ICSMA-18-310-01

Vulnerability from csaf_cisa - Published: 2018-11-06 00:00 - Updated: 2018-11-08 00:00
Summary
Roche Diagnostics Point of Care Handheld Medical Devices (Update A)

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could allow an attacker to gain unauthorized access to modify system settings or execute arbitrary code.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Switzerland
Recommended Practices
NCCIC recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:
Recommended Practices
NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS-CERT website in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities.

{
  "document": {
    "acknowledgments": [
      {
        "names": [
          "Niv Yehezkel"
        ],
        "organization": "Medigate",
        "summary": "reporting these vulnerabilities to Roche"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE",
        "url": "https://us-cert.cisa.gov/tlp/"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "legal_disclaimer",
        "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
        "title": "Legal Notice"
      },
      {
        "category": "summary",
        "text": "Successful exploitation of these vulnerabilities could allow an attacker to gain unauthorized access to modify system settings or execute arbitrary code.",
        "title": "Risk evaluation"
      },
      {
        "category": "other",
        "text": "Healthcare and Public Health",
        "title": "Critical infrastructure sectors"
      },
      {
        "category": "other",
        "text": "Worldwide",
        "title": "Countries/areas deployed"
      },
      {
        "category": "other",
        "text": "Switzerland",
        "title": "Company headquarters location"
      },
      {
        "category": "general",
        "text": "NCCIC recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Additional mitigation guidance and recommended practices are publicly available on the ICS-CERT website in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.",
        "title": "Recommended Practices"
      },
      {
        "category": "other",
        "text": "No known public exploits specifically target these vulnerabilities.",
        "title": "Exploitability"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-18-310-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2018/icsma-18-310-01.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-18-310-01 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-310-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/ics/alerts/ICS-ALERT-10-301-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/ics/tips/ICS-TIP-12-146-01B"
      }
    ],
    "title": "Roche Diagnostics Point of Care Handheld Medical Devices (Update A)",
    "tracking": {
      "current_release_date": "2018-11-08T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA CSAF Generator",
          "version": "1.0.0"
        }
      },
      "id": "ICSMA-18-310-01",
      "initial_release_date": "2018-11-06T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2018-11-06T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSMA-18-310-01 Roche Point of Care Handheld Medical Devices"
        },
        {
          "date": "2018-11-08T00:00:00.000000Z",
          "legacy_version": "A",
          "number": "2",
          "summary": "ICSMA-18-310-01 Roche Diagnostics Point of Care Handheld Medical Devices (Update A)"
        }
      ],
      "status": "final",
      "version": "2"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version",
                "name": "Accu-Chek Inform II",
                "product": {
                  "name": "Point of Care handheld medical devices - Accu-Chek Inform II",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "Point of Care handheld medical devices"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "CoaguChek Pro II",
                "product": {
                  "name": "Point of Care handheld medical devices - CoaguChek Pro II",
                  "product_id": "CSAFPID-0002"
                }
              }
            ],
            "category": "product_name",
            "name": "Point of Care handheld medical devices"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "CoaguChek XS Plus",
                "product": {
                  "name": "Point of Care handheld medical devices - CoaguChek XS Plus",
                  "product_id": "CSAFPID-0003"
                }
              }
            ],
            "category": "product_name",
            "name": "Point of Care handheld medical devices"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "CoaguChek XS Pro",
                "product": {
                  "name": "Point of Care handheld medical devices - CoaguChek XS Pro",
                  "product_id": "CSAFPID-0004"
                }
              }
            ],
            "category": "product_name",
            "name": "Point of Care handheld medical devices"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "cobas h 232 POC",
                "product": {
                  "name": "Point of Care handheld medical devices - cobas h 232 POC",
                  "product_id": "CSAFPID-0005"
                }
              }
            ],
            "category": "product_name",
            "name": "Point of Care handheld medical devices"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "base units (BU)|handheld base units (HBU)",
                "product": {
                  "name": "Point of Care handheld medical devices - Including the related base units (BU), base unit hubs and handheld base units (HBU).",
                  "product_id": "CSAFPID-0006"
                }
              }
            ],
            "category": "product_name",
            "name": "Point of Care handheld medical devices"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "Accu-Chek Inform II Base Unit Light",
                "product": {
                  "name": "Point of Care handheld medical devices - Accu-Chek Inform II Base Unit Light",
                  "product_id": "CSAFPID-0007"
                }
              }
            ],
            "category": "product_name",
            "name": "Point of Care handheld medical devices"
          },
          {
            "branches": [
              {
                "category": "product_version",
                "name": "Accu-Chek Inform II Base Unit NEW with Software 04.00.00 or newer",
                "product": {
                  "name": "Point of Care handheld medical devices - Accu-Chek Inform II Base Unit NEW with Software 04.00.00 or newer",
                  "product_id": "CSAFPID-0008"
                }
              }
            ],
            "category": "product_name",
            "name": "Point of Care handheld medical devices"
          }
        ],
        "category": "vendor",
        "name": "Roche Diagnostics"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2018-18561",
      "cwe": {
        "id": "CWE-287",
        "name": "Improper Authentication"
      },
      "notes": [
        {
          "category": "summary",
          "text": "Weak access credentials may enable attackers in the adjacent network to gain unauthorized service access via a service interface. CVE-2018-18561 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).. Affected products:CVE-2018-18561 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-18561"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Roche recommends the following mitigation procedures for connected devices (Ethernet and Wi-Fi):",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For non-connected devices:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For all affected products, Roche Diagnostic has scheduled release of new software updates with availability beginning November 2018.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For further information or concerns, please contact a local Roche Diagnostics office at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.roche.com/about/business/roche_worldwide.htm",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ],
          "url": "https://www.roche.com/about/business/roche_worldwide.htm"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2018-18562",
      "cwe": {
        "id": "CWE-78",
        "name": "Improper Neutralization of Special Elements used in an OS Command (\u0027OS Command Injection\u0027)"
      },
      "notes": [
        {
          "category": "summary",
          "text": "Insecure permissions in a service interface may allow authenticated attackers in the adjacent network to execute arbitrary commands on the operating systems. CVE-2018-18562 has been assigned to this vulnerability. A CVSS v3 base score of 8.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H).. Affected products:CVE-2018-18562 has been assigned to this vulnerability. A CVSS v3 base score of 8.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-18562"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Roche recommends the following mitigation procedures for connected devices (Ethernet and Wi-Fi):",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For non-connected devices:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For all affected products, Roche Diagnostic has scheduled release of new software updates with availability beginning November 2018.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For further information or concerns, please contact a local Roche Diagnostics office at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.roche.com/about/business/roche_worldwide.htm",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ],
          "url": "https://www.roche.com/about/business/roche_worldwide.htm"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 8.0,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2018-18563",
      "cwe": {
        "id": "CWE-434",
        "name": "Unrestricted Upload of File with Dangerous Type"
      },
      "notes": [
        {
          "category": "summary",
          "text": "A vulnerability in the software update mechanism allows an attacker in adjacent network to overwrite arbitrary files on the system through a crafted update package. CVE-2018-18563 has been assigned to this vulnerability. A CVSS v3 base score of 8.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H).. Affected products: Affected products:. Accu-Chek Inform II Instrument - all versions before 03.06.00 (serial number below 14000) / 04.03.00 (serial Number above 14000)CVE-2018-18563 has been assigned to this vulnerability. A CVSS v3 base score of 8.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-18563"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Roche recommends the following mitigation procedures for connected devices (Ethernet and Wi-Fi):",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For non-connected devices:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For all affected products, Roche Diagnostic has scheduled release of new software updates with availability beginning November 2018.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For further information or concerns, please contact a local Roche Diagnostics office at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.roche.com/about/business/roche_worldwide.htm",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ],
          "url": "https://www.roche.com/about/business/roche_worldwide.htm"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 8.0,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2018-18564",
      "cwe": {
        "id": "CWE-284",
        "name": "Improper Access Control"
      },
      "notes": [
        {
          "category": "summary",
          "text": "Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted message. CVE-2018-18564 has been assigned to this vulnerability. A CVSS v3 base score of 8.3 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H).. Affected Products:CVE-2018-18564 has been assigned to this vulnerability. A CVSS v3 base score of 8.3 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-18564"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Roche recommends the following mitigation procedures for connected devices (Ethernet and Wi-Fi):",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For non-connected devices:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For all affected products, Roche Diagnostic has scheduled release of new software updates with availability beginning November 2018.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For further information or concerns, please contact a local Roche Diagnostics office at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.roche.com/about/business/roche_worldwide.htm",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ],
          "url": "https://www.roche.com/about/business/roche_worldwide.htm"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 8.3,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2018-18565",
      "cwe": {
        "id": "CWE-284",
        "name": "Improper Access Control"
      },
      "notes": [
        {
          "category": "summary",
          "text": "Improper access control allows attackers in the adjacent network to change the instrument configuration. CVE-2018-18565 has been assigned to this vulnerability. A CVSS v3 base score of 8.2 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H).. Affected products:CVE-2018-18565 has been assigned to this vulnerability. A CVSS v3 base score of 8.2 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-18565"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Roche recommends the following mitigation procedures for connected devices (Ethernet and Wi-Fi):",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For non-connected devices:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For all affected products, Roche Diagnostic has scheduled release of new software updates with availability beginning November 2018.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "For further information or concerns, please contact a local Roche Diagnostics office at the following location:",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "https://www.roche.com/about/business/roche_worldwide.htm",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ],
          "url": "https://www.roche.com/about/business/roche_worldwide.htm"
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 8.2,
            "baseSeverity": "HIGH",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008"
          ]
        }
      ]
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.


Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…