ICSMA-22-244-01
Vulnerability from csaf_cisa - Published: 2022-09-01 06:00 - Updated: 2025-01-30 07:00Summary
Contec Health CMS8000 Patient Monitor (Update A)
Notes
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could allow a threat actor to cause a denial-of-service condition, modify firmware with physical access to the device, access a root shell, or employ hard-coded credentials to make configuration changes.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
China
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of this these vulnerabilities, such as:
Recommended Practices
Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
Recommended Practices
Locate control system networks and remote devices behind firewalls and isolating them from business networks.
Recommended Practices
When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Recommended Practices
CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Recommended Practices
Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.
Recommended Practices
No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.
{
"document": {
"acknowledgments": [
{
"names": [
"Level Nine"
],
"summary": "reporting these vulnerabilities to CISA"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities could allow a threat actor to cause a denial-of-service condition, modify firmware with physical access to the device, access a root shell, or employ hard-coded credentials to make configuration changes.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "China",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA recommends users take defensive measures to minimize the risk of exploitation of this these vulnerabilities, such as:",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Locate control system networks and remote devices behind firewalls and isolating them from business networks.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.",
"title": "Recommended Practices"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "central@cisa.dhs.gov",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-22-244-01 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2022/icsma-22-244-01.json"
},
{
"category": "self",
"summary": "ICSA Advisory ICSMA-22-244-01 - Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-22-244-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/ics/alerts/ICS-ALERT-10-301-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/resources-tools/resources/ics-recommended-practices"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/sites/default/files/publications/Cybersecurity_Best_Practices_for_Industrial_Control_Systems.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/topics/industrial-control-systems"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/ics/tips/ICS-TIP-12-146-01B"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/sites/default/files/publications/emailscams0905.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/ncas/tips/ST04-014"
}
],
"title": "Contec Health CMS8000 Patient Monitor (Update A)",
"tracking": {
"current_release_date": "2025-01-30T07:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-22-244-01",
"initial_release_date": "2022-09-01T06:00:00.000000Z",
"revision_history": [
{
"date": "2022-09-01T06:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "Initial Publication"
},
{
"date": "2025-01-30T07:00:00.000000Z",
"legacy_version": "Update A",
"number": "2",
"summary": "Update A - Updated Risk Evaluation section vulnerabilities, Added in CVSS 4.0 scores, updated discouraged CWEs, aligned title and mitigation recommendations with ICSMA-25-031-01."
}
],
"status": "final",
"version": "2"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c=smart3250-2.6.27-wlan2.1.7.cramfs",
"product": {
"name": "Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: \u003c=smart3250-2.6.27-wlan2.1.7.cramfs",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor"
},
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c=CMS7.820.075.08/0.740.75",
"product": {
"name": "Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: \u003c=CMS7.820.075.08/0.740.75",
"product_id": "CSAFPID-0002"
}
}
],
"category": "product_name",
"name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor"
},
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c=CMS7.820.120.01/0.930.95",
"product": {
"name": "Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: \u003c=CMS7.820.120.01/0.930.95",
"product_id": "CSAFPID-0003"
}
}
],
"category": "product_name",
"name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor"
}
],
"category": "vendor",
"name": "Contec Health"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2022-36385",
"cwe": {
"id": "CWE-1263",
"name": "Improper Physical Access Control"
},
"notes": [
{
"category": "summary",
"text": "A threat actor with momentary access to the device can plug a USB drive in and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "www.cve.org",
"url": "https://www.cve.org/CVERecord?id=CVE-2022-36385"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.1#CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/4.0#CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2022-38100",
"cwe": {
"id": "CWE-770",
"name": "Allocation of Resources Without Limits or Throttling"
},
"notes": [
{
"category": "summary",
"text": "The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "www.cve.org",
"url": "https://www.cve.org/CVERecord?id=CVE-2022-38100"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.1#CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/4.0#CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.5,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2022-38069",
"cwe": {
"id": "CWE-798",
"name": "Use of Hard-coded Credentials"
},
"notes": [
{
"category": "summary",
"text": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters.",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "www.cve.org",
"url": "https://www.cve.org/CVERecord?id=CVE-2022-38069"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.1#CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/4.0#CVSS:4.0/AV:P/AC:L/AT:N/PR:H/UI:N/VC:L/VI:L/VA:L/SC:N/SI:N/SA:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 4.3,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
"version": "3.1"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2022-38453",
"cwe": {
"id": "CWE-489",
"name": "Active Debug Code"
},
"notes": [
{
"category": "summary",
"text": "Multiple binary application files on the CMS8000 device are compiled with \u0027not stripped\u0027 and \u0027debug_info\u0027 compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "www.cve.org",
"url": "https://www.cve.org/CVERecord?id=CVE-2022-38453"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.1#CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/4.0#CVSS:4.0/AV:L/AC:H/AT:N/PR:H/UI:N/VC:L/VI:L/VA:N/SC:N/SI:N/SA:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 3.0,
"baseSeverity": "LOW",
"vectorString": "CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N",
"version": "3.1"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
},
{
"cve": "CVE-2022-3027",
"cwe": {
"id": "CWE-419",
"name": "Unprotected Primary Channel"
},
"notes": [
{
"category": "summary",
"text": "The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
},
"references": [
{
"category": "external",
"summary": "www.cve.org",
"url": "https://www.cve.org/CVERecord?id=CVE-2022-3027"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.1#CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/4.0#CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:P/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 5.7,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…