cb.06.18",
"scope": null,
"trust": 0.6,
"vendor": "philips",
"version": null
},
{
"model": "intelliview mx40 patient worn monitor",
"scope": "eq",
"trust": 0.3,
"vendor": "philips",
"version": "0"
},
{
"model": "intelliview mx40 patient worn monitor b.06.18",
"scope": "ne",
"trust": 0.3,
"vendor": "philips",
"version": null
},
{
"model": null,
"scope": "eq",
"trust": 0.2,
"vendor": "intellivue mx40",
"version": "*"
}
],
"sources": [
{
"db": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1"
},
{
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"db": "BID",
"id": "100813"
},
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"db": "NVD",
"id": "CVE-2017-9657"
}
]
},
"configurations": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/configurations#",
"children": {
"@container": "@list"
},
"cpe_match": {
"@container": "@list"
},
"data": {
"@container": "@list"
},
"nodes": {
"@container": "@list"
}
},
"data": [
{
"CVE_data_version": "4.0",
"nodes": [
{
"children": [
{
"children": [],
"cpe_match": [
{
"cpe23Uri": "cpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:*",
"cpe_name": [],
"versionEndExcluding": "b.06.18",
"vulnerable": true
}
],
"operator": "OR"
},
{
"children": [],
"cpe_match": [
{
"cpe23Uri": "cpe:2.3:h:philips:intellivue_mx40:-:*:*:*:*:*:*:*",
"cpe_name": [],
"vulnerable": false
}
],
"operator": "OR"
}
],
"cpe_match": [],
"operator": "AND"
}
]
}
],
"sources": [
{
"db": "NVD",
"id": "CVE-2017-9657"
}
]
},
"credits": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/credits#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "The vendor has reported the issue.",
"sources": [
{
"db": "BID",
"id": "100813"
}
],
"trust": 0.3
},
"cve": "CVE-2017-9657",
"cvss": {
"@context": {
"cvssV2": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV2#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/cvssV2"
},
"cvssV3": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV3#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/cvssV3/"
},
"severity": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/cvss/severity#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/severity"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
},
"@id": "https://www.variotdbs.pl/ref/sources"
}
},
"data": [
{
"cvssV2": [
{
"acInsufInfo": false,
"accessComplexity": "LOW",
"accessVector": "ADJACENT_NETWORK",
"authentication": "NONE",
"author": "NVD",
"availabilityImpact": "PARTIAL",
"baseScore": 3.3,
"confidentialityImpact": "NONE",
"exploitabilityScore": 6.5,
"impactScore": 2.9,
"integrityImpact": "NONE",
"obtainAllPrivilege": false,
"obtainOtherPrivilege": false,
"obtainUserPrivilege": false,
"severity": "LOW",
"trust": 1.0,
"userInteractionRequired": false,
"vectorString": "AV:A/AC:L/Au:N/C:N/I:N/A:P",
"version": "2.0"
},
{
"acInsufInfo": null,
"accessComplexity": "Low",
"accessVector": "Adjacent Network",
"authentication": "None",
"author": "NVD",
"availabilityImpact": "Partial",
"baseScore": 3.3,
"confidentialityImpact": "None",
"exploitabilityScore": null,
"id": "CVE-2017-9657",
"impactScore": null,
"integrityImpact": "None",
"obtainAllPrivilege": null,
"obtainOtherPrivilege": null,
"obtainUserPrivilege": null,
"severity": "Low",
"trust": 0.8,
"userInteractionRequired": null,
"vectorString": "AV:A/AC:L/Au:N/C:N/I:N/A:P",
"version": "2.0"
},
{
"accessComplexity": "LOW",
"accessVector": "ADJACENT_NETWORK",
"authentication": "NONE",
"author": "CNVD",
"availabilityImpact": "COMPLETE",
"baseScore": 6.1,
"confidentialityImpact": "NONE",
"exploitabilityScore": 6.5,
"id": "CNVD-2017-26428",
"impactScore": 6.9,
"integrityImpact": "NONE",
"severity": "MEDIUM",
"trust": 0.6,
"vectorString": "AV:A/AC:L/Au:N/C:N/I:N/A:C",
"version": "2.0"
},
{
"accessComplexity": "LOW",
"accessVector": "ADJACENT_NETWORK",
"authentication": "NONE",
"author": "IVD",
"availabilityImpact": "COMPLETE",
"baseScore": 6.1,
"confidentialityImpact": "NONE",
"exploitabilityScore": 6.5,
"id": "45eefca3-087c-45ad-b591-845fcd17fed1",
"impactScore": 6.9,
"integrityImpact": "NONE",
"severity": "MEDIUM",
"trust": 0.2,
"vectorString": "AV:A/AC:L/Au:N/C:N/I:N/A:C",
"version": "2.9 [IVD]"
}
],
"cvssV3": [
{
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"author": "NVD",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"exploitabilityScore": 2.8,
"impactScore": 3.6,
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"trust": 1.0,
"userInteraction": "NONE",
"vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.0"
},
{
"attackComplexity": "Low",
"attackVector": "Adjacent Network",
"author": "NVD",
"availabilityImpact": "High",
"baseScore": 6.5,
"baseSeverity": "Medium",
"confidentialityImpact": "None",
"exploitabilityScore": null,
"id": "CVE-2017-9657",
"impactScore": null,
"integrityImpact": "None",
"privilegesRequired": "None",
"scope": "Unchanged",
"trust": 0.8,
"userInteraction": "None",
"vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.0"
}
],
"severity": [
{
"author": "NVD",
"id": "CVE-2017-9657",
"trust": 1.8,
"value": "MEDIUM"
},
{
"author": "CNVD",
"id": "CNVD-2017-26428",
"trust": 0.6,
"value": "MEDIUM"
},
{
"author": "CNNVD",
"id": "CNNVD-201706-580",
"trust": 0.6,
"value": "MEDIUM"
},
{
"author": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1",
"trust": 0.2,
"value": "MEDIUM"
}
]
}
],
"sources": [
{
"db": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1"
},
{
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"db": "NVD",
"id": "CVE-2017-9657"
},
{
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
]
},
"description": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/description#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point. Philips IntelliVue MX40 Contains a data processing vulnerability.Service operation interruption (DoS) There is a possibility of being put into a state. The MX40 Patient Worn Monitor is primarily used as a traditional telemetry medical device as part of a surveillance and alarm system. Philips IntelliView MX40 Patient Worn Monitor is prone to multiple denial-of-service vulnerabilities. \nSuccessful exploits may allow attackers to crash the affected application, resulting in denial-of-service conditions. \nVersions prior to Philips IntelliView MX40 Patient Worn Monitor B.06.18 are vulnerable",
"sources": [
{
"db": "NVD",
"id": "CVE-2017-9657"
},
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"db": "BID",
"id": "100813"
},
{
"db": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1"
}
],
"trust": 2.61
},
"external_ids": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/external_ids#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"db": "NVD",
"id": "CVE-2017-9657",
"trust": 3.5
},
{
"db": "ICS CERT",
"id": "ICSMA-17-255-01",
"trust": 3.3
},
{
"db": "BID",
"id": "100813",
"trust": 1.9
},
{
"db": "CNVD",
"id": "CNVD-2017-26428",
"trust": 0.8
},
{
"db": "CNNVD",
"id": "CNNVD-201706-580",
"trust": 0.8
},
{
"db": "JVNDB",
"id": "JVNDB-2017-013369",
"trust": 0.8
},
{
"db": "IVD",
"id": "45EEFCA3-087C-45AD-B591-845FCD17FED1",
"trust": 0.2
}
],
"sources": [
{
"db": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1"
},
{
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"db": "BID",
"id": "100813"
},
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"db": "NVD",
"id": "CVE-2017-9657"
},
{
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
]
},
"id": "VAR-201804-0780",
"iot": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/iot#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": true,
"sources": [
{
"db": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1"
},
{
"db": "CNVD",
"id": "CNVD-2017-26428"
}
],
"trust": 1.6333333
},
"iot_taxonomy": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/iot_taxonomy#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"category": [
"ICS"
],
"sub_category": null,
"trust": 0.8
}
],
"sources": [
{
"db": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1"
},
{
"db": "CNVD",
"id": "CNVD-2017-26428"
}
]
},
"last_update_date": "2023-12-18T13:08:31.240000Z",
"patch": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/patch#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"title": "Philips IntelliVue MX40 WLAN Patient Wearable Monitor Vulnerabilities (11-SEP-2017)",
"trust": 0.8,
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"title": "Philips\u0027 IntelliView MX40 Patient Worn Monitor has an unexplained patch",
"trust": 0.6,
"url": "https://www.cnvd.org.cn/patchinfo/show/102127"
},
{
"title": "Philips IntelliVue MX40 Patient Worn Monitor Security vulnerabilities",
"trust": 0.6,
"url": "http://www.cnnvd.org.cn/web/xxk/bdxqbyid.tag?id=99852"
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
]
},
"problemtype_data": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/problemtype_data#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"problemtype": "CWE-755",
"trust": 1.0
},
{
"problemtype": "CWE-19",
"trust": 0.8
}
],
"sources": [
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"db": "NVD",
"id": "CVE-2017-9657"
}
]
},
"references": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/references#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"trust": 3.3,
"url": "https://ics-cert.us-cert.gov/advisories/icsma-17-255-01"
},
{
"trust": 1.6,
"url": "http://www.securityfocus.com/bid/100813"
},
{
"trust": 1.6,
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"trust": 0.8,
"url": "https://cve.mitre.org/cgi-bin/cvename.cgi?name=cve-2017-9657"
},
{
"trust": 0.8,
"url": "https://nvd.nist.gov/vuln/detail/cve-2017-9657"
},
{
"trust": 0.3,
"url": "http://www.usa.philips.com/"
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"db": "BID",
"id": "100813"
},
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"db": "NVD",
"id": "CVE-2017-9657"
},
{
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
]
},
"sources": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#",
"data": {
"@container": "@list"
}
},
"data": [
{
"db": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1"
},
{
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"db": "BID",
"id": "100813"
},
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"db": "NVD",
"id": "CVE-2017-9657"
},
{
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
]
},
"sources_release_date": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources_release_date#",
"data": {
"@container": "@list"
}
},
"data": [
{
"date": "2017-09-13T00:00:00",
"db": "IVD",
"id": "45eefca3-087c-45ad-b591-845fcd17fed1"
},
{
"date": "2017-09-13T00:00:00",
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"date": "2017-09-12T00:00:00",
"db": "BID",
"id": "100813"
},
{
"date": "2018-06-27T00:00:00",
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"date": "2018-04-30T15:29:00.163000",
"db": "NVD",
"id": "CVE-2017-9657"
},
{
"date": "2017-09-12T00:00:00",
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
]
},
"sources_update_date": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources_update_date#",
"data": {
"@container": "@list"
}
},
"data": [
{
"date": "2017-09-13T00:00:00",
"db": "CNVD",
"id": "CNVD-2017-26428"
},
{
"date": "2017-09-12T00:00:00",
"db": "BID",
"id": "100813"
},
{
"date": "2018-06-27T00:00:00",
"db": "JVNDB",
"id": "JVNDB-2017-013369"
},
{
"date": "2019-10-09T23:30:47.063000",
"db": "NVD",
"id": "CVE-2017-9657"
},
{
"date": "2019-10-17T00:00:00",
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
]
},
"threat_type": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/threat_type#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "remote or local",
"sources": [
{
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
],
"trust": 0.6
},
"title": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/title#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "Philips IntelliVue MX40 Data processing vulnerability",
"sources": [
{
"db": "JVNDB",
"id": "JVNDB-2017-013369"
}
],
"trust": 0.8
},
"type": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/type#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "lack of information",
"sources": [
{
"db": "CNNVD",
"id": "CNNVD-201706-580"
}
],
"trust": 0.6
}
}
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.