ICSMA-20-261-01

Vulnerability from csaf_cisa - Published: 2020-09-17 00:00 - Updated: 2020-09-17 00:00
Summary
ICSMA-20-261-01_Philips Clinical Collaboration Platform

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Summary
Northridge Hospital Medical Center reported these vulnerabilities to Philips.
Exploitability
No known public exploits specifically target this these vulnerabilities. These vulnerabilities are not exploitable remotely.

{
  "document": {
    "acknowledgments": [
      {
        "organization": "Northridge Hospital Medical Center",
        "summary": "reporting these vulnerabilities to Philips"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "summary",
        "text": "Northridge Hospital Medical Center reported these vulnerabilities to Philips.",
        "title": "Summary"
      },
      {
        "category": "other",
        "text": "No known public exploits specifically target this these vulnerabilities. These vulnerabilities are not exploitable remotely.",
        "title": "Exploitability"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "CISAservicedesk@cisa.dhs.gov",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-20-261-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2020/icsma-20-261-01.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-20-261-01 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-261-01"
      }
    ],
    "title": "ICSMA-20-261-01_Philips Clinical Collaboration Platform",
    "tracking": {
      "current_release_date": "2020-09-17T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA USCert CSAF Generator",
          "version": "1"
        }
      },
      "id": "ICSMA-20-261-01",
      "initial_release_date": "2020-09-17T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2020-09-17T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSMA-20-261-01 Philips Clinical Collaboration Platform"
        }
      ],
      "status": "final",
      "version": "1"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c= 12.2.1",
                "product": {
                  "name": "Clinical Collaboration Platform: Versions 12.2.1 and prior",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "Clinical Collaboration Platform"
          }
        ],
        "category": "vendor",
        "name": "Philips"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2020-14506",
      "cwe": {
        "id": "CWE-668",
        "name": "Exposure of Resource to Wrong Sphere"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The product receives input or data, but it does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly.CVE-2020-14506 has been assigned to this vulnerability. A CVSS v3 base score of 3.4 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:H/UI:N/S:U/C:L/I:L/A:N).",
          "title": "Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:H/UI:N/S:U/C:L/I:L/A:N"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Philips Clinical Collaboration Platform Version 12.2.5 was released in May 2020 to remediate CWE-16, CWE-352, CWE-83, and CWE-693. Philips requires manual intervention to remediate CWE-757.",
          "product_ids": [
            "CSAFPID-0001"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 3.4,
            "baseSeverity": "LOW",
            "vectorString": "CVSS:3.0/AV:L/AC:L/PR:H/UI:N/S:U/C:L/I:L/A:N",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ],
      "title": "CVE-2020-14506"
    },
    {
      "cve": "CVE-2020-14525",
      "cwe": {
        "id": "CWE-668",
        "name": "Exposure of Resource to Wrong Sphere"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The software does not neutralize or incorrectly neutralizes user-controllable input before it is placed in output used as a webpage that is served to other users.CVE-2020-14525 has been assigned to this vulnerability. A CVSS v3 base score of 3.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N).",
          "title": "Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Philips Clinical Collaboration Platform Version 12.2.5 was released in May 2020 to remediate CWE-16, CWE-352, CWE-83, and CWE-693. Philips requires manual intervention to remediate CWE-757.",
          "product_ids": [
            "CSAFPID-0001"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 3.5,
            "baseSeverity": "LOW",
            "vectorString": "CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ],
      "title": "CVE-2020-14525"
    },
    {
      "cve": "CVE-2020-16198",
      "cwe": {
        "id": "CWE-668",
        "name": "Exposure of Resource to Wrong Sphere"
      },
      "notes": [
        {
          "category": "summary",
          "text": "When an attacker claims to have a given identity, the software does not prove or insufficiently proves the claim is correct.CVE-2020-16198 has been assigned to this vulnerability. A CVSS v3 base score of 5.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L).",
          "title": "Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Philips Clinical Collaboration Platform Version 12.2.5 was released in May 2020 to remediate CWE-16, CWE-352, CWE-83, and CWE-693. Philips requires manual intervention to remediate CWE-757.",
          "product_ids": [
            "CSAFPID-0001"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 5.0,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ],
      "title": "CVE-2020-16198"
    },
    {
      "cve": "CVE-2020-16200",
      "cwe": {
        "id": "CWE-668",
        "name": "Exposure of Resource to Wrong Sphere"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The software does not properly control the allocation and maintenance of a limited resource, thereby enabling an attacker to influence the amount of resources consumed, eventually leading to the exhaustion of available resources.CVE-2020-16200 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).",
          "title": "Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Philips Clinical Collaboration Platform Version 12.2.5 was released in May 2020 to remediate CWE-16, CWE-352, CWE-83, and CWE-693. Philips requires manual intervention to remediate CWE-757.",
          "product_ids": [
            "CSAFPID-0001"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ],
      "title": "CVE-2020-16200"
    },
    {
      "cve": "CVE-2020-16247",
      "cwe": {
        "id": "CWE-668",
        "name": "Exposure of Resource to Wrong Sphere"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource.CVE-2020-16247 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:H).",
          "title": "Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001"
        ]
      },
      "references": [
        {
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:H"
        }
      ],
      "remediations": [
        {
          "category": "mitigation",
          "details": "Philips Clinical Collaboration Platform Version 12.2.5 was released in May 2020 to remediate CWE-16, CWE-352, CWE-83, and CWE-693. Philips requires manual intervention to remediate CWE-757.",
          "product_ids": [
            "CSAFPID-0001"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 6.8,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001"
          ]
        }
      ],
      "title": "CVE-2020-16247"
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.


Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…