Vulnerability from csaf_cisa - Published: 2018-04-17 00:00 - Updated: 2018-04-17 00:00
Summary
Abbott Laboratories Defibrillator

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities may allow a nearby attacker to gain unauthorized access to an ICD to issue commands, change settings, or otherwise interfere with the intended function of the ICD.
Recommended Practices
NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Recommended Practices
NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available in the ICS-CERT Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT website.
Exploitability
No known public exploits specifically target these vulnerabilities. High skill level is needed to exploit.

{
  "document": {
    "acknowledgments": [
      {
        "organization": "MedSec Holdings Ltd",
        "summary": "reporting these vulnerabilities to Abbott Laboratories and NCCIC"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE",
        "url": "https://us-cert.cisa.gov/tlp/"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "legal_disclaimer",
        "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
        "title": "Legal Notice"
      },
      {
        "category": "summary",
        "text": "Successful exploitation of these vulnerabilities may allow a nearby attacker to gain unauthorized access to an ICD to issue commands, change settings, or otherwise interfere with the intended function of the ICD.",
        "title": "Risk evaluation"
      },
      {
        "category": "general",
        "text": "NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Additional mitigation guidance and recommended practices are publicly available in the ICS-CERT Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT website.",
        "title": "Recommended Practices"
      },
      {
        "category": "other",
        "text": "No known public exploits specifically target these vulnerabilities. High skill level is needed to exploit.",
        "title": "Exploitability"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-18-107-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2018/icsma-18-107-01.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-18-107-01 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-107-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-107-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
      }
    ],
    "title": "Abbott Laboratories Defibrillator",
    "tracking": {
      "current_release_date": "2018-04-17T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA CSAF Generator",
          "version": "1.0.0"
        }
      },
      "id": "ICSMA-18-107-01",
      "initial_release_date": "2018-04-17T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2018-04-17T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSMA-18-107-01 Abbott Laboratories Defibrillator"
        }
      ],
      "status": "final",
      "version": "1"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Fortify Assura: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "Fortify Assura"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Promote Quadra: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0002"
                }
              }
            ],
            "category": "product_name",
            "name": "Promote Quadra"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Current: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0003"
                }
              }
            ],
            "category": "product_name",
            "name": "Current"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Promote: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0004"
                }
              }
            ],
            "category": "product_name",
            "name": "Promote"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Quadra Assura MP: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0005"
                }
              }
            ],
            "category": "product_name",
            "name": "Quadra Assura MP"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Unify Quadra: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0006"
                }
              }
            ],
            "category": "product_name",
            "name": "Unify Quadra"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Quadra Assura: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0007"
                }
              }
            ],
            "category": "product_name",
            "name": "Quadra Assura"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Ellipse: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0008"
                }
              }
            ],
            "category": "product_name",
            "name": "Ellipse"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Unify: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-0009"
                }
              }
            ],
            "category": "product_name",
            "name": "Unify"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Fortify: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-00010"
                }
              }
            ],
            "category": "product_name",
            "name": "Fortify"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "distributed  \u003c april 1, 2018",
                "product": {
                  "name": "Unify Assura: manufactured and distributed prior to April 1 2018",
                  "product_id": "CSAFPID-00011"
                }
              }
            ],
            "category": "product_name",
            "name": "Unify Assura"
          }
        ],
        "category": "vendor",
        "name": "Abbott Laboratories"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2017-12712",
      "cwe": {
        "id": "CWE-287",
        "name": "Improper Authentication"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The device \u0027s authentication algorithm, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the ICD or CRT-D via RF communications.CVE-2017-12712 has been assigned to this vulnerability. A CVSS v3 base score of 7.5 has been assigned; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009",
          "CSAFPID-00010",
          "CSAFPID-00011"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "nvd.nist.gov",
          "url": "https://nvd.nist.gov/vuln/detail/CVE-2017-12712"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Abbott has developed a firmware update to help mitigate the identified vulnerabilities.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "The firmware update provides additional security to reduce the risk of unauthorized access by bypassing authentication to the following high voltage device families that utilize wireless radio frequency (RF) communication: Fortify, Fortify Assura, Quadra Assura, Quadra Assura MP, Unify, Unify Assura, Unify Quadra, Promote Quadra, and Ellipse.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "The firmware update can be applied to an eligible implanted ICD or CRT-D via the Merlin PCS Programmer by a healthcare provider. Abbott and FDA have recommended the update to all eligible patients at the next regularly scheduled visit or when appropriate depending on the preferences of the patient and physician. ICDs and CRT-Ds manufactured beginning April 25, 2018, will have these updates preloaded on devices.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Abbott states that firmware updates should be approached with caution. As with any software update, firmware updates can cause devices to malfunction. Potential risks include discomfort due to back-up VVI pacing settings, reloading of previous firmware version due to incomplete upgrade, inability to treat VT/VF while in back-up mode given high voltage therapy is disabled, device remaining in back-up mode due to unsuccessful upgrade, and loss of currently-programmed device settings or diagnostic data. The Abbott Cybersecurity Medical Advisory Board has reviewed this firmware update and the associated risk of performing the update in the context of potential cybersecurity risk.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "While not intended to serve as a substitute for clinician judgment as to whether the firmware update is advisable for a particular patient, the Cybersecurity Medical Advisory Board recommends the following:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Abbott \u0027s older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Therefore, the Medical Advisory Boards recommends the following:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Patients and healthcare providers with questions can call the dedicated hotline at 1-800-722-3774 (U.S.) or visit https://www.sjm.com/cyberupdate for more information.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.sjm.com/cyberupdate%20"
        },
        {
          "category": "vendor_fix",
          "details": "Battery Performance Alert and Cybersecurity Firmware Updates for Certain Abbott (formerly St. Jude Medical) Implantable Cardiac Devices: FDA Safety Communication: FDA Safety Communication is available at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm604706.htm",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm604706.htm"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 7.5,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2017-12714",
      "cwe": {
        "id": "CWE-920",
        "name": "Improper Restriction of Power Consumption"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The ICDs and CRT-Ds do not restrict or limit the number of correctly formatted RF wake-up commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce device battery life. CVE-2017-12714 has been assigned to this vulnerability. A CVSS v3 base score of 5.3 has been assigned; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H).. Abbott is a U.S.-based company headquartered in Abbott Park, Illinois. Abbott is a U.S.-based company headquartered in Abbott Park, Illinois.. The affected ICDs and CRT-Ds are implantable medical devices designed to deliver high voltage electrical pulses to correct a fast or irregular heartbeat. According to Abbott, these devices are deployed across the healthcare and public health sector. Abbott indicates that these products are used worldwide.CVE-2017-12714 has been assigned to this vulnerability. A CVSS v3 base score of 5.3 has been assigned; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009",
          "CSAFPID-00010",
          "CSAFPID-00011"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "nvd.nist.gov",
          "url": "https://nvd.nist.gov/vuln/detail/CVE-2017-12714"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Abbott has developed a firmware update to help mitigate the identified vulnerabilities.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "The firmware update provides additional security to reduce the risk of unauthorized access by bypassing authentication to the following high voltage device families that utilize wireless radio frequency (RF) communication: Fortify, Fortify Assura, Quadra Assura, Quadra Assura MP, Unify, Unify Assura, Unify Quadra, Promote Quadra, and Ellipse.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "The firmware update can be applied to an eligible implanted ICD or CRT-D via the Merlin PCS Programmer by a healthcare provider. Abbott and FDA have recommended the update to all eligible patients at the next regularly scheduled visit or when appropriate depending on the preferences of the patient and physician. ICDs and CRT-Ds manufactured beginning April 25, 2018, will have these updates preloaded on devices.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Abbott states that firmware updates should be approached with caution. As with any software update, firmware updates can cause devices to malfunction. Potential risks include discomfort due to back-up VVI pacing settings, reloading of previous firmware version due to incomplete upgrade, inability to treat VT/VF while in back-up mode given high voltage therapy is disabled, device remaining in back-up mode due to unsuccessful upgrade, and loss of currently-programmed device settings or diagnostic data. The Abbott Cybersecurity Medical Advisory Board has reviewed this firmware update and the associated risk of performing the update in the context of potential cybersecurity risk.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "While not intended to serve as a substitute for clinician judgment as to whether the firmware update is advisable for a particular patient, the Cybersecurity Medical Advisory Board recommends the following:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Abbott \u0027s older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Therefore, the Medical Advisory Boards recommends the following:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Patients and healthcare providers with questions can call the dedicated hotline at 1-800-722-3774 (U.S.) or visit https://www.sjm.com/cyberupdate for more information.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.sjm.com/cyberupdate%20"
        },
        {
          "category": "vendor_fix",
          "details": "Battery Performance Alert and Cybersecurity Firmware Updates for Certain Abbott (formerly St. Jude Medical) Implantable Cardiac Devices: FDA Safety Communication: FDA Safety Communication is available at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm604706.htm",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ],
          "url": "https://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm604706.htm"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 5.3,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009",
            "CSAFPID-00010",
            "CSAFPID-00011"
          ]
        }
      ]
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.


Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…